Diagnostic performance of Sysmex XN-31 automated hematology analyzer compared to microscopy and PCR for detecting and quantifying malaria parasites in clinical settings of Oromia Regional State, Ethiopia
View through CrossRef
Abstract
Background
Malaria remains a significant public health challenge in Ethiopia. To achieve malaria free country with improved clinical care and support, innovative diagnostic tools are needed. The Sysmex XN-31 automated malaria analyzer, which detects malaria-infected red blood cells was assessed in clinical settings in Ethiopia. This study compared the performance of Sysmex XN-31 automated analyzer with microscopy and polymerase chain reaction (PCR) tests.
Methods
We conducted a facility-based cross-sectional study in two public hospitals from April to September 2024. Febrile patients suspected of malaria who visited the outpatient departments at Shashamene Comprehensive Specialized Hospital and Meki Primary Hospital were enrolled with consent. Finger prick and four milliliters of venous blood were collected in EDTA tubes from each study participant. The diagnostic performance of the Sysmex XN-31 hematology analyzer for malaria parasite detection and quantification was compared to expert microscopist and qPCR. We also evaluated the usability of the Sysmex XN-31 analyzer in near-patient settings at both sites. Statistical analysis was performed using SPSS 27 and Medical Statistical Calculator v23. Sensitivity, specificity, predictive values, and kappa statistics were calculated.
Results
Of the 400 suspected febrile cases, malaria positivity was 22.2% (n = 89), 21% (n = 84), 19.2% (n = 77), 18.8% (n = 75), and 16.2% (n = 65) by PCR, Sysmex XN-31, expert microscopist, field microscopist, and RDT, respectively. The sensitivity and specificity of Sysmex XN-31 were 93.5% and 96.6%, respectively, compared with the expert microscopist. However, with PCR as the reference, Sysmex XN-31’s sensitivity dropped to 84.3%, with a slight increase in specificity to 97%. The Sysmex-XN-31 exhibited high negative predictive value (NPV) reliability when compared with field and rapid diagnostic test (RDT) results, based on hypothetical community malaria prevalence. We found strong agreement (89.9%) between Sysmex XN-31 and the field microscopist, with a kappa value of 0.68. The usability test of the Sysmex XN-31 by most participants indicated that errors could be resolved easily, interruptions were manageable, and the interpretation of results was straightforward.
Conclusions
The Sysmex XN-31 automated hematology analyzer offers high sensitivity and specificity compared to expert microscopist and PCR methods, significantly enhancing malaria surveillance and elimination efforts. It can detect and count malaria parasites while also providing a CBC. With a rapid turnaround time of 55 samples in one hour, it is much faster than traditional testing methods.
Springer Science and Business Media LLC
Title: Diagnostic performance of Sysmex XN-31 automated hematology analyzer compared to microscopy and PCR for detecting and quantifying malaria parasites in clinical settings of Oromia Regional State, Ethiopia
Description:
Abstract
Background
Malaria remains a significant public health challenge in Ethiopia.
To achieve malaria free country with improved clinical care and support, innovative diagnostic tools are needed.
The Sysmex XN-31 automated malaria analyzer, which detects malaria-infected red blood cells was assessed in clinical settings in Ethiopia.
This study compared the performance of Sysmex XN-31 automated analyzer with microscopy and polymerase chain reaction (PCR) tests.
Methods
We conducted a facility-based cross-sectional study in two public hospitals from April to September 2024.
Febrile patients suspected of malaria who visited the outpatient departments at Shashamene Comprehensive Specialized Hospital and Meki Primary Hospital were enrolled with consent.
Finger prick and four milliliters of venous blood were collected in EDTA tubes from each study participant.
The diagnostic performance of the Sysmex XN-31 hematology analyzer for malaria parasite detection and quantification was compared to expert microscopist and qPCR.
We also evaluated the usability of the Sysmex XN-31 analyzer in near-patient settings at both sites.
Statistical analysis was performed using SPSS 27 and Medical Statistical Calculator v23.
Sensitivity, specificity, predictive values, and kappa statistics were calculated.
Results
Of the 400 suspected febrile cases, malaria positivity was 22.
2% (n = 89), 21% (n = 84), 19.
2% (n = 77), 18.
