Difference in IOP (Intraocular Pressure) with and without eye speculum during examination under general anaesthesia using perkins tonometer.
View through CrossRef
Objective: To compare intraocular pressure measurements obtained with and without an eyelid speculum during examination under general anaesthesia using Perkins tonometry. Study Design: Prospective Observational study. Setting: Department of Ophthalmology, Mughal Eye Hospital, Lahore. Periods: 1st October 2025 to 1st March 2026. Methods: Included 25 children (40 eyes) aged 1–12 years undergoing EUA. IOP was measured using a Perkins handheld applanation tonometer after induction of standardized general anaesthesia. Three readings were obtained first within 5 minutes of given general anaesthesia and subsequently after insertion of a pediatric eyelid speculum. Mean IOP values were compared using a paired t-test, and correlation analysis was performed to assess associations with demographic variables. Results: The mean age of participants was 3.5 ± 2.1 years. Mean baseline IOP without speculum was 11.2 ± 3.8 mmHg, which increased to 12.8 ± 3.2 mmHg after speculum application. The mean IOP difference was 1.6± 0.6 mmHg, which was statistically significant (p < 0.001). No significant correlation was observed between age and IOP change (r ≈ 0.10, p > 0.05). A weak positive correlation was observed between baseline IOP and IOP change (r ≈ 0.30, p < 0.05). Conclusion: Eyelid speculum use during examination under general anaesthesia results in a statistically significant increase in intraocular pressure measurements using Perkins tonometry. Although the magnitude of change is modest, it may be clinically relevant in pediatric glaucoma management, where treatment decisions are often based on single IOP readings.
Independent Medical Trust
Title: Difference in IOP (Intraocular Pressure) with and without eye speculum during examination under general anaesthesia using perkins tonometer.
Description:
Objective: To compare intraocular pressure measurements obtained with and without an eyelid speculum during examination under general anaesthesia using Perkins tonometry.
Study Design: Prospective Observational study.
Setting: Department of Ophthalmology, Mughal Eye Hospital, Lahore.
Periods: 1st October 2025 to 1st March 2026.
Methods: Included 25 children (40 eyes) aged 1–12 years undergoing EUA.
IOP was measured using a Perkins handheld applanation tonometer after induction of standardized general anaesthesia.
Three readings were obtained first within 5 minutes of given general anaesthesia and subsequently after insertion of a pediatric eyelid speculum.
Mean IOP values were compared using a paired t-test, and correlation analysis was performed to assess associations with demographic variables.
Results: The mean age of participants was 3.
5 ± 2.
1 years.
Mean baseline IOP without speculum was 11.
2 ± 3.
8 mmHg, which increased to 12.
8 ± 3.
2 mmHg after speculum application.
The mean IOP difference was 1.
6± 0.
6 mmHg, which was statistically significant (p < 0.
001).
No significant correlation was observed between age and IOP change (r ≈ 0.
10, p > 0.
05).
A weak positive correlation was observed between baseline IOP and IOP change (r ≈ 0.
30, p < 0.
05).
Conclusion: Eyelid speculum use during examination under general anaesthesia results in a statistically significant increase in intraocular pressure measurements using Perkins tonometry.
Although the magnitude of change is modest, it may be clinically relevant in pediatric glaucoma management, where treatment decisions are often based on single IOP readings.
Related Results
Implantable arterial blood pressure sensor
Implantable arterial blood pressure sensor
Although blood pressure measurements are well developed, the existing technology in the field of Arterial Blood Pressure (ABP) measurements is not sufficient in terms of long-term ...
Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer
Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer
Objective: To compare the Intraocular Pressure measured by the Goldmann Applanation tonometer with by iCare ic100 tonometer in normal human eye to determine the accuracy and reli...
Comparison of Intraocular Pressure between Air Puff Tonometer and Goldmann Applanation Tonometer at a Tertiary Eye Care Centre
Comparison of Intraocular Pressure between Air Puff Tonometer and Goldmann Applanation Tonometer at a Tertiary Eye Care Centre
Purpose: The purpose of this study was to evaluate the role of air puff (AP) tonometer by comparing the measurements of intraocular pressure (IOP) made using this device with those...
Effect of Fasting on Intraocular Pressure
Effect of Fasting on Intraocular Pressure
Purpose: To determine the effect of fasting on intraocular pressure in fasting individuals.
Study Design: Cross sectional observational study.
Place and Duration of S...
Transpalpebral tonometry after penetrating keratoplasty
Transpalpebral tonometry after penetrating keratoplasty
Aim: to assess the efficacy and safety of EASYTON transpalpebral intraocular pressure (IOP) tonometer in the early postoperative period after penetrating keratoplasty (PK).Patients...
Accuracy of Icare Rebound Tonometer and Its Comparison with Goldman Applanation Tonometer
Accuracy of Icare Rebound Tonometer and Its Comparison with Goldman Applanation Tonometer
Purpose: To determine accuracy of iCare rebound tonometer (IRT) in terms of agreement with Goldman Applanation Tonometer (GAT) and effect of Central corneal thickness (CCT) on its...
Effect of Supra Choroid Triamcinolone Acetate On Intraocular Pressure
Effect of Supra Choroid Triamcinolone Acetate On Intraocular Pressure
Intraocular pressure is a critical parameter in maintaining the physiological function of the eye. Dysregulation of Intraocular pressure can lead to ocular pathologies. Objective: ...
Features of measurement of intraocular pressure in children
Features of measurement of intraocular pressure in children
This review discusses the results of various studies conducted in recent years on the comparison of modern methods of measuring intraocular pressure (IOP) in children: pneumotonome...