SU‐GG‐T‐161: Commissioning IMSure 3.1 for IMRT Second Calculation Utility Program

View through CrossRef
Purpose: A program for IMRT QA for dose point calculation (IMSure Ver 3.1 from Standard Imaging) was commissioned for applications at our clinic. The purpose of this study was to set out acceptability criteria of IMRT plans, and to determine the percentage discrepancy range of point dose calculation between IMSure QA software and Eclipse treatment planning system from Varian Medical Systems. Method and Materials: This program uses the patented 3‐source model algorithm for calculating a point dose in water for IMRT plans. The data for output factors and scatter factors were measured using a PinPoint ion chamber with an active volume of 0.01 cm3, and a brass cap was used for measurement in air. The 6 MV beam was commissioned with a value of 0.14 cm for the dosimetric leaf gap which gave the closest agreement with the treatment planning system. Results: Twenty patients planned on Eclipse with 7—field IMRT for head and neck were transferred using DICOM RT to IMSure. Dose calculation at isocentre was assessed with IMSure and compared with the treatment planning results. Percentage discrepancy in the isocentre dose calculation ranged from −2.0% to 2.0% with a standard deviation of 1%. In other words, the agreement in isocentre dose calculation for 13 patients between the treatment planning system and the IMRT QA software was within +/− 1%. Measurements with an ion chamber and a cylindrical phantom were also performed, and found to be consistent with the above results. Conclusion: Our investigation of the IMSure QA software for IMRT showed good agreement with Eclipse treatment planning system to within 1% for 68% of patients sampled in our study. Our results showed that IMSure software can be a reliable tool for IMRT QA with discrepancies with Eclipse not exceeding +/− 2%.
Title: SU‐GG‐T‐161: Commissioning IMSure 3.1 for IMRT Second Calculation Utility Program
Description:
Purpose: A program for IMRT QA for dose point calculation (IMSure Ver 3.
1 from Standard Imaging) was commissioned for applications at our clinic.
The purpose of this study was to set out acceptability criteria of IMRT plans, and to determine the percentage discrepancy range of point dose calculation between IMSure QA software and Eclipse treatment planning system from Varian Medical Systems.
Method and Materials: This program uses the patented 3‐source model algorithm for calculating a point dose in water for IMRT plans.
The data for output factors and scatter factors were measured using a PinPoint ion chamber with an active volume of 0.
01 cm3, and a brass cap was used for measurement in air.
The 6 MV beam was commissioned with a value of 0.
14 cm for the dosimetric leaf gap which gave the closest agreement with the treatment planning system.
Results: Twenty patients planned on Eclipse with 7—field IMRT for head and neck were transferred using DICOM RT to IMSure.
Dose calculation at isocentre was assessed with IMSure and compared with the treatment planning results.
Percentage discrepancy in the isocentre dose calculation ranged from −2.
0% to 2.
0% with a standard deviation of 1%.
In other words, the agreement in isocentre dose calculation for 13 patients between the treatment planning system and the IMRT QA software was within +/− 1%.
Measurements with an ion chamber and a cylindrical phantom were also performed, and found to be consistent with the above results.
Conclusion: Our investigation of the IMSure QA software for IMRT showed good agreement with Eclipse treatment planning system to within 1% for 68% of patients sampled in our study.
Our results showed that IMSure software can be a reliable tool for IMRT QA with discrepancies with Eclipse not exceeding +/− 2%.

Related Results

Integrated commissioning for integrated care
Integrated commissioning for integrated care
Background: To support the success of integrated care, it must be underpinned by integrated commissioning. This workshop introduces the Integrated System Commissioning Program, dra...
Poster — Thur Eve — 71: Improved dose accuracy for plan checking IMRT breast plans
Poster — Thur Eve — 71: Improved dose accuracy for plan checking IMRT breast plans
Dose verification as part of plan checking is a critical component of high quality patient care. IMSure QA is a software platform used at the BC Cancer Agency that facilitates dose...
SU‐GG‐T‐211: An Independent MU Calculation Program for Dynamic IMRT
SU‐GG‐T‐211: An Independent MU Calculation Program for Dynamic IMRT
Purpose: The performance of an independent monitor unit calculation program is evaluated to verify the monitor unit (MU) for dynamic IMRT fields. The MU calculation program runs on...
Dosimetric comparison of 3DCRT and IMRT in radical chemoradiotherapy of squamous cell carcinoma esophagus
Dosimetric comparison of 3DCRT and IMRT in radical chemoradiotherapy of squamous cell carcinoma esophagus
ABSTRACT Background: Radical chemoradiation is the standard of treatment for locally advanced squamous cell carcinoma of esophagus and for patien...
Metaheuristic-Based IMRT Optimization for Lung Cancer Treatment under Respiratory Motion
Metaheuristic-Based IMRT Optimization for Lung Cancer Treatment under Respiratory Motion
Introduction: Intensity-Modulated Radiation Therapy (IMRT) is a widely adopted cancer treatment technique due to its precision in delivering radiation doses to tumors while minimiz...
Ten-Year Outcomes of IMRT With Chemotherapy Versus IMRT Alone for Stage Ⅱ-Ⅳa Nasopharyngeal Carcinoma: A Retrospective Study
Ten-Year Outcomes of IMRT With Chemotherapy Versus IMRT Alone for Stage Ⅱ-Ⅳa Nasopharyngeal Carcinoma: A Retrospective Study
Abstract Background Concurrent chemoradiotherapy (CCRT) is the cornerstone of treatment for patients with locoregionally advanced nasopharyngeal carcinoma (NPC). However, ...
Analysis of commissioning and procurement of adult social care in Scotland
Analysis of commissioning and procurement of adult social care in Scotland
Background: Strategic planning and commissioning of adult social care (ASC) involves forecasting population needs across a defined area, together with availability of services or r...

Back to Top