Results of the recalibration of creatinine measurement with the modular Beckman Coulter® Jaffe creatinine method
View through CrossRef
AbstractSerum creatinine is important for detecting the beginning of a decline in kidney function. The Beckman Coulter Jaffe reagents for measuring creatinine have been standardized to the internationally accepted reference method: isotope dilution mass spectrometry (IDMS). The impact of this recalibration on the Beckman CoulterRecalibrated Jaffe (calibrator set points IDMS traceable) and classic National Institute of Standards and Technology (NIST) creatinine methods (traceable to NIST 914a) were compared with an enzymatic method (Sentinel, traceable to NIST SRM 967). All measurements were performed on Synchron DxC 800 systems. Imprecision of the routine methods was studied using the Clinical and Laboratory Standards Institute (CLSI) protocols and laboratory quality survey. Thirteen plasma pools, with concentrations <354 μmol/L, were analyzed with a gas chromatography isotope dilution mass spectrometry (GC-IDMS) method and routine methods. Total error of 8.2% based on biological variability, set on the GC-IDMS values and acceptance criteria (bias <5%, imprecision <8% at concentrations ≥88.4 μmol/L and a maximum 10% increase in the relative error of estimated glomerular filtration rate (eGFR) of the National Kidney Disease Educational Program (NKDEP) were used for evaluating analytical performance of the routine methods studied.After recalibration of the Jaffe method, median bias (μmol/L) decreased from 12.4 (95% CI: 9.1–21.2) to 7.3 (95% CI: 1.5–10.5). Imprecision of the Jaffe method is in agreement with the claim of the manufacturer, namely <9 μmol/L or <3% below or above 292 μmol/L. Below creatinine of 88.4 μmol/L, imprecision of the recalibrated Jaffe and enzymatic methods is between 4.1% and 6.9%, and 5.0% and 7.1%, respectively, and significantly different (p=0.02 for both the Jaffe and enzymatic methods) from the goal, based on biological variability, of 2.7%. For the adult pools, all recalibrated Jaffe and enzymatic results fall within the total error of 8.2%. In the pediatric samples, one-third of the recalibrated Jaffe and three of the six enzymatic results fall within this total error goal.Recalibration significantly reduced bias of the Jaffe method. For pediatric samples, recalibrated Jaffe results do not comply with either the imprecision goal or the total error based on biological variability. Adult recalibrated Jaffe results are in compliance with the goals based on biological variability and with the acceptance criteria from the NKDEP.
Title: Results of the recalibration of creatinine measurement with the modular Beckman Coulter® Jaffe creatinine method
Description:
AbstractSerum creatinine is important for detecting the beginning of a decline in kidney function.
The Beckman Coulter Jaffe reagents for measuring creatinine have been standardized to the internationally accepted reference method: isotope dilution mass spectrometry (IDMS).
The impact of this recalibration on the Beckman CoulterRecalibrated Jaffe (calibrator set points IDMS traceable) and classic National Institute of Standards and Technology (NIST) creatinine methods (traceable to NIST 914a) were compared with an enzymatic method (Sentinel, traceable to NIST SRM 967).
All measurements were performed on Synchron DxC 800 systems.
Imprecision of the routine methods was studied using the Clinical and Laboratory Standards Institute (CLSI) protocols and laboratory quality survey.
Thirteen plasma pools, with concentrations <354 μmol/L, were analyzed with a gas chromatography isotope dilution mass spectrometry (GC-IDMS) method and routine methods.
Total error of 8.
2% based on biological variability, set on the GC-IDMS values and acceptance criteria (bias <5%, imprecision <8% at concentrations ≥88.
4 μmol/L and a maximum 10% increase in the relative error of estimated glomerular filtration rate (eGFR) of the National Kidney Disease Educational Program (NKDEP) were used for evaluating analytical performance of the routine methods studied.
After recalibration of the Jaffe method, median bias (μmol/L) decreased from 12.
4 (95% CI: 9.
1–21.
