Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the NIM, China, and the BIPM in low-energy x-rays
View through CrossRef
A key comparison has been made between the air-kerma standards of the NIM and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the combined standard uncertainty of 3.7 parts in 10
3
. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.
KEY WORDS FOR SEARCH
Radiation dosimetry, key comparison, low-energy x-rays, air kerma
Main text
To reach the main text of this paper, click on
Final Report
. Note that this text is that which appears in Appendix B of the BIPM key comparison database
kcdb.bipm.org/
.
The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
Title: Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the NIM, China, and the BIPM in low-energy x-rays
Description:
A key comparison has been made between the air-kerma standards of the NIM and the BIPM in the low-energy x-ray range.
The results show the standards to be in agreement at the level of the combined standard uncertainty of 3.
7 parts in 10
3
.
The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.
KEY WORDS FOR SEARCH
Radiation dosimetry, key comparison, low-energy x-rays, air kerma
Main text
To reach the main text of this paper, click on
Final Report
.
Note that this text is that which appears in Appendix B of the BIPM key comparison database
kcdb.
bipm.
org/
.
The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
Related Results
Bilateral comparison of 10 pF and 100 pF standards (ongoing BIPM key comparisons BIPM.EM-K14.a and 14.b) between the NIMT (Thailand) and the BIPM
Bilateral comparison of 10 pF and 100 pF standards (ongoing BIPM key comparisons BIPM.EM-K14.a and 14.b) between the NIMT (Thailand) and the BIPM
This report presents the results of a bilateral comparison of 10 pF and 100 pF capacitance standards between the NIMT (Thailand) and the BIPM carried out in 2018. The comparison at...
Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NIM, China and the BIPM in medium-energy x-rays
Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NIM, China and the BIPM in medium-energy x-rays
A key comparison has been made between the air-kerma standards of the NIM, Republic of China and the BIPM in the medium-energy x ray range. The results show the standards to be in ...
Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the NIM, China and the BIPM in mammography x-rays
Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the NIM, China and the BIPM in mammography x-rays
A first key comparison has been made between the air-kerma standards of the NIM, Republic of China and the BIPM in mammography x-ray beams. The results show the standards to be in ...
Comparison of the Josephson Voltage Standards of the NIMT and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b)
Comparison of the Josephson Voltage Standards of the NIMT and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b)
A comparison of the Josephson array voltage standards of the Bureau International des Poids et Mesures (BIPM) and the National Institute of Metrology - (Thailand), NIMT, was carrie...
Monte Carlo simulation of air kerma rate constants for different radionuclides
Monte Carlo simulation of air kerma rate constants for different radionuclides
The air kerma rate constant values for different radionuclides are often used in radiation shielding calculations, calibration of instruments, estimating the absorbed dose from bra...
Final report, On-going Key Comparison BIPM.QM-K1, Ozone at ambient level, comparison with NIM Oct.2023
Final report, On-going Key Comparison BIPM.QM-K1, Ozone at ambient level, comparison with NIM Oct.2023
Main text
As part of the ongoing key comparison BIPM.QM-K1, a comparison has been performed between the ozone national standard of China maintained by the National I...
Comparison of Anteroposterior Vs. Axial X-rays for Assessing Bone Loss in Recurrent Shoulder Dislocation
Comparison of Anteroposterior Vs. Axial X-rays for Assessing Bone Loss in Recurrent Shoulder Dislocation
Background: Recurrent shoulder dislocation frequently causes the glenoid and humeral head to gradually lose bone, which decreases joint stability and raises the risk of another dis...
A reference facility for the comparison of CO2 in Air isotope ratio standards
A reference facility for the comparison of CO2 in Air isotope ratio standards
The commercial development of laser-based instruments over the last decade that can measure real-time isotopic ratio variations of greenhouse gases, and notably CO2, has allowed th...