ANALYTICAL AND CLINICAL EVALUATION OF SYSMEX UF1000I FOR AUTOMATED SCREENING OF CEREBROSPINAL FLUIDS ANALITIČKA I KLINIČKA EVALUACIJA UREĐAJA SYSMEX UF1000I ZA AUTOMATSKI SKRINING CEREBROSPINALNIH TEČNOSTI

View through CrossRef
Summary Background: We evaluated the performance of Sysmex UF- 1000i for cell counting and differential cell count, as well as for assessment of bacteria load in cerebrospinal fluid (CSF), as a potential approach for the rapid screening of meningitis or bacterial encephalitis. Methods: We analyzed 77 consecutive CSF samples, 34 of which (44%) displayed leukocyte count >5 white blood cell (WBQ/^iL with optical microscopy. Results on the UF-1000i were compared with those obtained by microscopic analysis. Imprecision was evaluated by testing three CSF samples with leukocyte values between 3.5 and 28.8 WBC/nL in 10 repli- cates. Carry-over was evaluated with the Broughton formula on three CSF pools with leukocyte counts between 93.5 and 132.5 WBC/^L. Linearity was assessed according to CLSI document EP6-A. In the presence of bacteria, identification and antibiogram were performed with Vitex (Biomerieux), except for Neisserie meningitidis (ApiNH, Biomerieux). Sensitivity tests were performed with Vitex and disc diffusion. Results: Optimal correlation was found between UF-1000i and optical microscopy, displaying Pearson's correlation of 0.99 and mean bias of-3.5 WBC/^L (95% Cl, from -7.0 to 0.0 WBC/nL). Imprecision varied between 12 and 16%. Li- nearity was excellent, 4-278 WBC/nL. Carry-over was neg- ligible. ROC analysis yielded AUC of 0.99 for both WBC and bacterial counts. The agreement at threshold >4 WBC/nL was 0.91, with sensitivity and specificity of 1.00 and 0.84. At S19 bacteria^nL cut-off, accuracy was 0.98, sensitivity 1.00 and specificity 0.97. Conclusions: According to these results, CSF screening with UF-I000i seems a reliable approach in terms of instrument performance, turnaround time and overall laboratory effi- ciency.
Title: ANALYTICAL AND CLINICAL EVALUATION OF SYSMEX UF1000I FOR AUTOMATED SCREENING OF CEREBROSPINAL FLUIDS ANALITIČKA I KLINIČKA EVALUACIJA UREĐAJA SYSMEX UF1000I ZA AUTOMATSKI SKRINING CEREBROSPINALNIH TEČNOSTI
Description:
Summary Background: We evaluated the performance of Sysmex UF- 1000i for cell counting and differential cell count, as well as for assessment of bacteria load in cerebrospinal fluid (CSF), as a potential approach for the rapid screening of meningitis or bacterial encephalitis.
Methods: We analyzed 77 consecutive CSF samples, 34 of which (44%) displayed leukocyte count >5 white blood cell (WBQ/^iL with optical microscopy.
Results on the UF-1000i were compared with those obtained by microscopic analysis.
Imprecision was evaluated by testing three CSF samples with leukocyte values between 3.
5 and 28.
8 WBC/nL in 10 repli- cates.
Carry-over was evaluated with the Broughton formula on three CSF pools with leukocyte counts between 93.
5 and 132.
5 WBC/^L.
Linearity was assessed according to CLSI document EP6-A.
In the presence of bacteria, identification and antibiogram were performed with Vitex (Biomerieux), except for Neisserie meningitidis (ApiNH, Biomerieux).
Sensitivity tests were performed with Vitex and disc diffusion.
Results: Optimal correlation was found between UF-1000i and optical microscopy, displaying Pearson's correlation of 0.
99 and mean bias of-3.
5 WBC/^L (95% Cl, from -7.
0 to 0.
0 WBC/nL).
Imprecision varied between 12 and 16%.
Li- nearity was excellent, 4-278 WBC/nL.
Carry-over was neg- ligible.
ROC analysis yielded AUC of 0.
99 for both WBC and bacterial counts.
The agreement at threshold >4 WBC/nL was 0.
91, with sensitivity and specificity of 1.
00 and 0.
84.
At S19 bacteria^nL cut-off, accuracy was 0.
98, sensitivity 1.
00 and specificity 0.
97.
Conclusions: According to these results, CSF screening with UF-I000i seems a reliable approach in terms of instrument performance, turnaround time and overall laboratory effi- ciency.

Related Results

Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF‐1000i
Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF‐1000i
AbstractBackgroundWe evaluated the new body fluid module on Sysmex UF1000‐i (UF1000i‐BF) for analysis of white blood cell (WBC) and red blood cell (RBC) in cerebrospinal fluid (CSF...
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Abstract The rapid growth of open access publishing (OAP) has significantly improved the accessibility and dissemination of scientific knowledge. However, this expansion has also c...
PROGRAM PENGEMBANGAN DESA MITRA PADA KADER KIA DAN BKR DALAM UPAYA INTEGRASI SKRINING CATIN
PROGRAM PENGEMBANGAN DESA MITRA PADA KADER KIA DAN BKR DALAM UPAYA INTEGRASI SKRINING CATIN
Abstrak: Skrining prakonsepsi sebagai persiapan dan perencanaan sebelum kehamilan terjadi sehingga saat kehamilan terjadi kondisi pasangan telah siap secara fisik, mental, sosial, ...
CONTEXTUAL TEACHING & LEARNING (CTL) UNTUK MENINGKATKAN PENGETAHUAN DALAM SCREENING KESEHATAN
CONTEXTUAL TEACHING & LEARNING (CTL) UNTUK MENINGKATKAN PENGETAHUAN DALAM SCREENING KESEHATAN
Salah satu peran tenaga kesehatan adalah transformasi layanan primer dengan memperkuat aktivitas promotif preventif untuk dapat membuat masyarakat sehat, meningkatkan kapasitas lay...

Back to Top