Abstract 13177: The Association Between Insufficient Valve Expansion and Aortic Valve Calcification for Transcatheter Valve Implantation With Self-Expandable Valve

View through CrossRef
Background: Transcatheter aortic valve implantation (TAVI) has emerged as the preferred treatment for severe aortic stenosis. Nevertheless, severe calcification remains a factor leading to insufficient valve expansion and paravalvular leakage (PVL) after TAVI, mainly when using the self-expandable valve. We aimed to investigate the relationship between calcification and valve expansion using pre- and post-operative computed tomography (CT) scans, by comparing with valve phantoms as a reference standard. Methods: We enrolled 62 consecutive patients (mean age: 84.2 ± 4.4 years, 23 male) who underwent TAVI with the Evolut PRO+. We measured the aortic valve's calcium volume (Ca-Vol) by contrast-enhanced CT. The shorter diameter (mm) and the area (cm 2 ) of the base of the implanted valve were measured. We performed CT scans of valve phantoms for each size and defined the measured shorter diameter and the base area as the reference standard. We calculated the insufficiency of valve expansion as the ratio of the patient valve's shorter diameter and base area to that of the phantom valve. PVL was defined as more than mild regurgitation by transthoracic echocardiography. Results: The base area of valve phantoms was 3.6, 4.1, 6.4, and 8.4 cm 2 for 23, 26, 29, and 34 mm, respectively. The shorter diameter and base area of the implanted valve were 15.7 ± 2.6, 15.2 ± 2.8, 17.3 ± 4.0, 23.9 ± 1.8 mm, and 2.7 ± 0.7, 2.5 ± 0.6, 3.4 ± 1.0, and 6.2 ± 1.6 cm 2 for 23, 26, 29, and 34 mm (n=4, 31, 24, and 3, respectively). Significant inverse correlations were observed between Ca-Vol and shorter diameter ratio (r 2 = 0.1, P = 0.009), area ratio (r 2 = 0.1, P = 0.006), and PVL (r 2 = 0.4, P < 0.001), respectively. On the other hand, there was no significant correlation between area ratio and PVL (r 2 = 0.004, P = 0.6). Conclusion: Severe calcification in the aortic valve was associated with insufficient valve expansion and PVL for the Evolut PRO+.
Title: Abstract 13177: The Association Between Insufficient Valve Expansion and Aortic Valve Calcification for Transcatheter Valve Implantation With Self-Expandable Valve
Description:
Background: Transcatheter aortic valve implantation (TAVI) has emerged as the preferred treatment for severe aortic stenosis.
Nevertheless, severe calcification remains a factor leading to insufficient valve expansion and paravalvular leakage (PVL) after TAVI, mainly when using the self-expandable valve.
We aimed to investigate the relationship between calcification and valve expansion using pre- and post-operative computed tomography (CT) scans, by comparing with valve phantoms as a reference standard.
Methods: We enrolled 62 consecutive patients (mean age: 84.
2 ± 4.
4 years, 23 male) who underwent TAVI with the Evolut PRO+.
We measured the aortic valve's calcium volume (Ca-Vol) by contrast-enhanced CT.
The shorter diameter (mm) and the area (cm 2 ) of the base of the implanted valve were measured.
We performed CT scans of valve phantoms for each size and defined the measured shorter diameter and the base area as the reference standard.
We calculated the insufficiency of valve expansion as the ratio of the patient valve's shorter diameter and base area to that of the phantom valve.
PVL was defined as more than mild regurgitation by transthoracic echocardiography.
Results: The base area of valve phantoms was 3.
6, 4.
1, 6.
4, and 8.
4 cm 2 for 23, 26, 29, and 34 mm, respectively.
The shorter diameter and base area of the implanted valve were 15.
7 ± 2.
6, 15.
2 ± 2.
8, 17.
3 ± 4.
0, 23.
9 ± 1.
8 mm, and 2.
7 ± 0.
7, 2.
5 ± 0.
6, 3.
4 ± 1.
0, and 6.
2 ± 1.
6 cm 2 for 23, 26, 29, and 34 mm (n=4, 31, 24, and 3, respectively).
Significant inverse correlations were observed between Ca-Vol and shorter diameter ratio (r 2 = 0.
1, P = 0.
009), area ratio (r 2 = 0.
1, P = 0.
006), and PVL (r 2 = 0.
4, P < 0.
001), respectively.
On the other hand, there was no significant correlation between area ratio and PVL (r 2 = 0.
004, P = 0.
6).
Conclusion: Severe calcification in the aortic valve was associated with insufficient valve expansion and PVL for the Evolut PRO+.

Related Results

ALDOSTERONE IS A VASCULAR CALCIFICATION PROMOTING FACTOR
ALDOSTERONE IS A VASCULAR CALCIFICATION PROMOTING FACTOR
Objectives Aim: To observe aortic and cardiac aldosterone expression and its receptor changes in rats with vascular calcification, and exogenous aldosterone effec...
Emerging Evidence of IgG4-Related Disease in Pericarditis: A Systematic Review
Emerging Evidence of IgG4-Related Disease in Pericarditis: A Systematic Review
Abstract Introduction Immunoglobulin G4-related disease (IgG4-RD) is a recently identified immune-mediated condition that is debilitating and often overlooked. While IgG4-RD has be...
CLINICAL STUDY OF OLD AGED PATIENTS WITH AORTIC VALVE CALCIFICATION
CLINICAL STUDY OF OLD AGED PATIENTS WITH AORTIC VALVE CALCIFICATION
Objectives To evaluate the incidence of aortic valve calcification, and the correlation with valve function and commonly encountered disease in the aged ...
The Critical Role of Presettable and Expandable Artificial Surgical Valves in the Lifetime Management of Valvular Heart Disease
The Critical Role of Presettable and Expandable Artificial Surgical Valves in the Lifetime Management of Valvular Heart Disease
With the trend of younger patients undergoing surgical valve replacement and increased life expectancy, how to account for subsequent therapies during the initial valve replacement...
Blood pressure, hypertension, and the risk of aortic aneurysm in the UK Biobank
Blood pressure, hypertension, and the risk of aortic aneurysm in the UK Biobank
Abstract Background Although an association between elevated blood pressure and risk of aortic aneurysm is established, f...
Abstract 4369914: Successful Transfemoral TAVR in a Patient with Chronic Aortic Dissection and Severe Aortic Insufficiency
Abstract 4369914: Successful Transfemoral TAVR in a Patient with Chronic Aortic Dissection and Severe Aortic Insufficiency
Transcatheter aortic valve replacement is a well-established treatment for high-surgical-risk patients with severe aortic disease, providing a less invasive alternative to traditio...

Back to Top