On the fatigue resistance of endodontic files subjected to electrochemical polishing and an autoclave’s sterilisation cycle

View through CrossRef
Purpose The purpose of this paper is to improve the surface finish and, therefore, the fatigue resistance of some endodontic files by electrochemical polishing; the influence of applying an autoclave’s sterilising cycle to the endodontic files under study was also assessed during the research herein presented. Design/methodology/approach In total, 22 Ni-Ti endodontic instruments of two sizes, namely, ref. Hyflex 0.04/20 and 0.06/20 (taper: 0.04; 0.06 and tip diameter: 0.20 mm, respectively), were submitted to rotational bending tests at 500 rpm; the radius and the angle of curvature imposed to the endodontic instruments during the experimental tests were equal to 4.7 mm and 45°, respectively. In addition, 18 endodontic files (out of the 22) were subjected to electrochemical polishing before in vitro fatigue tests had been carried out. Several combinations of electrochemical parameters were considered during the preliminary experiments involving electrolyte, voltage, flow rate and duration. The endodontic files under study were weighted using a high-precision balance before and after polishing, and mass variation was determined. Also, a sterilising cycle was applied to six polished endodontic instruments, which were placed inside an autoclave at 134°C, for 30 min, and under a pressure of 2.16 bar. Findings The fatigue resistance of as-received endodontic files ref. 0.04/20 is much larger (+480 per cent) than the fatigue strength of endodontic files ref. 0.06/20. The weight variation due to electropolishing was comprehended between 0.53 mg and 1.01 mg, and the fatigue resistance of all polished endodontic instruments increased between 30 per cent (0.06/20) and 80 per cent (0.04/20) when compared with the fatigue resistance of the as-received files. The application of an autoclave’s sterilisation cycle had a negative impact on the fatigue resistance of instruments ref. 0.04/20 and a positive impact on the durability of instruments ref. 0.06/20. Originality/value The application of an electrochemical polishing process increased the fatigue resistance of the endodontic files under study significantly, and the values for the variables of the electrochemical process were never published elsewhere. The surface finish improvement was obtained by immersing the endodontic files in a Struers’ Electrolyte (ref. E2) for 3 s while applying 30 volts and a flow rate of 1.
Title: On the fatigue resistance of endodontic files subjected to electrochemical polishing and an autoclave’s sterilisation cycle
Description:
Purpose The purpose of this paper is to improve the surface finish and, therefore, the fatigue resistance of some endodontic files by electrochemical polishing; the influence of applying an autoclave’s sterilising cycle to the endodontic files under study was also assessed during the research herein presented.
Design/methodology/approach In total, 22 Ni-Ti endodontic instruments of two sizes, namely, ref.
Hyflex 0.
04/20 and 0.
06/20 (taper: 0.
04; 0.
06 and tip diameter: 0.
20 mm, respectively), were submitted to rotational bending tests at 500 rpm; the radius and the angle of curvature imposed to the endodontic instruments during the experimental tests were equal to 4.
7 mm and 45°, respectively.
In addition, 18 endodontic files (out of the 22) were subjected to electrochemical polishing before in vitro fatigue tests had been carried out.
Several combinations of electrochemical parameters were considered during the preliminary experiments involving electrolyte, voltage, flow rate and duration.
The endodontic files under study were weighted using a high-precision balance before and after polishing, and mass variation was determined.
Also, a sterilising cycle was applied to six polished endodontic instruments, which were placed inside an autoclave at 134°C, for 30 min, and under a pressure of 2.
16 bar.
Findings The fatigue resistance of as-received endodontic files ref.
0.
04/20 is much larger (+480 per cent) than the fatigue strength of endodontic files ref.
0.
06/20.
The weight variation due to electropolishing was comprehended between 0.
53 mg and 1.
01 mg, and the fatigue resistance of all polished endodontic instruments increased between 30 per cent (0.
06/20) and 80 per cent (0.
04/20) when compared with the fatigue resistance of the as-received files.
The application of an autoclave’s sterilisation cycle had a negative impact on the fatigue resistance of instruments ref.
0.
04/20 and a positive impact on the durability of instruments ref.
0.
06/20.
Originality/value The application of an electrochemical polishing process increased the fatigue resistance of the endodontic files under study significantly, and the values for the variables of the electrochemical process were never published elsewhere.
The surface finish improvement was obtained by immersing the endodontic files in a Struers’ Electrolyte (ref.
E2) for 3 s while applying 30 volts and a flow rate of 1.

Related Results

Recent Patents for Optical Component Polishing Technology
Recent Patents for Optical Component Polishing Technology
Background: Fluid jet polishing, Airbag polishing, Magnetorheological polishing, and Ion beam polishing are all emerging polishing technologies commonly used for processing optical...
Fatigue resistance of rotary endodontic files subjected to planar and non-planar curvatures induced by in vitro tooth canals
Fatigue resistance of rotary endodontic files subjected to planar and non-planar curvatures induced by in vitro tooth canals
Purpose The purpose of this paper is to assess the fatigue life of different sizes of HyFlex CM™ endodontic files when submitted to planar or to non-planar curvatures, and to two d...
Health care and contraceptive decision-making autonomy and use of female sterilisation among married women in Malawi
Health care and contraceptive decision-making autonomy and use of female sterilisation among married women in Malawi
IntroductionFemale sterilisation is the most common contraceptive method used globally. The use of female sterilisation is disproportionately low in sub-Sahara Africa (SSA) at just...
EFEKTIVITAS YOGA UNTUK MENGURANGI FATIGUE PADA PASIEN KANKER YANG MENJALANI KEMOTERAPI
EFEKTIVITAS YOGA UNTUK MENGURANGI FATIGUE PADA PASIEN KANKER YANG MENJALANI KEMOTERAPI
ABSTRAKLatar Belakang : Cancer Related Fatigue (CRF) adalah manifestasi klinis yang serius dan gejala umum yang dialami oleh pasien kanker. Fatigue adalah salah satu yang paling se...
Sterilisation of Dental Implant Surgical Instruments, Office Area and Implant Components: A Narrative Review
Sterilisation of Dental Implant Surgical Instruments, Office Area and Implant Components: A Narrative Review
Implant surgical procedures require meticulous attention to sterilisation protocols to prevent infections and ensure successful outcomes. Various sterilisation methods, such as ste...
Understanding the Effects of Sterilisation Methods on Orthodontic Archwires Properties: A Review
Understanding the Effects of Sterilisation Methods on Orthodontic Archwires Properties: A Review
Sterilisation is the removal of all microorganisms, vegetative or spore-forming, from an article, surface or medium that can be accomplished through steam autoclaving, dry heat and...
Study on Polishing Characteristics of Pyramidal Structured Polishing Pad
Study on Polishing Characteristics of Pyramidal Structured Polishing Pad
We conducted a polishing test to clarify the change in polishing characteristics resulting from the wear of a pyramidal-structured polishing tool, and discuss the polishing mechani...

Back to Top