The effects of scatter in x‐ray computed tomography
View through CrossRef
The effects of detection of scattered radiation in x‐ray transmission CT are studied both theoretically and experimentally. It is shown that scatter induces nonlinear errors in the measurement of attenuation values which can lead to cupping, streaks, and CT number inaccuracies. It is shown that scatter effects predominate over beam spectrum hardening effects for large body parts, and that the artifact propensity is a direct function of the scatter‐to‐primary ratio. The presence of scatter induced streaks were demonstrated experimentally on both a third‐ and a fourth‐generation CT scanner using an appropriate water‐equivalent phantom. A simple model with constant scatter background leads to a correction algorithm which was tested on several phantoms and one human pelvis. The algorithm worked well on the smaller phantoms but was less successful for the larger objects. Nevertheless, it still gave substantial improvement in the pelvic scan. We demonstrated that, at least in the pelvis, scatter is a more significant source of error than beam hardening and that improved scatter correction algorithms are needed. The consequences for the quantitative interpretation of CT numbers for clinical diagnoses are discussed.
Title: The effects of scatter in x‐ray computed tomography
Description:
The effects of detection of scattered radiation in x‐ray transmission CT are studied both theoretically and experimentally.
It is shown that scatter induces nonlinear errors in the measurement of attenuation values which can lead to cupping, streaks, and CT number inaccuracies.
It is shown that scatter effects predominate over beam spectrum hardening effects for large body parts, and that the artifact propensity is a direct function of the scatter‐to‐primary ratio.
The presence of scatter induced streaks were demonstrated experimentally on both a third‐ and a fourth‐generation CT scanner using an appropriate water‐equivalent phantom.
A simple model with constant scatter background leads to a correction algorithm which was tested on several phantoms and one human pelvis.
The algorithm worked well on the smaller phantoms but was less successful for the larger objects.
Nevertheless, it still gave substantial improvement in the pelvic scan.
We demonstrated that, at least in the pelvis, scatter is a more significant source of error than beam hardening and that improved scatter correction algorithms are needed.
The consequences for the quantitative interpretation of CT numbers for clinical diagnoses are discussed.
Related Results
Single‐scan patient‐specific scatter correction in computed tomography using peripheral detection of scatter and compressed sensing scatter retrieval
Single‐scan patient‐specific scatter correction in computed tomography using peripheral detection of scatter and compressed sensing scatter retrieval
Purpose:X‐ray scatter results in a significant degradation of image quality in computed tomography (CT), representing a major limitation in cone‐beam CT (CBCT) and large field‐of‐v...
TU‐D‐I‐611‐06: A Direct, Empirical Method for X‐Ray Scatter Correction in Digital Radiography and Cone‐Beam CT
TU‐D‐I‐611‐06: A Direct, Empirical Method for X‐Ray Scatter Correction in Digital Radiography and Cone‐Beam CT
Purpose: X‐ray scatter poses a severe physical limitation to image quality in cone‐beam CT (CBCT), resulting in contrast reduction, image artifacts, and lack of CT number accuracy....
[RETRACTED] Keanu Reeves CBD Gummies v1
[RETRACTED] Keanu Reeves CBD Gummies v1
[RETRACTED]Keanu Reeves CBD Gummies ==❱❱ Huge Discounts:[HURRY UP ] Absolute Keanu Reeves CBD Gummies (Available)Order Online Only!! ❰❰= https://www.facebook.com/Keanu-Reeves-CBD-G...
Recalculation of scatter fractions for homogeneous and heterogeneous geometries using Geant4 Monte Carlo simulations
Recalculation of scatter fractions for homogeneous and heterogeneous geometries using Geant4 Monte Carlo simulations
Abstract
The primary aim of this study was to determine the scatter fraction from patient, particularly in the presence of patient heterogeneities, using Monte Carlo...
Modified fast adaptive scatter kernel superposition (mfASKS) correction and its dosimetric impact on CBCT‐based proton therapy dose calculation
Modified fast adaptive scatter kernel superposition (mfASKS) correction and its dosimetric impact on CBCT‐based proton therapy dose calculation
PurposeWhile cone beam computed tomography (CBCT) is able to provide patient anatomical information, its image quality is severely degraded due to scatter contamination, which degr...
Characterization of surgically transposed ovaries in integrated PET/CT scan in patients with cervical cancer
Characterization of surgically transposed ovaries in integrated PET/CT scan in patients with cervical cancer
AbstractBackground. The purpose of this study was to determine the ovarian findings on integrated positron emission tomography/computed tomography scans during follow‐up in cervica...
Using dual-energy CT to discriminate sediment facies in a varved sequence 
Using dual-energy CT to discriminate sediment facies in a varved sequence 
The attenuation of X-ray is influenced by the density (electron density, ρe) and the elemental composition (effective atomic number, Zeff) of the object being imaged. The ...
Towards Safer Batteries- 4D Imaging of Abuse Mechanisms in Lithium-Ion Batteries Using Synchrotron X-Ray Computed Tomography
Towards Safer Batteries- 4D Imaging of Abuse Mechanisms in Lithium-Ion Batteries Using Synchrotron X-Ray Computed Tomography
Higher energy density materials are being pushed by the research community to make lithium-ion batteries a better competitor to chemical fossil fuels for transport applications. Th...