Comparison of Intraocular Pressure between Air Puff Tonometer and Goldmann Applanation Tonometer at a Tertiary Eye Care Centre

View through CrossRef
Purpose: The purpose of this study was to evaluate the role of air puff (AP) tonometer by comparing the measurements of intraocular pressure (IOP) made using this device with those made using a Goldmann applanation tonometer (GAT) at Chittagong Eye Infirmary and Training Complex.  Method: An observational and comparative study was carried out at Chittagong Eye Infirmary and Training Complex from January 2016 to January 2017. Two techniques for IOP measurements using the standard GAT and the non-contact tonometer (NCT) were compared. A total of 400 eyes from 200 patients were included in the study. Results: The mean IOP as measured by GAT in the right eye was 14.50±5.59mmHg and in the left eye was 14.87±7.03mmHg while that as measured by NCT in the right eye was 15.97±6.12mmHg and in the left eye was 15.94±6.98mmHg. The mean difference between the two methods of measurement in the right eye was 1.47±0.53mmHg and the left eye was 1.07±0.05mmHg. The readings obtained by NCT were higher than those obtained by GAT. There was no statistically significant difference found in IOP measurements between GAT and NCT according to patient’s age, gender or laterality of eyes.  Conclusion: There was a significant difference in the measurement of IOP between GAT and NCT. Goldmann applanation tonometry remains the most suitable and reliable method for measuring IOP.
Title: Comparison of Intraocular Pressure between Air Puff Tonometer and Goldmann Applanation Tonometer at a Tertiary Eye Care Centre
Description:
Purpose: The purpose of this study was to evaluate the role of air puff (AP) tonometer by comparing the measurements of intraocular pressure (IOP) made using this device with those made using a Goldmann applanation tonometer (GAT) at Chittagong Eye Infirmary and Training Complex.
  Method: An observational and comparative study was carried out at Chittagong Eye Infirmary and Training Complex from January 2016 to January 2017.
Two techniques for IOP measurements using the standard GAT and the non-contact tonometer (NCT) were compared.
A total of 400 eyes from 200 patients were included in the study.
Results: The mean IOP as measured by GAT in the right eye was 14.
50±5.
59mmHg and in the left eye was 14.
87±7.
03mmHg while that as measured by NCT in the right eye was 15.
97±6.
12mmHg and in the left eye was 15.
94±6.
98mmHg.
The mean difference between the two methods of measurement in the right eye was 1.
47±0.
53mmHg and the left eye was 1.
07±0.
05mmHg.
The readings obtained by NCT were higher than those obtained by GAT.
There was no statistically significant difference found in IOP measurements between GAT and NCT according to patient’s age, gender or laterality of eyes.
  Conclusion: There was a significant difference in the measurement of IOP between GAT and NCT.
Goldmann applanation tonometry remains the most suitable and reliable method for measuring IOP.

Related Results

Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer
Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer
  Objective: To compare the Intraocular Pressure measured by the Goldmann Applanation tonometer with by iCare ic100 tonometer in normal human eye to determine the accuracy and reli...
Changes in Intraocular Pressure After Laser in situ Keratomileusis
Changes in Intraocular Pressure After Laser in situ Keratomileusis
ABSTRACT PURPOSE: To evaluate changes in intraocular pressure (IOP) measurements by Goldmann applanation tonometry after laser in situ keratomileusis (LASIK) for myopia a...
Effect of Corneal Hydration on Goldmann Applanation Tonometry and Corneal Topography
Effect of Corneal Hydration on Goldmann Applanation Tonometry and Corneal Topography
ABSTRACT BACKGROUND: Measured ocular tension is reported to be positively related to corneal thickness. It has been speculated that an edematous cornea will cause low mea...
Implantable arterial blood pressure sensor
Implantable arterial blood pressure sensor
Although blood pressure measurements are well developed, the existing technology in the field of Arterial Blood Pressure (ABP) measurements is not sufficient in terms of long-term ...
Effect of Fasting on Intraocular Pressure
Effect of Fasting on Intraocular Pressure
Purpose:  To determine the effect of fasting on intraocular pressure in fasting individuals. Study Design:  Cross sectional observational study. Place and Duration of S...
Effect of Supra Choroid Triamcinolone Acetate On Intraocular Pressure
Effect of Supra Choroid Triamcinolone Acetate On Intraocular Pressure
Intraocular pressure is a critical parameter in maintaining the physiological function of the eye. Dysregulation of Intraocular pressure can lead to ocular pathologies. Objective: ...

Back to Top