Time‐dependent diffusivity and kurtosis in phantoms and patients with head and neck cancer
View through CrossRef
PurposeTo assess the reliability of measuring diffusivity, diffusional kurtosis, and cellular‐interstitial water exchange time with long diffusion times (100–800 ms) using stimulated‐echo DWI.MethodsTime‐dependent diffusion MRI was tested on two well‐established diffusion phantoms and in 5 patients with head and neck cancer. Measurements were conducted using an in‐house diffusion‐weighted STEAM‐EPI pulse sequence with multiple diffusion times at a fixed TE on three scanners. We used the weighted linear least‐squares fit method to estimate time‐dependent diffusivity, , and diffusional kurtosis, . Additionally, the Kärger model was used to estimate cellular‐interstitial water exchange time () from .ResultsDiffusivity measured by time‐dependent STEAM‐EPI measurements and commercial SE‐EPI showed comparable results withR2above 0.98 and overall 5.4 ± 3.0% deviation across diffusion times. Diffusional kurtosis phantom data showed expected patterns: constant and = 0 for negative controls and slow varying and for samples made of nanoscopic vesicles. Time‐dependent diffusion MRI in patients with head and neck cancer found that the Kärger model could be considered valid in 72% ± 23% of the voxels in the metastatic lymph nodes. The median cellular‐interstitial water exchange time estimated for lesions was between 58.5 ms and 70.6 ms.ConclusionsBased on two well‐established diffusion phantoms, we found that time‐dependent diffusion MRI measurements can provide stable diffusion and kurtosis values over a wide range of diffusion times and across multiple MRI systems. Moreover, estimation of cellular‐interstitial water exchange time can be achieved using the Kärger model for the metastatic lymph nodes in patients with head and neck cancer.
Title: Time‐dependent diffusivity and kurtosis in phantoms and patients with head and neck cancer
Description:
PurposeTo assess the reliability of measuring diffusivity, diffusional kurtosis, and cellular‐interstitial water exchange time with long diffusion times (100–800 ms) using stimulated‐echo DWI.
MethodsTime‐dependent diffusion MRI was tested on two well‐established diffusion phantoms and in 5 patients with head and neck cancer.
Measurements were conducted using an in‐house diffusion‐weighted STEAM‐EPI pulse sequence with multiple diffusion times at a fixed TE on three scanners.
We used the weighted linear least‐squares fit method to estimate time‐dependent diffusivity, , and diffusional kurtosis, .
Additionally, the Kärger model was used to estimate cellular‐interstitial water exchange time () from .
ResultsDiffusivity measured by time‐dependent STEAM‐EPI measurements and commercial SE‐EPI showed comparable results withR2above 0.
98 and overall 5.
4 ± 3.
0% deviation across diffusion times.
Diffusional kurtosis phantom data showed expected patterns: constant and = 0 for negative controls and slow varying and for samples made of nanoscopic vesicles.
Time‐dependent diffusion MRI in patients with head and neck cancer found that the Kärger model could be considered valid in 72% ± 23% of the voxels in the metastatic lymph nodes.
The median cellular‐interstitial water exchange time estimated for lesions was between 58.
5 ms and 70.
6 ms.
ConclusionsBased on two well‐established diffusion phantoms, we found that time‐dependent diffusion MRI measurements can provide stable diffusion and kurtosis values over a wide range of diffusion times and across multiple MRI systems.
Moreover, estimation of cellular‐interstitial water exchange time can be achieved using the Kärger model for the metastatic lymph nodes in patients with head and neck cancer.
Related Results
SU‐FF‐I‐26: Tissue Equivalent Phantoms for the Evaluation of Tube Current Modulated CT Dose and Image Quality
SU‐FF‐I‐26: Tissue Equivalent Phantoms for the Evaluation of Tube Current Modulated CT Dose and Image Quality
Purpose: To develop and test new, flexible, tissue‐equivalent phantoms for the evaluation of tube current modulation dose reduction and image quality in CT. The developed phantom m...
Photothermal Response of Tissue Phantoms Containing Multi-Walled Carbon Nanotubes
Photothermal Response of Tissue Phantoms Containing Multi-Walled Carbon Nanotubes
Inclusion of multi-walled carbon nanotubes (MWNTs) into tissue prior to laser therapy has the potential to enhance the selectivity and effectiveness of cancer therapy by providing ...
Edoxaban and Cancer-Associated Venous Thromboembolism: A Meta-analysis of Clinical Trials
Edoxaban and Cancer-Associated Venous Thromboembolism: A Meta-analysis of Clinical Trials
Abstract
Introduction
Cancer patients face a venous thromboembolism (VTE) risk that is up to 50 times higher compared to individuals without cancer. In 2010, direct oral anticoagul...
Are Cervical Ribs Indicators of Childhood Cancer? A Narrative Review
Are Cervical Ribs Indicators of Childhood Cancer? A Narrative Review
Abstract
A cervical rib (CR), also known as a supernumerary or extra rib, is an additional rib that forms above the first rib, resulting from the overgrowth of the transverse proce...
Diffusion of small solutes in cartilage as measured by nuclear magnetic resonance (NMR) spectroscopy and imaging
Diffusion of small solutes in cartilage as measured by nuclear magnetic resonance (NMR) spectroscopy and imaging
AbstractThe ability of water and solutes to move through the cartilage matrix is important to the normal function of cartilage and is presumed to be altered in degenerative disease...
Comparative study of kurtosis and L-kurtosis for bearing fault classification in induction motors
Comparative study of kurtosis and L-kurtosis for bearing fault classification in induction motors
This study investigates the effectiveness of L-kurtosis as a robust alternative to traditional kurtosis for identifying and categorizing rolling bearing faults in vibration signals...
Differential Diagnosis of Neurogenic Thoracic Outlet Syndrome: A Review
Differential Diagnosis of Neurogenic Thoracic Outlet Syndrome: A Review
Abstract
Thoracic outlet syndrome (TOS) is a complex and often overlooked condition caused by the compression of neurovascular structures as they pass through the thoracic outlet. ...
Multicenter Repeatability Study of a Novel Quantitative Diffusion Kurtosis Imaging Phantom
Multicenter Repeatability Study of a Novel Quantitative Diffusion Kurtosis Imaging Phantom
Quantitative kurtosis phantoms are sought by multicenter clinical trials to establish accuracy and precision of quantitative imaging biomarkers on the basis of diffusion kurtosis i...