A method for calibrating forward scatter meters indoors
View through CrossRef
Abstract
In this study, we developed a method for calibrating forward scatter meters (FSMs) against transmissometers for measuring visibility under laboratory conditions, and we designed and constructed a novel fog chamber for calibrating FSMs indoors. The fog chamber was used to simulate a controlled visibility environment, and a transmissometer was installed on both ends of the fog chamber as the standard. The developed method involved a comparison of the FSMs and transmissometer in a group of fog events. FSMs were installed on both sides of the optical path between the transmitter and receiver of the transmissometer. The chamber door was closed, and, after creating fog to simulate low-visibility conditions, measurements were collected from the FSMs and transmissometer every minute while the fog dissipated very slowly. The measurement errors of the FSM could then be calculated. To evaluate this method, tests were conducted in the fog chamber, and the results show that the fog chamber could simulate visibility from 10 m to 10 000 m, while the FSMs could only be calibrated below a visibility of 1500 m due to the baseline of the transmissometer. The fog homogeneity was controlled within 1.2% or less, and the stability was limited to 5% per 30 min. These parameters are reasonably homogeneous and stable for FSM—transmissometer comparison. Three PWD22 FSMs were calibrated using this method, and the calibration uncertainties were estimated to be in the range of 7.8% to 9.2%.
Title: A method for calibrating forward scatter meters indoors
Description:
Abstract
In this study, we developed a method for calibrating forward scatter meters (FSMs) against transmissometers for measuring visibility under laboratory conditions, and we designed and constructed a novel fog chamber for calibrating FSMs indoors.
The fog chamber was used to simulate a controlled visibility environment, and a transmissometer was installed on both ends of the fog chamber as the standard.
The developed method involved a comparison of the FSMs and transmissometer in a group of fog events.
FSMs were installed on both sides of the optical path between the transmitter and receiver of the transmissometer.
The chamber door was closed, and, after creating fog to simulate low-visibility conditions, measurements were collected from the FSMs and transmissometer every minute while the fog dissipated very slowly.
The measurement errors of the FSM could then be calculated.
To evaluate this method, tests were conducted in the fog chamber, and the results show that the fog chamber could simulate visibility from 10 m to 10 000 m, while the FSMs could only be calibrated below a visibility of 1500 m due to the baseline of the transmissometer.
The fog homogeneity was controlled within 1.
2% or less, and the stability was limited to 5% per 30 min.
These parameters are reasonably homogeneous and stable for FSM—transmissometer comparison.
Three PWD22 FSMs were calibrated using this method, and the calibration uncertainties were estimated to be in the range of 7.
8% to 9.
2%.
Related Results
Single‐scan patient‐specific scatter correction in computed tomography using peripheral detection of scatter and compressed sensing scatter retrieval
Single‐scan patient‐specific scatter correction in computed tomography using peripheral detection of scatter and compressed sensing scatter retrieval
Purpose:X‐ray scatter results in a significant degradation of image quality in computed tomography (CT), representing a major limitation in cone‐beam CT (CBCT) and large field‐of‐v...
PET scatter estimation using deep learning U-Net architecture
PET scatter estimation using deep learning U-Net architecture
Abstract
Objective.
Positron emission tomography (PET) image reconstruction needs to be corrected for scatter in order to...
Impact of scatter correction on personalized dosimetry in selective internal radiotherapy using 166Ho-PLLA: a single-center study including Monte-Carlo simulation, phantom and patient imaging
Impact of scatter correction on personalized dosimetry in selective internal radiotherapy using 166Ho-PLLA: a single-center study including Monte-Carlo simulation, phantom and patient imaging
Abstract
Background
Developments in transarterial radioembolization led to the conception of new microspheres loaded with holmium-166 (166Ho). Howev...
TU‐D‐I‐611‐06: A Direct, Empirical Method for X‐Ray Scatter Correction in Digital Radiography and Cone‐Beam CT
TU‐D‐I‐611‐06: A Direct, Empirical Method for X‐Ray Scatter Correction in Digital Radiography and Cone‐Beam CT
Purpose: X‐ray scatter poses a severe physical limitation to image quality in cone‐beam CT (CBCT), resulting in contrast reduction, image artifacts, and lack of CT number accuracy....
Reproducibility and Correlation of X-Ray Meters for Monitoring Scatter Radiation During Fluoroscopic Procedures
Reproducibility and Correlation of X-Ray Meters for Monitoring Scatter Radiation During Fluoroscopic Procedures
Background:
The reliability and accuracy of solid-state radiation dosimeters used during clinical procedures using pulsed low-energy x ray is uncertain.
...
Recalculation of scatter fractions for homogeneous and heterogeneous geometries using Geant4 Monte Carlo simulations
Recalculation of scatter fractions for homogeneous and heterogeneous geometries using Geant4 Monte Carlo simulations
Abstract
The primary aim of this study was to determine the scatter fraction from patient, particularly in the presence of patient heterogeneities, using Monte Carlo...
Assessment of Compliance of Smart Meters for Smart Distribution System Applications for the Indian Scenario
Assessment of Compliance of Smart Meters for Smart Distribution System Applications for the Indian Scenario
The existing distribution system is being modernised with Information and Communication technology to make the grid smarter, reliable, stable, safer, resilient and sustainable. The...
Modified fast adaptive scatter kernel superposition (mfASKS) correction and its dosimetric impact on CBCT‐based proton therapy dose calculation
Modified fast adaptive scatter kernel superposition (mfASKS) correction and its dosimetric impact on CBCT‐based proton therapy dose calculation
PurposeWhile cone beam computed tomography (CBCT) is able to provide patient anatomical information, its image quality is severely degraded due to scatter contamination, which degr...