Corneal Biomechanics in Normal and Subclinical Keratoconus Eyes
View through CrossRef
Abstract
Background
The diagnosis of keratoconus, as the most prevalent corneal ectatic disorder, at the subclinical stage gained great attention due to the increased acceptance of refractive surgeries. This study aimed to assess the pattern of the corneal biomechanical properties derived from Corneal Visualization Scheimpflug Technology (Corvis ST) and evaluate the diagnostic value of these parameters in distinguishing subclinical keratoconus (SKC) from normal eyes.
Methods
This prospective study was conducted on 73 SKC and 69 normal eyes. Subclinical keratoconus eyes were defined as corneas with no clinical evidence of keratoconus and suspicious tomographic and tomographic features. Following a complete ophthalmic examination, topographic and tomographic corneal assessment via Pentacam HR, and corneal biomechanical evaluation utilizing Corvis ST were done.
Results
Subclinical keratoconus eyes presented significantly higher Deformation Amplitude (DA) ratio, Tomographic Biomechanical Index (TBI), and Corvis Biomechanical Index (CBI) rates than the control group. Conversely, Ambrósio Relational Thickness to the Horizontal profile (ARTh), and Stiffness Parameter at the first Applanation (SPA1) showed significantly lower rates in SKC eyes. In diagnosing SKC from normal eyes, TBI (AUC: 0.858, Cut-off value: > 0.33, Youden index: 0.55), ARTh (AUC: 0.813, Cut-off value: ≤ 488.1, Youden index: 0.58), and CBI (AUC: 0.804, Cut-off value: > 0.47, Youden index: 0.49) appeared as good indicators.
Conclusions
Integrated Corvis ST parameters could be valuable in distinguishing SKC eyes from normal ones.
Research Square Platform LLC
Title: Corneal Biomechanics in Normal and Subclinical Keratoconus Eyes
Description:
Abstract
Background
The diagnosis of keratoconus, as the most prevalent corneal ectatic disorder, at the subclinical stage gained great attention due to the increased acceptance of refractive surgeries.
This study aimed to assess the pattern of the corneal biomechanical properties derived from Corneal Visualization Scheimpflug Technology (Corvis ST) and evaluate the diagnostic value of these parameters in distinguishing subclinical keratoconus (SKC) from normal eyes.
Methods
This prospective study was conducted on 73 SKC and 69 normal eyes.
Subclinical keratoconus eyes were defined as corneas with no clinical evidence of keratoconus and suspicious tomographic and tomographic features.
Following a complete ophthalmic examination, topographic and tomographic corneal assessment via Pentacam HR, and corneal biomechanical evaluation utilizing Corvis ST were done.
Results
Subclinical keratoconus eyes presented significantly higher Deformation Amplitude (DA) ratio, Tomographic Biomechanical Index (TBI), and Corvis Biomechanical Index (CBI) rates than the control group.
Conversely, Ambrósio Relational Thickness to the Horizontal profile (ARTh), and Stiffness Parameter at the first Applanation (SPA1) showed significantly lower rates in SKC eyes.
In diagnosing SKC from normal eyes, TBI (AUC: 0.
858, Cut-off value: > 0.
33, Youden index: 0.
55), ARTh (AUC: 0.
813, Cut-off value: ≤ 488.
1, Youden index: 0.
58), and CBI (AUC: 0.
804, Cut-off value: > 0.
47, Youden index: 0.
49) appeared as good indicators.
Conclusions
Integrated Corvis ST parameters could be valuable in distinguishing SKC eyes from normal ones.
Related Results
Corneal densitometry analysis in patients diagnosed with keratoconus
Corneal densitometry analysis in patients diagnosed with keratoconus
AbstractPurposeThe purpose of this study is to carry out a comparative study of corneal densitometry (COD) using Pentacam HR in two groups of patients. A control group of normal ey...
Correlation of the KISA% index and Scheimpflug tomography in ‘normal’, ‘subclinical’, ‘keratoconus‐suspect’ and ‘clinically manifest’ keratoconus eyes
Correlation of the KISA% index and Scheimpflug tomography in ‘normal’, ‘subclinical’, ‘keratoconus‐suspect’ and ‘clinically manifest’ keratoconus eyes
AbstractPurposeTo analyse tomographic changes in eyes classified as ‘normal’, ‘keratoconus‐suspect’ and ‘clinically manifest keratoconus’ based on the established KISA% definition ...
Corneal Deformation Response in Patients With keratoconus and in Healthy Subjects
Corneal Deformation Response in Patients With keratoconus and in Healthy Subjects
Abstract
Background
To compare the corneal deformation response between patients with keratoconus and healthy subjects and to identify potential characteristics associated...
A predictive model for early diagnosis of keratoconus
A predictive model for early diagnosis of keratoconus
Abstract
Background: The diagnosis of keratoconus in the early stages of the disease is necessary to initiate an early treatment of keratoconus. Furthermore to avoid possib...
A predictive model for early diagnosis of keratoconus
A predictive model for early diagnosis of keratoconus
Abstract
Background: The diagnosis of keratoconus in the early stages of the disease is necessary to initiate an early treatment of keratoconus. Furthermore, to avoid possi...
Predictive model for early diagnosis of keratoconus
Predictive model for early diagnosis of keratoconus
Abstract
Background To describe the topographic, pachymetric and aberrometry characteristics in patients with keratoconus, subclinical keratoconus/forme fruste and normal c...
Assessment and Correlation of Corneal Endothelial Cell Changes in Different Stages of Keratoconus in Non–contact Lens Wearers
Assessment and Correlation of Corneal Endothelial Cell Changes in Different Stages of Keratoconus in Non–contact Lens Wearers
Supplemental digital content is available in the text.
SIGNIFICANCE
The endothelial changes in keratoc...
Topographic Keratoconus Incidence in Greece Diagnosed in Routine Consecutive Cataract Procedures: A Consecutive Case Series of 1250 Cases over 5 Years
Topographic Keratoconus Incidence in Greece Diagnosed in Routine Consecutive Cataract Procedures: A Consecutive Case Series of 1250 Cases over 5 Years
Background: Scheimpflug tomography has for many years been an integral part of our pre-operative assessment in cataract extraction. We retrospectively reviewed the incidence of top...