2) to 7.
3 (95% CI: 1.
5–10.
5).
Imprecision of the Jaffe method is in agreement with the claim of the manufacturer, namely <9 μmol/L or <3% below or above 292 μmol/L.
Below creatinine of 88.
4 μmol/L, imprecision of the recalibrated Jaffe and enzymatic methods is between 4.
1% and 6.
9%, and 5.
0% and 7.
1%, respectively, and significantly different (p=0.
02 for both the Jaffe and enzymatic methods) from the goal, based on biological variability, of 2.
7%.
For the adult pools, all recalibrated Jaffe and enzymatic results fall within the total error of 8.
2%.
In the pediatric samples, one-third of the recalibrated Jaffe and three of the six enzymatic results fall within this total error goal.
Recalibration significantly reduced bias of the Jaffe method.
For pediatric samples, recalibrated Jaffe results do not comply with either the imprecision goal or the total error based on biological variability.
Adult recalibrated Jaffe results are in compliance with the goals based on biological variability and with the acceptance criteria from the NKDEP.
Related Results
520. Development of Cefepime-taniborbactam MIC Antimicrobial Susceptibility Test (AST) for Enterobacterales and Pseudomonas aeruginosa on MicroScan Dried Gram-negative MIC Panels
520. Development of Cefepime-taniborbactam MIC Antimicrobial Susceptibility Test (AST) for Enterobacterales and Pseudomonas aeruginosa on MicroScan Dried Gram-negative MIC Panels
Abstract
Background
Automated AST devices are critical to inform appropriate care of patients with bacterial infections. Cefepim...
Coulter Method for the detection of Physiological parameters v1
Coulter Method for the detection of Physiological parameters v1
Instrument A Coulter counter [1] [2] is an apparatus for counting and sizing particles suspended in electrolytes. It is used for cells, bacteria, prokaryotic cells and virus parti...
Measurement of serum creatinine levels with liquid chromatography-tandem mass spectrometry: comparison with Jaffe and enzymatic methods
Measurement of serum creatinine levels with liquid chromatography-tandem mass spectrometry: comparison with Jaffe and enzymatic methods
Abstract
Objectives
Our aim was to validate a mass spectrometric creatinine method and compare this method with Jaffe and...
Two Suspected Cases of Immunoglobulin-Mediated Interference Causing Falsely Low Vancomycin Concentrations with the Beckman PETINIA Method
Two Suspected Cases of Immunoglobulin-Mediated Interference Causing Falsely Low Vancomycin Concentrations with the Beckman PETINIA Method
OBJECTIVE
To describe 2 recent cases of suspected immunoglobulin-mediated interference with the Beckman Coulter particle–enhanced turbidimetric inhibition immun...
S-JAFFE = JAFFE: A Note on Dataset Rebranding and Misattribution
S-JAFFE = JAFFE: A Note on Dataset Rebranding and Misattribution
This note clarifies that the entity referred to in parts of the label-distribution-learning (LDL) literature as S-JAFFE is not distinct from JAFFE, the Japanese Female Facial Expre...
Where you aim – not how you aim – affects implicit recalibration in visuomotor adaptation
Where you aim – not how you aim – affects implicit recalibration in visuomotor adaptation
Abstract
The influence of explicit strategies on implicit recalibration during visuomotor adaptation has become a central question in motor learning. Because the tw...
Evaluating kidney function using a point-of-care creatinine test in Ugandan children with severe malaria: a prospective cohort study
Evaluating kidney function using a point-of-care creatinine test in Ugandan children with severe malaria: a prospective cohort study
Abstract
Background
Acute kidney injury (AKI) disproportionately affects individuals in low-and middle-income countries (LMIC). However, LMIC—partic...
Determination of Reticulocytes: Three Methods Compared
Determination of Reticulocytes: Three Methods Compared
Abstract
Determination of reticulocytes in peripheral blood is a valuable tool for getting information about erythropoiesis of an individual. For many years, reticul...