Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer

View through CrossRef
  Objective: To compare the Intraocular Pressure measured by the Goldmann Applanation tonometer with by iCare ic100 tonometer in normal human eye to determine the accuracy and reliability of both methods.  Study Design: Comparative cross-sectional study. Place and Duration of Study: Ophthalmology Department, Combined Military Hospital, Quetta Pakistan, from Oct 2022 to Apr 2023. Methodology: Approval was secured prior to enrolling every individual, protecting their privacy throughout the process. All instruments adhered to the manufacturer's guidelines for calibration. A single ophthalmologist assessed the intraocular pressure of all participants. In our study we included 400 eyes (200 right and 200 left) among 200 patients. Initially, study eyes were inspected using a slit lamp. Subsequently, iCare tonometer and Goldman applanation tonometer were employed to gauge the intraocular pressure. Results: Out of 200, 87 patients (43.5%) were male and 113(56.5%) were females. Median age of the patients was 51.00(45-58) years range from 22-77 years. Mean IOP measured by iCare was 19.01±7.86 mmHg and by Goldman applanation tonomter was 20.14±8.15 mmHg. The correlation coefficient value between the readings of two instruments was r=0.866 with significant p value (p< 0.001). Conclusion: The iCare ic100 consistently provides high intraocular pressure measurements (2-3mmHg) compared to the Goldmann Applanation tonometer across various sub Groups.
Title: Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer
Description:
  Objective: To compare the Intraocular Pressure measured by the Goldmann Applanation tonometer with by iCare ic100 tonometer in normal human eye to determine the accuracy and reliability of both methods.
  Study Design: Comparative cross-sectional study.
Place and Duration of Study: Ophthalmology Department, Combined Military Hospital, Quetta Pakistan, from Oct 2022 to Apr 2023.
Methodology: Approval was secured prior to enrolling every individual, protecting their privacy throughout the process.
All instruments adhered to the manufacturer's guidelines for calibration.
A single ophthalmologist assessed the intraocular pressure of all participants.
In our study we included 400 eyes (200 right and 200 left) among 200 patients.
Initially, study eyes were inspected using a slit lamp.
Subsequently, iCare tonometer and Goldman applanation tonometer were employed to gauge the intraocular pressure.
Results: Out of 200, 87 patients (43.
5%) were male and 113(56.
5%) were females.
Median age of the patients was 51.
00(45-58) years range from 22-77 years.
Mean IOP measured by iCare was 19.
01±7.
86 mmHg and by Goldman applanation tonomter was 20.
14±8.
15 mmHg.
The correlation coefficient value between the readings of two instruments was r=0.
866 with significant p value (p< 0.
001).
Conclusion: The iCare ic100 consistently provides high intraocular pressure measurements (2-3mmHg) compared to the Goldmann Applanation tonometer across various sub Groups.

Related Results

Accuracy of Icare Rebound Tonometer and Its Comparison with Goldman Applanation Tonometer
Accuracy of Icare Rebound Tonometer and Its Comparison with Goldman Applanation Tonometer
Purpose:  To determine accuracy of iCare rebound tonometer (IRT) in terms of agreement with Goldman Applanation Tonometer (GAT) and effect of Central corneal thickness (CCT) on its...
Implantable arterial blood pressure sensor
Implantable arterial blood pressure sensor
Although blood pressure measurements are well developed, the existing technology in the field of Arterial Blood Pressure (ABP) measurements is not sufficient in terms of long-term ...
Effect of Fasting on Intraocular Pressure
Effect of Fasting on Intraocular Pressure
Purpose:  To determine the effect of fasting on intraocular pressure in fasting individuals. Study Design:  Cross sectional observational study. Place and Duration of S...
Effect of a tight necktie on intraocular pressure
Effect of a tight necktie on intraocular pressure
Introduction. Any factor causing constriction of the neck may lead to an increase in intraocular pressure. A tight necktie may result in increasing intraocular pressure, whic...
Effect of Corneal Hydration on Goldmann Applanation Tonometry and Corneal Topography
Effect of Corneal Hydration on Goldmann Applanation Tonometry and Corneal Topography
ABSTRACT BACKGROUND: Measured ocular tension is reported to be positively related to corneal thickness. It has been speculated that an edematous cornea will cause low mea...
Changes in Intraocular Pressure After Laser in situ Keratomileusis
Changes in Intraocular Pressure After Laser in situ Keratomileusis
ABSTRACT PURPOSE: To evaluate changes in intraocular pressure (IOP) measurements by Goldmann applanation tonometry after laser in situ keratomileusis (LASIK) for myopia a...

Back to Top