Evaluation of Concomitant Imaging Dose in 4D-CBCT Guided Thoracic Radiotherapy
View through CrossRef
Introduction
This retrospective study evaluated the imaging dose of 4D-CBCT in patients treated with thoracic radiotherapy.
Methods
A geometric model of a 4D-CBCT system was built for Monte Carlo (MC) calculations. Percentage depth dose (PDD) and profiles of the system were measured and compared with MC calculations to verify the model. Eight lung cancer patients were retrospectively selected for 4D-CBCT imaging dose assessment. For each patient, the imaging dose was calculated using the CBCT log file, simulating CT and the MC model. Organ at risk (OAR) dose parameters from the imaging dose were calculated, and the effect of the imaging dose on OAR dose parameters when the imaging dose was added to the treatment dose was analyzed.
Results
Deviations between MC calculations and measurements for PDD and profiles were within 1.1% and 5%, respectively. For a pre-treatment 4D-CBCT scan, the mean lung and heart doses were 9.5 ± 1.4 mGy (mean value ± standard deviation, hereafter the same) and 9.7 ± 1.4 mGy, and the maximum spinal cord, esophagus and ribs doses were 17.6 ± 4.2 mGy, 17.1 ± 3.3 mGy and 37.8 ± 5.3 mGy, respectively. For an in-treatment 4D-CBCT scan, the dose parameters were 5.6 ± 1.8 mGy, 5.6 ± 1.5 mGy, 10.3 ± 3.5 mGy, 10.3 ± 4.3 mGy and 22.4 ± 5.3 mGy, respectively. The OAR dose parameters for in-treatment 4D-CBCT scans showed a relatively strong linear relationship with beam MU and number of frames. For patients receiving 5-fraction treatments and one pre-treatment and one in-treatment 4D-CBCT for each fraction, the increases in lung V20, lung V5, mean lung and heart doses, maximum spinal cord, esophagus and ribs doses were 0.04 ± 0.03%, 0.29 ± 0.23%, 76.3 ± 14.4 mGy, 76.5 ± 12.0 mGy, 76.3 ± 20.2 mGy, 70.5 ± 13.1 mGy and 101.2 ± 20.5 mGy, respectively. The increase in OAR dose parameters was proportional to the number of fractions.
Conclusion
The imaging dose of pre-treatment and in-treatment 4D-CBCT was calculated using a validated MC model. For 5-fraction treatments, the imaging dose will likely have minimal clinical impact on OAR dose parameters.
Title: Evaluation of Concomitant Imaging Dose in 4D-CBCT Guided Thoracic Radiotherapy
Description:
Introduction
This retrospective study evaluated the imaging dose of 4D-CBCT in patients treated with thoracic radiotherapy.
Methods
A geometric model of a 4D-CBCT system was built for Monte Carlo (MC) calculations.
Percentage depth dose (PDD) and profiles of the system were measured and compared with MC calculations to verify the model.
Eight lung cancer patients were retrospectively selected for 4D-CBCT imaging dose assessment.
For each patient, the imaging dose was calculated using the CBCT log file, simulating CT and the MC model.
Organ at risk (OAR) dose parameters from the imaging dose were calculated, and the effect of the imaging dose on OAR dose parameters when the imaging dose was added to the treatment dose was analyzed.
Results
Deviations between MC calculations and measurements for PDD and profiles were within 1.
1% and 5%, respectively.
For a pre-treatment 4D-CBCT scan, the mean lung and heart doses were 9.
5 ± 1.
4 mGy (mean value ± standard deviation, hereafter the same) and 9.
7 ± 1.
4 mGy, and the maximum spinal cord, esophagus and ribs doses were 17.
6 ± 4.
2 mGy, 17.
1 ± 3.
3 mGy and 37.
8 ± 5.
3 mGy, respectively.
For an in-treatment 4D-CBCT scan, the dose parameters were 5.
6 ± 1.
8 mGy, 5.
6 ± 1.
5 mGy, 10.
3 ± 3.
