Measurement of serum creatinine levels with liquid chromatography-tandem mass spectrometry: comparison with Jaffe and enzymatic methods
View through CrossRef
Abstract
Objectives
Our aim was to validate a mass spectrometric creatinine method and compare this method with Jaffe and enzymatic serum creatinine methods.
Methods
90 samples were included. The levels were classified into three groups according to serum creatinine results as Group 1: Lower (n=30) (0.16–0.59 mg/dL), Group 2: Normal (n=30) (0.62–1.18 mg/dL) and Group 3: Higher (n=30) (1.33–3.88 mg/dL). Jaffe and enzymatic creatinine measurements were performed on the Beckman Coulter AU5800 autoanalyzer.
Results
Serum creatinine was linear from 0.039 up to 10 mg/dL, CV and bias values were ranged between 1.9–3.8% and 2–15%. Correlation coefficients were 0.990 (95% confidence interval 0.984–0.993), 0.992 (95% confidence interval 0.988–0.995) and 0.994 (95% confidence interval 0.991–0.996) for LC-MS/MS-Enzymatic, LC-MS/MS-Jaffe and Enzymatic-Jaffe, respectively.
Conclusions
Although, Jaffe method for serum creatinine measurement is still much more practical and cheap, so in use for routine practice, tandem mass spectrometric detection of serum creatinine can be used as an accurate and specific method for verification of discordant clinical results, existence of possible interferences and serum levels under 0.5 mg/dL creatinine results such as pediatric or pregnant populations.
Title: Measurement of serum creatinine levels with liquid chromatography-tandem mass spectrometry: comparison with Jaffe and enzymatic methods
Description:
Abstract
Objectives
Our aim was to validate a mass spectrometric creatinine method and compare this method with Jaffe and enzymatic serum creatinine methods.
Methods
90 samples were included.
The levels were classified into three groups according to serum creatinine results as Group 1: Lower (n=30) (0.
16–0.
59 mg/dL), Group 2: Normal (n=30) (0.
62–1.
18 mg/dL) and Group 3: Higher (n=30) (1.
33–3.
88 mg/dL).
Jaffe and enzymatic creatinine measurements were performed on the Beckman Coulter AU5800 autoanalyzer.
Results
Serum creatinine was linear from 0.
039 up to 10 mg/dL, CV and bias values were ranged between 1.
9–3.
8% and 2–15%.
Correlation coefficients were 0.
990 (95% confidence interval 0.
984–0.
993), 0.
992 (95% confidence interval 0.
988–0.
995) and 0.
994 (95% confidence interval 0.
991–0.
996) for LC-MS/MS-Enzymatic, LC-MS/MS-Jaffe and Enzymatic-Jaffe, respectively.
Conclusions
Although, Jaffe method for serum creatinine measurement is still much more practical and cheap, so in use for routine practice, tandem mass spectrometric detection of serum creatinine can be used as an accurate and specific method for verification of discordant clinical results, existence of possible interferences and serum levels under 0.
5 mg/dL creatinine results such as pediatric or pregnant populations.
Related Results
7
th
International Symposium on Enabling Technologies for Life Sciences (ETP)
7
th
International Symposium on Enabling Technologies for Life Sciences (ETP)
The seventh in the series of ETP Symposia (see
Rapid Communications in Mass Spectrometry
2012,
26
, ...
Results of the recalibration of creatinine measurement with the modular Beckman Coulter® Jaffe creatinine method
Results of the recalibration of creatinine measurement with the modular Beckman Coulter® Jaffe creatinine method
AbstractSerum creatinine is important for detecting the beginning of a decline in kidney function. The Beckman Coulter Jaffe reagents for measuring creatinine have been standardize...
Simultaneous Determination of Cortisol and Testosterone from Rat Serum by Liquid Chromatography-Tandem Mass Spectrometry
Simultaneous Determination of Cortisol and Testosterone from Rat Serum by Liquid Chromatography-Tandem Mass Spectrometry
A liquid chromatography-electrospray ionization tandem mass spectrometry in a positive mode (LC-ESI-MS/MS) was developed to determine simultaneously the levels of cortisol and test...
Hubungan kontrol glikemik dengan petanda gangguan ginjal dini pada pasien diabetes melitus tipe 2
Hubungan kontrol glikemik dengan petanda gangguan ginjal dini pada pasien diabetes melitus tipe 2
Background: Glycemic control in type 2 diabetes mellitus (DM) which can be assessed by examination of fasting blood sugar (FBG) and glycosylated hemoglobin (HbA1c) is thought to be...
Emerging Evidence of IgG4-Related Disease in Pericarditis: A Systematic Review
Emerging Evidence of IgG4-Related Disease in Pericarditis: A Systematic Review
Abstract
Introduction
Immunoglobulin G4-related disease (IgG4-RD) is a recently identified immune-mediated condition that is debilitating and often overlooked. While IgG4-RD has be...
Chemical extraction of sulfachloropyridazine from feather samples v1
Chemical extraction of sulfachloropyridazine from feather samples v1
Analytical methodology for the detection of sulfachloropyridazine (SCP) in samples of feathersvia LC-MS/MS was implemented based on techniques previously published by other authors...
Chemical extraction of sulfachloropyridazine from muscle samples v1
Chemical extraction of sulfachloropyridazine from muscle samples v1
Analytical methodology for the detection of sulfachloropyridazine (SCP) in samples of feathersvia LC-MS/MS was implemented based on techniques previously published by other authors...
Chemical extraction of sulfachloropyridazine from liver samples v1
Chemical extraction of sulfachloropyridazine from liver samples v1
Analytical methodology for the detection of sulfachloropyridazine (SCP) in samples of feathersvia LC-MS/MS was implemented based on techniques previously published by other authors...