Baselines of daily quality assurance for proton pencil beam scanning using the Sphinx Compact
View through CrossRef
Quality assurance (QA) is a systematic process used to determine whether the product meets specified requirements. The American Association of Physicists in Medicine (AAPM) published a comprehensive proton therapy machine QA in TG-224 to provide confidence in use. King Chulalongkorn Memorial Hospital (KCMH) recently operated the first proton therapy center in Thailand, and the data of daily QA are not available or established. This study aims to apply the control chart for establishing the baselines of daily QA for the proton pencil beam scanning system using the Sphinx Compact detector.
The Sphinx Compact was used to perform the proton daily QA test for a total of 93 days and analyzed by myQA software according to the AAPM TG-224 recommendation. The baselines obtained by a control chart were divided into 5 subjects: 1) the depth verification including the distal and proximal depth, distal falloff, and peak width of 100 MeV, 150 MeV, and 200 MeV, 2) spot characteristics including position, sigma, skewness, and intensity of 70 MeV, 100 MeV, 125 MeV, 150 MeV, 175 MeV, and 200 MeV, 3) x-ray/proton coincidence, 4) homogeneity and 5) dose output. The upper and lower control limits (UCL and LCL) of the control chart were calculated from 50 consecutive data in each test.
All the results of control limits were found to be less than the limits recommended by AAPM TG-224. The highest limits of distal depth, distal fall-off, proximal depth, and peak width were ± 0.5 mm in all tasks from 200 MeV. For spot characteristics, the highest limits of spot position, spot sigma, spot skewness, and intensity were ± 1.2 mm, ± 2.9 mm, ± 0.621, and ± 1.8 % respectively. For x-ray/proton coincidence, the limits were found to be ± 0.4 mm and ± 0.2 mm in X and Y directions, respectively. The limits of uniform field homogeneity and dose output consistency were ± 0.7 % and ± 0.010 nC, respectively.
In conclusion, the control charts obtained by Sphinx Compact can be applied to set the baselines of daily QA data for proton pencil beam scanning at Her Royal Highness Princess Maha Chakri Sirindhorn Proton Therapy Center, KCMH.
Title: Baselines of daily quality assurance for proton pencil beam scanning using the Sphinx Compact
Description:
Quality assurance (QA) is a systematic process used to determine whether the product meets specified requirements.
The American Association of Physicists in Medicine (AAPM) published a comprehensive proton therapy machine QA in TG-224 to provide confidence in use.
King Chulalongkorn Memorial Hospital (KCMH) recently operated the first proton therapy center in Thailand, and the data of daily QA are not available or established.
This study aims to apply the control chart for establishing the baselines of daily QA for the proton pencil beam scanning system using the Sphinx Compact detector.
The Sphinx Compact was used to perform the proton daily QA test for a total of 93 days and analyzed by myQA software according to the AAPM TG-224 recommendation.
The baselines obtained by a control chart were divided into 5 subjects: 1) the depth verification including the distal and proximal depth, distal falloff, and peak width of 100 MeV, 150 MeV, and 200 MeV, 2) spot characteristics including position, sigma, skewness, and intensity of 70 MeV, 100 MeV, 125 MeV, 150 MeV, 175 MeV, and 200 MeV, 3) x-ray/proton coincidence, 4) homogeneity and 5) dose output.
The upper and lower control limits (UCL and LCL) of the control chart were calculated from 50 consecutive data in each test.
All the results of control limits were found to be less than the limits recommended by AAPM TG-224.
The highest limits of distal depth, distal fall-off, proximal depth, and peak width were ± 0.
5 mm in all tasks from 200 MeV.
For spot characteristics, the highest limits of spot position, spot sigma, spot skewness, and intensity were ± 1.
2 mm, ± 2.
9 mm, ± 0.
621, and ± 1.
8 % respectively.
For x-ray/proton coincidence, the limits were found to be ± 0.
4 mm and ± 0.
2 mm in X and Y directions, respectively.
The limits of uniform field homogeneity and dose output consistency were ± 0.
7 % and ± 0.
010 nC, respectively.
In conclusion, the control charts obtained by Sphinx Compact can be applied to set the baselines of daily QA data for proton pencil beam scanning at Her Royal Highness Princess Maha Chakri Sirindhorn Proton Therapy Center, KCMH.
.
Related Results
SU‐E‐T‐47: A Monte Carlo Model of a Spot Scanning Proton Beam Based On a Synchrotron Proton Therapy Accelerator
SU‐E‐T‐47: A Monte Carlo Model of a Spot Scanning Proton Beam Based On a Synchrotron Proton Therapy Accelerator
Purpose:To build the model of a spot scanning proton beam for the dose calculation of a synchrotron proton therapy accelerator, which is capable of accelerating protons from 50 up ...
TH‐E‐ValA‐02: Simultaneous Multi‐Pencil Fan‐Beam‐Based Intensity‐Modulated Proton Therapy.
TH‐E‐ValA‐02: Simultaneous Multi‐Pencil Fan‐Beam‐Based Intensity‐Modulated Proton Therapy.
Purpose: Intensity‐modulated proton therapy (IMPT) will improve the conformality of proton radiotherapy while preserving target homogeneity and low integral dose characteristics. I...
SU‐E‐T‐470: Beam Performance of the Radiance 330 Proton Therapy System
SU‐E‐T‐470: Beam Performance of the Radiance 330 Proton Therapy System
Purpose:The ProTom Radiance 330 proton radiotherapy system is a fully functional, compact proton radiotherapy system that provides advanced proton delivery capabilities. It support...
Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields
Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields
PurposeMagnetic resonance guidance in proton therapy (MRPT) is expected to improve its current performance. The combination of magnetic fields with clinical proton beam lines poses...
Intense-Proton-Beam Transport through an Insulator Beam Guide
Intense-Proton-Beam Transport through an Insulator Beam Guide
In this paper we study intense-proton-beam transport through an insulator guide. In our previous papers [Jpn. J. Appl. Phys. 34 (1995) L520, Jpn. J. Appl. Phys. 35 (1996) L112...
Exploring the Efficacy of Once and Twice Weekly Thyroxine Dosing: A Promising Approach for Hypothyroidism Management
Exploring the Efficacy of Once and Twice Weekly Thyroxine Dosing: A Promising Approach for Hypothyroidism Management
Abstract
Introduction
Hypothyroidism is a common endocrine disorder, in which the management involves daily intake of thyroxine. However, adherence to a daily medication regimen po...
La convention de courtage en matière d'assurance
La convention de courtage en matière d'assurance
La convention de courtage d’assurance constitue un accord d’intermédiation d’assurance conclu entre un courtier d’assurance et un preneur d’assurance en vue de la conclusion ou la ...
Focusing effect of channel target on ultra-intense laser-accelerated proton beam
Focusing effect of channel target on ultra-intense laser-accelerated proton beam
In laser proton acceleration, the inevitable transverse divergence of proton beam restricts its applications in many fields. In this paper, a structured target with a properly wide...