Biometry and Intraocular Lens Power Calculation by Combined Scheimpflug-Placido Disc versus Optical Interferometry Devices
View through CrossRef
Purpose: To compare the results of the current gold standard, laser interferometry, and keratometry by the IOL-Master, with a newly developed Galilei G6 using raytracing software Okulix for intraocular lens (IOL) power calculations.
Methods: For comparison of the IOL-power calculation of both devices, we analyzed the difference between the actual one-month postoperative subjective refraction and the theoretically calculated target refraction before cataract surgery. The IOL was selected according to the IOL Master recommendation aiming for emmetropia after surgery. We analyzed the differences of the measurements of the basic biometric data in 205 healthy eyes by each device.
Results: Our study included 205 healthy, unoperated eyes from 117 patients (61 women, 56 men) aged 20 to 75 years. Twenty-two eyes of cataract patients were also included in this retrospective study design. The mean difference between the prediction of the postoperative refraction and the refraction actually achieved was 0.03 D for the IOL Master and –0.23 D for the Galilei G6. The difference was not statistically significant (P = 0.059). The difference between the IOL power calculation of the IOL Master and the calculation of the G6 was not statistically significant (P = 0.064). The difference between the predicted refraction of the G6 and the refraction achieved after one month was also not statistically significant (P = 0.12) and neither was the difference between the predicted refraction of the IOL Master and the achieved refraction (P = 0.39). The mean axial length was calculated as 24.21 ± 0.80 mm using the IOL Master and 24.27 ± 0.82 mm using the Galilei G6 device. The mean value regarding anterior chamber depth (ACD) of the IOL master was 3.46 ± 0.23 mm and for the Galilei was G6 3.51 ± 0.25 mm. When comparing the white to white (WTW) values of the IOL master, it showed mean values of 12.32 ± 0.31 and Galilei showed mean values of G6 12.21 ± 0.28. All of these differences (between Galileo and IOL Master measurements) were statistically significant (P < 0.001).
Conclusion: Both the laser interferometry/keratometry performed by the IOL Master and the interferometry/raytracing biometry strategy performed by the Galilei G6 demonstrated equal results when executing the IOL power calculation before cataract surgery in eyes with no prior ocular surgery.
Title: Biometry and Intraocular Lens Power Calculation by Combined Scheimpflug-Placido Disc versus Optical Interferometry Devices
Description:
Purpose: To compare the results of the current gold standard, laser interferometry, and keratometry by the IOL-Master, with a newly developed Galilei G6 using raytracing software Okulix for intraocular lens (IOL) power calculations.
Methods: For comparison of the IOL-power calculation of both devices, we analyzed the difference between the actual one-month postoperative subjective refraction and the theoretically calculated target refraction before cataract surgery.
The IOL was selected according to the IOL Master recommendation aiming for emmetropia after surgery.
We analyzed the differences of the measurements of the basic biometric data in 205 healthy eyes by each device.
Results: Our study included 205 healthy, unoperated eyes from 117 patients (61 women, 56 men) aged 20 to 75 years.
Twenty-two eyes of cataract patients were also included in this retrospective study design.
The mean difference between the prediction of the postoperative refraction and the refraction actually achieved was 0.
03 D for the IOL Master and –0.
23 D for the Galilei G6.
The difference was not statistically significant (P = 0.
059).
The difference between the IOL power calculation of the IOL Master and the calculation of the G6 was not statistically significant (P = 0.
064).
The difference between the predicted refraction of the G6 and the refraction achieved after one month was also not statistically significant (P = 0.
12) and neither was the difference between the predicted refraction of the IOL Master and the achieved refraction (P = 0.
39).
The mean axial length was calculated as 24.
21 ± 0.
80 mm using the IOL Master and 24.
27 ± 0.
82 mm using the Galilei G6 device.
The mean value regarding anterior chamber depth (ACD) of the IOL master was 3.
46 ± 0.
23 mm and for the Galilei was G6 3.
51 ± 0.
25 mm.
When comparing the white to white (WTW) values of the IOL master, it showed mean values of 12.
32 ± 0.
31 and Galilei showed mean values of G6 12.
21 ± 0.
28.
All of these differences (between Galileo and IOL Master measurements) were statistically significant (P < 0.
001).
Conclusion: Both the laser interferometry/keratometry performed by the IOL Master and the interferometry/raytracing biometry strategy performed by the Galilei G6 demonstrated equal results when executing the IOL power calculation before cataract surgery in eyes with no prior ocular surgery.
Related Results
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Objective: To compare the axial length and intraocular lens power calculated by using ultrasound biometry and optical biometry in patients with cataract
Methodology: It was prospec...
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Abstract
Background: To find agreement between the calculated intraocular lens (IOL) power from using the SRK/T based preoperative biometry and the intraoperative wavefront...
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Abstract
Background: As people now are expecting more precise and predictable results from cataract surgery. We need to find agreement between the calculated intraocular le...
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Abstract
Background : To find agreement between the calculated intraocular lens (IOL) power from using the SRK/T based preoperative biometry and the intraoperative wavefron...
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Comparison of IOL Power Calculated by Preoperative Biometry versus Intraoperative Wavefront Aberrometry in Thai Cataract Patients.
Abstract
Background: To find agreement between the calculated intraocular lens (IOL) power from using the SRK/T based preoperative biometry and the intraoperative wavefront...
Comparative Analysis of Refractive Outcomes Using Optical and Ultrasound Biometry in Phacoemulsification Cataract Surgery
Comparative Analysis of Refractive Outcomes Using Optical and Ultrasound Biometry in Phacoemulsification Cataract Surgery
Background: Accurate biometry is critical for optimal visual outcomes in cataract surgery. Both optical and ultrasound biometry are widely used, but their comparative effectiveness...
Repeatability and Agreement of Horizontal Corneal Diameter Measurements Between Scanning-Slit Topography, Dual Rotating Scheimpflug Camera With Placido Disc Tomography, Placido Disc Topography, and Optical Coherence Tomography
Repeatability and Agreement of Horizontal Corneal Diameter Measurements Between Scanning-Slit Topography, Dual Rotating Scheimpflug Camera With Placido Disc Tomography, Placido Disc Topography, and Optical Coherence Tomography
Purpose:
The purpose of this study was to assess the repeatability and agreement of horizontal white-to-white (WTW) measurements using 4 different imaging modalities in...
Impact and Correction of an Anterior Phakic Intraocular Lens on Swept-Source Optical Coherence Tomography Biometry
Impact and Correction of an Anterior Phakic Intraocular Lens on Swept-Source Optical Coherence Tomography Biometry
Purpose:
To evaluate the impact of anterior chamber phakic intraocular lens (pIOL) on swept-source optical coherence tomography (SS-OCT) biometric measurements and IOL ...