8% (n = 75), and 16.
2% (n = 65) by PCR, Sysmex XN-31, expert microscopist, field microscopist, and RDT, respectively.
The sensitivity and specificity of Sysmex XN-31 were 93.
5% and 96.
6%, respectively, compared with the expert microscopist.
However, with PCR as the reference, Sysmex XN-31’s sensitivity dropped to 84.
3%, with a slight increase in specificity to 97%.
The Sysmex-XN-31 exhibited high negative predictive value (NPV) reliability when compared with field and rapid diagnostic test (RDT) results, based on hypothetical community malaria prevalence.
We found strong agreement (89.
9%) between Sysmex XN-31 and the field microscopist, with a kappa value of 0.
68.
The usability test of the Sysmex XN-31 by most participants indicated that errors could be resolved easily, interruptions were manageable, and the interpretation of results was straightforward.
Conclusions
The Sysmex XN-31 automated hematology analyzer offers high sensitivity and specificity compared to expert microscopist and PCR methods, significantly enhancing malaria surveillance and elimination efforts.
It can detect and count malaria parasites while also providing a CBC.
With a rapid turnaround time of 55 samples in one hour, it is much faster than traditional testing methods.
Related Results
The Diagnostic Performance of a Sysmex XN‐31 Automated Malaria Analyzer vs. Expert Microscopy
The Diagnostic Performance of a Sysmex XN‐31 Automated Malaria Analyzer vs. Expert Microscopy
ABSTRACTIntroductionMalaria is a common and life‐threatening infection. Malaria diagnosis needs to be fast and reliable. Although malaria microscopy is currently the gold standard,...
Impact And Acceptability of Seasonal Malaria Chemoprevention (SMC) And Pharmacovigilance Campaign Among Under-Five Children In Nigeria: An Explanatory Study
Impact And Acceptability of Seasonal Malaria Chemoprevention (SMC) And Pharmacovigilance Campaign Among Under-Five Children In Nigeria: An Explanatory Study
Globally, an estimated 249 million malaria cases occurred in 2022, leading to 608,000 malaria deaths in a single year. Malaria is one of the most severe public health problems, wit...
Shifts in Indonesia’s malaria landscape: an analysis of 2010-2019 routine surveillance data
Shifts in Indonesia’s malaria landscape: an analysis of 2010-2019 routine surveillance data
Abstract
Background
Indonesia faces challenges in achieving its goal of eliminating malaria by 2030, with cases stagnating betw...
JIT 2023 - Jornadas de Jóvenes Investigadores Tecnológicos
JIT 2023 - Jornadas de Jóvenes Investigadores Tecnológicos
Es un honor presentar este libro que compila los trabajos de investigación y desarrollo presentados en las Jornadas de Jóvenes Investigadores Tecnológicos (JIT) 2023. Este evento s...
Malariology (A Continuing Education Activity)
Malariology (A Continuing Education Activity)
Malariology is the scientific study of Malaria. Malaria is a life-threatening disease caused by parasites that are transmitted to people through the bites of infected female Anophe...
Evaluating performance of multiplex real time PCR for the diagnosis of malaria at elimination targeted low transmission settings of Ethiopia
Evaluating performance of multiplex real time PCR for the diagnosis of malaria at elimination targeted low transmission settings of Ethiopia
Abstract
Background
Malaria incidence has declined in Ethiopia in the past 10 years. Current malaria diagnostic tests, in...
Evaluating Performance of Multiplex Real Time PCR for the Diagnosis of Malaria at Elimination Targeted Low Transmission Settings of Ethiopia.
Evaluating Performance of Multiplex Real Time PCR for the Diagnosis of Malaria at Elimination Targeted Low Transmission Settings of Ethiopia.
Abstract
Background: Malaria incidence has declined in Ethiopia in the past ten years. Current malaria diagnostic tests, including light microscopy and antigen-detecting ra...
Malaria Incidence and Associated Risk Factors in and Around Anger Gute Town, Western Ethiopia
Malaria Incidence and Associated Risk Factors in and Around Anger Gute Town, Western Ethiopia
Abstract
Background: Malaria is one of the most important causes of morbidity and mortality in sub-Saharan Africa including Ethiopia. About 75% of the total area of the cou...