The medtronic melody® transcatheter pulmonary valve implanted at 24‐mm diameter—it works
View through CrossRef
ObjectivesWe report the Melody valve implanted and/or expanded to 24‐mm diameter.BackgroundThe Medtronic Melody valve has been implanted up to 22 mm in the pulmonary position for over a decade.MethodsA retrospective chart review was performed on 82 patients who underwent Melody valve implant. Technical implant method, pre‐ and postimplant echocardiographic findings, and initial follow‐up were reviewed.ResultsBetween 04/2008 and 12/2011, 13 Melody valves were successfully implanted in 11 patients, median age 35 years (range 16–61 years), in the pulmonary (bioprosthetic valve, right ventricle to pulmonary artery conduit, native valve) position (n = 9), tricuspid position (bioprosthetic valve n = 3), and aortic position (bioprosthetic valve n = 1). Ten valves were delivered on a 24‐mm balloon in balloon catheter and three were implanted using a 22‐mm Ensemble balloon delivery system, followed by postdilation using a 24‐mm × 2‐cm Atlas balloon catheter. Postimplant, the median peak systolic gradient across the pulmonary valve was 7 mm Hg and median gradient across the tricuspid valve was 3 mm Hg. There was no change in gradient across the Melody valve in the aortic position where valve prosthesis–patient mismatch was present. Postimplant intracardiac echocardiography demonstrated none or mild valve regurgitation. No more than mild regurgitation was noted at a median follow‐up of 9.5 months.ConclusionsThe Melody valve can be implanted at 24 mm in the stenotic/regurgitant bioprosthetic pulmonary, tricuspid, and aortic valve, dysfunctional right ventricle to pulmonary artery conduit, and the native right ventricular outflow tract, whereas the valve remains competent with only mild regurgitation. © 2013 Wiley Periodicals, Inc.
Title: The medtronic melody® transcatheter pulmonary valve implanted at 24‐mm diameter—it works
Description:
ObjectivesWe report the Melody valve implanted and/or expanded to 24‐mm diameter.
BackgroundThe Medtronic Melody valve has been implanted up to 22 mm in the pulmonary position for over a decade.
MethodsA retrospective chart review was performed on 82 patients who underwent Melody valve implant.
Technical implant method, pre‐ and postimplant echocardiographic findings, and initial follow‐up were reviewed.
ResultsBetween 04/2008 and 12/2011, 13 Melody valves were successfully implanted in 11 patients, median age 35 years (range 16–61 years), in the pulmonary (bioprosthetic valve, right ventricle to pulmonary artery conduit, native valve) position (n = 9), tricuspid position (bioprosthetic valve n = 3), and aortic position (bioprosthetic valve n = 1).
Ten valves were delivered on a 24‐mm balloon in balloon catheter and three were implanted using a 22‐mm Ensemble balloon delivery system, followed by postdilation using a 24‐mm × 2‐cm Atlas balloon catheter.
Postimplant, the median peak systolic gradient across the pulmonary valve was 7 mm Hg and median gradient across the tricuspid valve was 3 mm Hg.
There was no change in gradient across the Melody valve in the aortic position where valve prosthesis–patient mismatch was present.
Postimplant intracardiac echocardiography demonstrated none or mild valve regurgitation.
No more than mild regurgitation was noted at a median follow‐up of 9.
5 months.
ConclusionsThe Melody valve can be implanted at 24 mm in the stenotic/regurgitant bioprosthetic pulmonary, tricuspid, and aortic valve, dysfunctional right ventricle to pulmonary artery conduit, and the native right ventricular outflow tract, whereas the valve remains competent with only mild regurgitation.
© 2013 Wiley Periodicals, Inc.
Related Results
Abstract 13177: The Association Between Insufficient Valve Expansion and Aortic Valve Calcification for Transcatheter Valve Implantation With Self-Expandable Valve
Abstract 13177: The Association Between Insufficient Valve Expansion and Aortic Valve Calcification for Transcatheter Valve Implantation With Self-Expandable Valve
Background:
Transcatheter aortic valve implantation (TAVI) has emerged as the preferred treatment for severe aortic stenosis. Nevertheless, severe calcification remains...
Early Outcomes of Percutaneous Pulmonary Valve Implantation with Pulsta and Melody Valves: The First Report from Korea
Early Outcomes of Percutaneous Pulmonary Valve Implantation with Pulsta and Melody Valves: The First Report from Korea
Percutaneous pulmonary valve implantation (PPVI) is used to treat pulmonary stenosis (PS) or pulmonary regurgitation (PR). We described our experience with PPVI, specifically valve...
The Critical Role of Presettable and Expandable Artificial Surgical Valves in the Lifetime Management of Valvular Heart Disease
The Critical Role of Presettable and Expandable Artificial Surgical Valves in the Lifetime Management of Valvular Heart Disease
With the trend of younger patients undergoing surgical valve replacement and increased life expectancy, how to account for subsequent therapies during the initial valve replacement...
“O ARRASTA POVO DO PARÁ” A (RE)EXISTÊNCIA DAS RUAS: Possibilidades do Melody e Tecno-Melody a partir das Intersubjetividades Culturais
“O ARRASTA POVO DO PARÁ” A (RE)EXISTÊNCIA DAS RUAS: Possibilidades do Melody e Tecno-Melody a partir das Intersubjetividades Culturais
“O Arrasta Povo do Pará” A (Re)Existência das Ruas: Possibilidades do Melody e Tecno-Melody a partir das Intersubjetividades Culturais trata da contextualização do ensino do melody...
An Acoustically Controlled Down-Hole Safety Valve (SCSSSV)
An Acoustically Controlled Down-Hole Safety Valve (SCSSSV)
American Institute of Mining, Metallurgical, and Petroleum Engineers, Inc.
Abstract
A surface-controlled subsurface safety valve...
Automatic Melody Composition Using Enhanced GAN
Automatic Melody Composition Using Enhanced GAN
In traditional music composition, the composer has a special knowledge of music and combines emotion and creative experience to create music. As computer technology has evolved, va...
Incidence, predictors and outcomes of Valve-in-valve (ViV) Transcatheter aortic valve replacement (TAVR): a systematic review and meta-analysis
Incidence, predictors and outcomes of Valve-in-valve (ViV) Transcatheter aortic valve replacement (TAVR): a systematic review and meta-analysis
Abstract
Background
Surgical aortic valve replacement has been the treatment of choice for patients with aortic valve disease be...
Comparison of Heart Valve Circumference Examined Before and After 10% Formalin Fixation
Comparison of Heart Valve Circumference Examined Before and After 10% Formalin Fixation
Objective: To compare the heart valve circumference before and after 10% formalin fixation. Materials and Methods: The study analyzed 63 Thai human cadaveric hearts. Each heart val...