Biomechanical comparison of humeral nails with different distal locking mechanisms: Insafelock nails versus conventional locking nails
View through CrossRef
Objective: The aim of this study was to compare the biomechanical resistance to rotational and axial forces of a conventional locking nail with a newly designed intramedullary humeral nail developed for humeral shaft fractures with a secure locking mechanism through the distal part of the nail.
Methods: InSafeLOCK humeral nail system (group 1, TST, Istanbul, Turkey) and Expert humeral nail system (group 2, DePuy Synthes, Bettlach, Switzerland) of the same size (9 300 mm) were examined. In total, 24 fourth-generation humerus sawbones were used in the experiment. Osteotomy was performed at the humerus shaft, and a defect was created by removing 1 cm of bone. After pre-loading 5000 cycles at a frequency of 2 Hz and a force of 50e250 N for axial loading and 5000 torsion torques between 0.5 Nm and 6.5 Nm at a 2 Hz frequency for torsional loading, the failure load values of each load were recorded. Distal interlocking was performed with an endopin in group 1, while a double cortex screw was used in group 2.
Results: All samples successfully passed the cyclic loading. The initial and final stiffness values were similar between the groups after axial loading (p ¼ 0.873 and p ¼ 0.522, respectively). The mean axial failure load values in groups 1 and 2 were 2627 ± 164 N and 7141 ± 1491 N, respectively. A significant difference was found in the axial failure load values (p ¼ 0.004). Significant differences were observed between the initial and final torsional stiffness between the two groups (p ¼ 0.004 and p ¼ 0.004, respectively). No significant difference was found in the failure load values after torsional loading (11791 ± 2055 N.mm and 16997 ± 5440 N.mm) (p ¼ 0.055).
Conclusion: These results provide a biomechanical demonstration of the adequate stability of both nails after axial and rotational loading. The reliability of the newly developed InSafeLOCK humeral nail system, which does not require fluoroscopic control and an additional incision for distal locking, supports its use in the clinic.
Title: Biomechanical comparison of humeral nails with different distal locking mechanisms: Insafelock nails versus conventional locking nails
Description:
Objective: The aim of this study was to compare the biomechanical resistance to rotational and axial forces of a conventional locking nail with a newly designed intramedullary humeral nail developed for humeral shaft fractures with a secure locking mechanism through the distal part of the nail.
Methods: InSafeLOCK humeral nail system (group 1, TST, Istanbul, Turkey) and Expert humeral nail system (group 2, DePuy Synthes, Bettlach, Switzerland) of the same size (9 300 mm) were examined.
In total, 24 fourth-generation humerus sawbones were used in the experiment.
Osteotomy was performed at the humerus shaft, and a defect was created by removing 1 cm of bone.
After pre-loading 5000 cycles at a frequency of 2 Hz and a force of 50e250 N for axial loading and 5000 torsion torques between 0.
5 Nm and 6.
5 Nm at a 2 Hz frequency for torsional loading, the failure load values of each load were recorded.
Distal interlocking was performed with an endopin in group 1, while a double cortex screw was used in group 2.
Results: All samples successfully passed the cyclic loading.
The initial and final stiffness values were similar between the groups after axial loading (p ¼ 0.
873 and p ¼ 0.
522, respectively).
The mean axial failure load values in groups 1 and 2 were 2627 ± 164 N and 7141 ± 1491 N, respectively.
A significant difference was found in the axial failure load values (p ¼ 0.
004).
Significant differences were observed between the initial and final torsional stiffness between the two groups (p ¼ 0.
004 and p ¼ 0.
004, respectively).
No significant difference was found in the failure load values after torsional loading (11791 ± 2055 N.
mm and 16997 ± 5440 N.
mm) (p ¼ 0.
055).
Conclusion: These results provide a biomechanical demonstration of the adequate stability of both nails after axial and rotational loading.
The reliability of the newly developed InSafeLOCK humeral nail system, which does not require fluoroscopic control and an additional incision for distal locking, supports its use in the clinic.
Related Results
Asymptomatic Osteonecrosis of the Trochlea in an Adolescent: A Case Report
Asymptomatic Osteonecrosis of the Trochlea in an Adolescent: A Case Report
Abstract
Introduction
Osteonecrosis, also known as avascular necrosis, aseptic necrosis, or ischemic necrosis, results from a temporary or permanent halt in blood flow to a portion...
Case report: triplane fracture of distal humeral epiphysis diagnosed using ultrasonography
Case report: triplane fracture of distal humeral epiphysis diagnosed using ultrasonography
Introduction: We encountered a rare case of distal humeral triplane fracture. Ultrasonography was useful for the diagnosis. Case Report: A boy, aged 5 years 11 months, fell from a...
Periprosthetic Humeral Fractures in Shoulder Arthroplasty
Periprosthetic Humeral Fractures in Shoulder Arthroplasty
» The reported combined rates of intraoperative and postoperative periprosthetic humeral fractures range widely, from 1.2% to 19.4%.» The risk factors for an intraoperative humeral...
Pathogenic Organisms Associated with Artificial Fingernails Worn by Healthcare Workers
Pathogenic Organisms Associated with Artificial Fingernails Worn by Healthcare Workers
Objective:To determine differences in the identity and quantity of microbial flora from healthcare workers (HCWs) wearing artificial nails compared with control HCWs with native na...
Reduced Surgical Time and Higher Accuracy of Distal Locking with the Electromagnetic Targeting System in Humeral Shaft Intramedullary Nailing
Reduced Surgical Time and Higher Accuracy of Distal Locking with the Electromagnetic Targeting System in Humeral Shaft Intramedullary Nailing
ObjectiveTo evaluate the efficacy of the electromagnetic distal targeting system in the treatment of humeral shaft fractures.MethodsFrom January 2012 to December 2018, we retrospec...
Comparative study of operative treatment of distal femur fractures using retrograde intramedullary nail versus locking plate; Retrospective study
Comparative study of operative treatment of distal femur fractures using retrograde intramedullary nail versus locking plate; Retrospective study
Introduction: The overall incidence of distal femur fractures is approximately 37 per 100,000 person-years.1 These fractures are either caused by high energy trauma in younger grou...
Association of Supraspinatus Tear Type and Localisation Between Humeral Head Bone Cyst
Association of Supraspinatus Tear Type and Localisation Between Humeral Head Bone Cyst
Objective: This study aims to comprehensively explore the association between different types and locations of supraspinatus tendon tears and the presence of humeral head bone cyst...
Effect of angular stability and other locking parameters on the mechanical performance of intramedullary nails
Effect of angular stability and other locking parameters on the mechanical performance of intramedullary nails
AbstractTo extend the indications of intramedullary nails for distal or proximal fractures, nails with angle stable locking options have been developed. Studies on the mechanical e...