5 mGy, 10.
3 ± 4.
3 mGy and 22.
4 ± 5.
3 mGy, respectively.
The OAR dose parameters for in-treatment 4D-CBCT scans showed a relatively strong linear relationship with beam MU and number of frames.
For patients receiving 5-fraction treatments and one pre-treatment and one in-treatment 4D-CBCT for each fraction, the increases in lung V20, lung V5, mean lung and heart doses, maximum spinal cord, esophagus and ribs doses were 0.
04 ± 0.
03%, 0.
29 ± 0.
23%, 76.
3 ± 14.
4 mGy, 76.
5 ± 12.
0 mGy, 76.
3 ± 20.
2 mGy, 70.
5 ± 13.
1 mGy and 101.
2 ± 20.
5 mGy, respectively.
The increase in OAR dose parameters was proportional to the number of fractions.
Conclusion
The imaging dose of pre-treatment and in-treatment 4D-CBCT was calculated using a validated MC model.
For 5-fraction treatments, the imaging dose will likely have minimal clinical impact on OAR dose parameters.
Related Results
Blunt Chest Trauma and Chylothorax: A Systematic Review
Blunt Chest Trauma and Chylothorax: A Systematic Review
Abstract
Introduction: Although traumatic chylothorax is predominantly associated with penetrating injuries, instances following blunt trauma, as a rare and challenging condition, ...
Evaluation of a novel CBCT conversion method implemented in a treatment planning system
Evaluation of a novel CBCT conversion method implemented in a treatment planning system
Abstract
Background
To evaluate a novel CBCT conversion algorithm for dose calculation implemented in a research version ...
Evaluasi Dosis Tambahan dari Cone Beam Computed Tomography (CBCT) pada Image Guided Radiotherapy (IGRT) Kasus Kanker Head and Neck, Payudara, dan Pelvis
Evaluasi Dosis Tambahan dari Cone Beam Computed Tomography (CBCT) pada Image Guided Radiotherapy (IGRT) Kasus Kanker Head and Neck, Payudara, dan Pelvis
Cone Beam CT (CBCT) is an imaging system used to verify the position in Image Guided Radiotherapy (IGRT). CBCT uses wide-cone beam radiation, which is wider than that of a regular ...
Radiographic imaging in relation to the mandibular third molar: tooth characteristics, modality choice, optimization, and absorbed dose
Radiographic imaging in relation to the mandibular third molar: tooth characteristics, modality choice, optimization, and absorbed dose
Radiographic imaging of the mandibular third molar (M3) is a valuable diagnostic tool. It provides information on tooth position, root morphology, and relations to surrounding anat...
Comparing Unet training with three different datasets to correct CBCT images for prostate radiotherapy dose calculations
Comparing Unet training with three different datasets to correct CBCT images for prostate radiotherapy dose calculations
Abstract
Image intensity correction is crucial to enable cone beam computed tomography (CBCT) based radiotherapy dose calculations. This study evaluated three dif...
Presentation and Management of Cervical Thoracic Duct Cyst: A Systematic Review of the Literature
Presentation and Management of Cervical Thoracic Duct Cyst: A Systematic Review of the Literature
Abstract
Introduction
Thoracic duct cysts are an uncommon phenomenon, especially within the cervical region. Due to its limited reported cases, very little is known about its etiol...
Phrenic Nerve Block for Management of Post-Thoracic Outlet Decompression Cough: A Case Report and Literature Review
Phrenic Nerve Block for Management of Post-Thoracic Outlet Decompression Cough: A Case Report and Literature Review
Abstract
Introduction
Thoracic outlet syndrome is a group of disorders arising from compressive forces on the neurovascular bundle in that region due to different etiologies. This...
Comparison between hybrid CBCT and CBCT in detection of vertical root fracture
Comparison between hybrid CBCT and CBCT in detection of vertical root fracture
Cone beam computerized tomography (CBCT) has been wildly used in many fields of dentistry, especially to investigate the vertical root fracture (VRF). CBCT systems were consisted o...