Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF‐1000i
View through CrossRef
AbstractBackgroundWe evaluated the new body fluid module on Sysmex UF1000‐i (UF1000i‐BF) for analysis of white blood cell (WBC) and red blood cell (RBC) in cerebrospinal fluid (CSF).MethodsWBC and RBC counting were compared between UF1000i‐BF and Fuchs–Rosenthal counting chamber in 67 CSF samples. This study also included the evaluation of between‐day precision, limit of blank (LoB), limit of detection (LoD), functional sensitivity (limit of quantitation, LoQ), carryover and linearity. Diagnostic agreement for differentiation between normal and increased WBC counts (≥5.0 × 106/L) was also assessed.ResultsThe agreement between UF1000i‐BF and manual WBC counts was otpiaml in all CSF samples (r = 0.99; y = 1.05x + 0.09). A modest overestimation was noticed in samples with WBC < 30 × 106/L (r = 0.95; y = 1.21x − 0.15). A good agreement was observed for RBC counts (r = 0.98; y = 1.15x + 0.55), particularly in samples with RBC ≥ 18 × 106/L (r = 0.98; y = 1.01x + 8.90). Between‐day precision was good, with coefficient of variations (CVs) lower than 7.2% for both WBC and RBC. The LoBs were 0.1 × 106 WBC/L and 1.2 × 106 RBC/L, the LoDs were 0.7 × 106 WBC/L and 5.5 × 106 RBC/L, the LoQs were 2.4 × 106 WBC/L and 18.0 × 106 RBC/L, respectively. Linearity was excellent (r = 1.00 for both WBC and RBC). Carryover was negligible. Excellent diagnostic agreement was obtained at 4.5 × 106 WBC/L cut‐off (sensitivity, 100%; specificity, 97.4%).ConclusionThe UF1000i‐BF provides rapid and accurate WBC and RBC counts in clinically relevant values of CSF cells. The use of UF1000i‐BF may hence allow to replace routine optical counting, except for samples displaying abnormal WBC counts or abnormal scattergram distribution, for which differential cell counts may still be required.
Title: Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF‐1000i
Description:
AbstractBackgroundWe evaluated the new body fluid module on Sysmex UF1000‐i (UF1000i‐BF) for analysis of white blood cell (WBC) and red blood cell (RBC) in cerebrospinal fluid (CSF).
MethodsWBC and RBC counting were compared between UF1000i‐BF and Fuchs–Rosenthal counting chamber in 67 CSF samples.
This study also included the evaluation of between‐day precision, limit of blank (LoB), limit of detection (LoD), functional sensitivity (limit of quantitation, LoQ), carryover and linearity.
Diagnostic agreement for differentiation between normal and increased WBC counts (≥5.
0 × 106/L) was also assessed.
ResultsThe agreement between UF1000i‐BF and manual WBC counts was otpiaml in all CSF samples (r = 0.
99; y = 1.
05x + 0.
09).
A modest overestimation was noticed in samples with WBC < 30 × 106/L (r = 0.
95; y = 1.
21x − 0.
15).
A good agreement was observed for RBC counts (r = 0.
98; y = 1.
15x + 0.
55), particularly in samples with RBC ≥ 18 × 106/L (r = 0.
98; y = 1.
01x + 8.
90).
Between‐day precision was good, with coefficient of variations (CVs) lower than 7.
2% for both WBC and RBC.
The LoBs were 0.
1 × 106 WBC/L and 1.
2 × 106 RBC/L, the LoDs were 0.
7 × 106 WBC/L and 5.
5 × 106 RBC/L, the LoQs were 2.
4 × 106 WBC/L and 18.
0 × 106 RBC/L, respectively.
Linearity was excellent (r = 1.
00 for both WBC and RBC).
Carryover was negligible.
Excellent diagnostic agreement was obtained at 4.
5 × 106 WBC/L cut‐off (sensitivity, 100%; specificity, 97.
4%).
ConclusionThe UF1000i‐BF provides rapid and accurate WBC and RBC counts in clinically relevant values of CSF cells.
The use of UF1000i‐BF may hence allow to replace routine optical counting, except for samples displaying abnormal WBC counts or abnormal scattergram distribution, for which differential cell counts may still be required.
Related Results
Tijelo u opusu Janka Polića Kamova
Tijelo u opusu Janka Polića Kamova
The doctoral disertation is dedicated to the concept of the body in the works of Janko Polić Kamov. The body is approached as a signifier system on the basis of which numerous and ...
Canine Cerebrospinal Fluid Analysis Using Two New Automated Techniques: The Sysmex XN-V Body Fluid Mode and an Artificial-Intelligence-Based Algorithm
Canine Cerebrospinal Fluid Analysis Using Two New Automated Techniques: The Sysmex XN-V Body Fluid Mode and an Artificial-Intelligence-Based Algorithm
Cerebrospinal fluid analysis is an important diagnostic test when assessing a neurological canine patient. For this analysis, the total nucleated cell count and differential cell c...
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Objective: To determine the frequency of common chromosomal aberrations in local population idiopathic determine the frequency of common chromosomal aberrations in local population...
Diagnostic performance of Sysmex XN-31 automated hematology analyzer compared to microscopy and PCR for detecting and quantifying malaria parasites in clinical settings of Oromia Regional State, Ethiopia
Diagnostic performance of Sysmex XN-31 automated hematology analyzer compared to microscopy and PCR for detecting and quantifying malaria parasites in clinical settings of Oromia Regional State, Ethiopia
Abstract
Background
Malaria remains a significant public health challenge in Ethiopia. To achieve malaria free country with improved clinical care and support, inn...
ANALYTICAL AND CLINICAL EVALUATION OF SYSMEX UF1000I FOR AUTOMATED SCREENING OF CEREBROSPINAL FLUIDS ANALITIČKA I KLINIČKA EVALUACIJA UREĐAJA SYSMEX UF1000I ZA AUTOMATSKI SKRINING CEREBROSPINALNIH TEČNOSTI
ANALYTICAL AND CLINICAL EVALUATION OF SYSMEX UF1000I FOR AUTOMATED SCREENING OF CEREBROSPINAL FLUIDS ANALITIČKA I KLINIČKA EVALUACIJA UREĐAJA SYSMEX UF1000I ZA AUTOMATSKI SKRINING CEREBROSPINALNIH TEČNOSTI
Summary
Background: We evaluated the performance of Sysmex UF- 1000i for cell counting and differential cell count, as well as for assessment of bacteria load in cerebrospinal...
Complex Collision Tumors: A Systematic Review
Complex Collision Tumors: A Systematic Review
Abstract
Introduction: A collision tumor consists of two distinct neoplastic components located within the same organ, separated by stromal tissue, without histological intermixing...
[RETRACTED] Prima Weight Loss Dragons Den UK v1
[RETRACTED] Prima Weight Loss Dragons Den UK v1
[RETRACTED]Prima Weight Loss Dragons Den UK :-Obesity is a not kidding medical issue brought about by devouring an excessive amount of fat, eating terrible food sources, and practi...
[RETRACTED] Prima Weight Loss Dragons Den UK v1
[RETRACTED] Prima Weight Loss Dragons Den UK v1
[RETRACTED]Prima Weight Loss Dragons Den UK :-Obesity is a not kidding medical issue brought about by devouring an excessive amount of fat, eating terrible food sources, and practi...