Modified fast adaptive scatter kernel superposition (mfASKS) correction and its dosimetric impact on CBCT‐based proton therapy dose calculation

View through CrossRef
PurposeWhile cone beam computed tomography (CBCT) is able to provide patient anatomical information, its image quality is severely degraded due to scatter contamination, which degrades the accuracy of CBCT‐based dose distribution estimation in proton therapy. In this work, we combined two existing scatter kernel correction methods: the point‐spread function (PSF)‐based scatter kernel derivation method and the fast adaptive scatter kernel superposition (fASKS) model, and evaluated the impact of the modified fASKS (mfASKS) correction on the accuracy of proton dose distribution estimation. To evaluate feasibility of the mfASKS approach using accurate scatter distributions, both Monte Carlo simulations and experiments were performed for an on‐board CBCT machine integrated with a proton therapy machine.MethodsWe developed a strategy to modify central intensity, constant intensity, and amplitude of the scatter kernels derived from PSFs for the fASKS model. A parameter required for the fASKS model was derived by optimizing uniformity in the mfASKS‐corrected reconstructed images. Subsequently, the mfASKS model was used to remove scatter in CBCT imaging. We quantitatively compared the Hounsfield Unit (HU) and proton stopping power ratio (SPR) images for five different phantoms. To assess improvement of dose calculation accuracy, a series of proton treatment plans were produced using the CBCT images with and without the mfASKS correction.ResultsThe accuracies of both HU and SPR intensity quantifications are improved as a result of the mfASKS correction. Mean absolute water‐equivalent path length difference to the true value decreases from 10.3 to 0.934 mm for the Gammex phantom (simulation). At the same time, mfASKS is able to offer more accurate dose distributions, especially at the distal fall‐off region where noticeable dose overestimation is observed in the uncorrected scenario. Mean absolute relative error of proton range in the pelvic phantom improves from 5.03% to 2.57% (experiment).ConclusionsmfASKS enables more accurate CBCT‐based proton dose calculation. This technique has significant implications in image‐guided radiotherapy and dose verifications in adaptive proton therapy.
Title: Modified fast adaptive scatter kernel superposition (mfASKS) correction and its dosimetric impact on CBCT‐based proton therapy dose calculation
Description:
PurposeWhile cone beam computed tomography (CBCT) is able to provide patient anatomical information, its image quality is severely degraded due to scatter contamination, which degrades the accuracy of CBCT‐based dose distribution estimation in proton therapy.
In this work, we combined two existing scatter kernel correction methods: the point‐spread function (PSF)‐based scatter kernel derivation method and the fast adaptive scatter kernel superposition (fASKS) model, and evaluated the impact of the modified fASKS (mfASKS) correction on the accuracy of proton dose distribution estimation.
To evaluate feasibility of the mfASKS approach using accurate scatter distributions, both Monte Carlo simulations and experiments were performed for an on‐board CBCT machine integrated with a proton therapy machine.
MethodsWe developed a strategy to modify central intensity, constant intensity, and amplitude of the scatter kernels derived from PSFs for the fASKS model.
A parameter required for the fASKS model was derived by optimizing uniformity in the mfASKS‐corrected reconstructed images.
Subsequently, the mfASKS model was used to remove scatter in CBCT imaging.
We quantitatively compared the Hounsfield Unit (HU) and proton stopping power ratio (SPR) images for five different phantoms.
To assess improvement of dose calculation accuracy, a series of proton treatment plans were produced using the CBCT images with and without the mfASKS correction.
ResultsThe accuracies of both HU and SPR intensity quantifications are improved as a result of the mfASKS correction.
Mean absolute water‐equivalent path length difference to the true value decreases from 10.
3 to 0.
934 mm for the Gammex phantom (simulation).
At the same time, mfASKS is able to offer more accurate dose distributions, especially at the distal fall‐off region where noticeable dose overestimation is observed in the uncorrected scenario.
Mean absolute relative error of proton range in the pelvic phantom improves from 5.
03% to 2.
57% (experiment).
ConclusionsmfASKS enables more accurate CBCT‐based proton dose calculation.
This technique has significant implications in image‐guided radiotherapy and dose verifications in adaptive proton therapy.

Related Results

Comparing Unet training with three different datasets to correct CBCT images for prostate radiotherapy dose calculations
Comparing Unet training with three different datasets to correct CBCT images for prostate radiotherapy dose calculations
Abstract Image intensity correction is crucial to enable cone beam computed tomography (CBCT) based radiotherapy dose calculations. This study evaluated three dif...
Evaluation of a novel CBCT conversion method implemented in a treatment planning system
Evaluation of a novel CBCT conversion method implemented in a treatment planning system
Abstract Background To evaluate a novel CBCT conversion algorithm for dose calculation implemented in a research version ...
Radiographic imaging in relation to the mandibular third molar: tooth characteristics, modality choice, optimization, and absorbed dose
Radiographic imaging in relation to the mandibular third molar: tooth characteristics, modality choice, optimization, and absorbed dose
Radiographic imaging of the mandibular third molar (M3) is a valuable diagnostic tool. It provides information on tooth position, root morphology, and relations to surrounding anat...
Comparison between hybrid CBCT and CBCT in detection of vertical root fracture
Comparison between hybrid CBCT and CBCT in detection of vertical root fracture
Cone beam computerized tomography (CBCT) has been wildly used in many fields of dentistry, especially to investigate the vertical root fracture (VRF). CBCT systems were consisted o...
Barriers Toward Better CBCT Knowledge, Attitude, and Practice
Barriers Toward Better CBCT Knowledge, Attitude, and Practice
Objective: This study attempts to evaluate the knowledge, attitude, and clinical practice of Iraqi dentists toward CBCT use, referral patterns, observed challenges, and the impacts...
TU‐D‐I‐611‐06: A Direct, Empirical Method for X‐Ray Scatter Correction in Digital Radiography and Cone‐Beam CT
TU‐D‐I‐611‐06: A Direct, Empirical Method for X‐Ray Scatter Correction in Digital Radiography and Cone‐Beam CT
Purpose: X‐ray scatter poses a severe physical limitation to image quality in cone‐beam CT (CBCT), resulting in contrast reduction, image artifacts, and lack of CT number accuracy....
CBCT-based synthetic CT generated using CycleGAN with HU correction for adaptive radiotherapy of nasopharyngeal carcinoma
CBCT-based synthetic CT generated using CycleGAN with HU correction for adaptive radiotherapy of nasopharyngeal carcinoma
Abstract This study aims to utilize a hybrid approach of phantom correction and deep learning for synthesized CT (sCT) images generation based on cone-beam CT (CB...

Back to Top