Poster — Thur Eve — 71: Improved dose accuracy for plan checking IMRT breast plans

View through CrossRef
Dose verification as part of plan checking is a critical component of high quality patient care. IMSure QA is a software platform used at the BC Cancer Agency that facilitates dose verification for both conformal and IMRT plans. We have recently initiated treating breast tangents using IMRT at the Fraser Valley Centre and noted increased dose discrepancies (mean difference of −3%) between Eclipse and IMSure's QA module. We identified two potential sources of error: air flash and tissue heterogeneity. We extend our generated fluences 3cm past the breast contour and into air to account for breathing, set‐up uncertainties and swelling. IMSure does not account for the fluence in air or air flash. We present an air‐flash‐correction factor based on the ratios of TMRs and Phantom Scatter Factors which use the field sizes of fields with and without the air flash. In addition, we present a method to improve the heterogeneity correction used by IMSure to better match that used by AAA. Effectively we remove the IMSure's inherent heterogeneity correction and manually apply a AAA‐based heterogeneity‐correction factor. We evaluated our correction factors on a sample of 8 patients (32 fields) using ANOVA methods to determine which dose corrections most accurately reproduce Eclipse's values. We found the air‐flash correction coupled with IMSure's inherent‐heterogeneity correction has the best dose accuracy (mean difference improved from −3% to 0.3%). The AAA‐heterogeneity correction alone also improved the accuracy (mean difference improved from −3% to − 1.5%), which is acceptable for plan checking purposes.
Title: Poster — Thur Eve — 71: Improved dose accuracy for plan checking IMRT breast plans
Description:
Dose verification as part of plan checking is a critical component of high quality patient care.
IMSure QA is a software platform used at the BC Cancer Agency that facilitates dose verification for both conformal and IMRT plans.
We have recently initiated treating breast tangents using IMRT at the Fraser Valley Centre and noted increased dose discrepancies (mean difference of −3%) between Eclipse and IMSure's QA module.
We identified two potential sources of error: air flash and tissue heterogeneity.
We extend our generated fluences 3cm past the breast contour and into air to account for breathing, set‐up uncertainties and swelling.
IMSure does not account for the fluence in air or air flash.
We present an air‐flash‐correction factor based on the ratios of TMRs and Phantom Scatter Factors which use the field sizes of fields with and without the air flash.
In addition, we present a method to improve the heterogeneity correction used by IMSure to better match that used by AAA.
Effectively we remove the IMSure's inherent heterogeneity correction and manually apply a AAA‐based heterogeneity‐correction factor.
We evaluated our correction factors on a sample of 8 patients (32 fields) using ANOVA methods to determine which dose corrections most accurately reproduce Eclipse's values.
We found the air‐flash correction coupled with IMSure's inherent‐heterogeneity correction has the best dose accuracy (mean difference improved from −3% to 0.
3%).
The AAA‐heterogeneity correction alone also improved the accuracy (mean difference improved from −3% to − 1.
5%), which is acceptable for plan checking purposes.

Related Results

Desmoid-Type Fibromatosis of The Breast: A Case Series
Desmoid-Type Fibromatosis of The Breast: A Case Series
Abstract IntroductionDesmoid-type fibromatosis (DTF), also called aggressive fibromatosis, is a rare, benign, locally aggressive condition. Mammary DTF originates from fibroblasts ...
Dosimetric comparison of 3DCRT and IMRT in radical chemoradiotherapy of squamous cell carcinoma esophagus
Dosimetric comparison of 3DCRT and IMRT in radical chemoradiotherapy of squamous cell carcinoma esophagus
ABSTRACT Background: Radical chemoradiation is the standard of treatment for locally advanced squamous cell carcinoma of esophagus and for patien...
Breast Carcinoma within Fibroadenoma: A Systematic Review
Breast Carcinoma within Fibroadenoma: A Systematic Review
Abstract Introduction Fibroadenoma is the most common benign breast lesion; however, it carries a potential risk of malignant transformation. This systematic review provides an ove...
SU‐GG‐T‐572: Are Constant Dose Rate VMAT Plans Clinically Acceptable? — A Site‐By‐Site Plan Comparison Study
SU‐GG‐T‐572: Are Constant Dose Rate VMAT Plans Clinically Acceptable? — A Site‐By‐Site Plan Comparison Study
Purpose: To evaluate the adequacy of constant dose rate (CDR) VMAT plans for clinical use of commonly treated sites. The clinical acceptability of CDR VMAT plans was evaluated base...
Characterizing Groundwater Quality, Recharge and Distribution under Anthropogenic conditions
Characterizing Groundwater Quality, Recharge and Distribution under Anthropogenic conditions
Awareness concerning sustainable groundwater management is gaining traction and calls for adequate understanding of the complexities of natural and anthropogenic processes and how ...

Back to Top