Fatigue resistance of rotary endodontic files subjected to planar and non-planar curvatures induced by in vitro tooth canals
View through CrossRef
Purpose
The purpose of this paper is to assess the fatigue life of different sizes of HyFlex CM™ endodontic files when submitted to planar or to non-planar curvatures, and to two different rotational speeds, namely 500 rpm or 250 rpm. The influence of superimposing back and forth motion to rotational bending of endodontic files was also assessed.
Design/methodology/approach
In all, 64 HyFlex CM™ files of different sizes, namely ref. 0.04/20, 0.06/20, 0.04/35 and 0.06/35, were submitted to rotational bending tests at two different rotational speeds. The planar radius of curvature imposed to the endodontic files was about 4.5 mm, along an angle of 45°, in order to simulate an apical canal that would induce severe loading to the files during clinical treatment. Additionally, 11 Hyflex CM™ files ref. 0.04/20 were submitted to rotational bending tests at 500 rpm inside a mandible first molar manufactured through selective laser melting (SLM), aiming to simulate non-planar curvatures of a real tooth canal.
Findings
When considering planar curvature, the endodontic file ref. 0.06/20, tested at 250 rpm, has shown the highest fatigue resistance (4,185 revolutions, 1,004 seconds), while the lowest fatigue resistance was registered for instrument ref. 0.04/35 submitted to 500 rpm (747 revolutions, 89 seconds). Hence, depending on the rotational speed, surface finish and the size of the endodontic files tested (taper and tip’s diameter), significant differences in fatigue resistance were noticed. If non-planar curvatures were considered, the minimum fatigue resistance was equal to 107 seconds, and back and forth motion allowed extending the minimum fatigue lifetime to 140 seconds.
Originality/value
The fatigue resistance of endodontic files is frequently determined through in vitro fatigue tests carried out under single planar curvature or eventually under double planar curvatures. However, non-planar loadings are frequently induced on endodontic instruments when treating root canals with severe multiplanar curvatures. In the research herein presented, a mandible first molar was 3D printed by using the SLM technique in an AISI 316 L stainless steel and more realistic experimental fatigue tests were carried out.
Title: Fatigue resistance of rotary endodontic files subjected to planar and non-planar curvatures induced by in vitro tooth canals
Description:
Purpose
The purpose of this paper is to assess the fatigue life of different sizes of HyFlex CM™ endodontic files when submitted to planar or to non-planar curvatures, and to two different rotational speeds, namely 500 rpm or 250 rpm.
The influence of superimposing back and forth motion to rotational bending of endodontic files was also assessed.
Design/methodology/approach
In all, 64 HyFlex CM™ files of different sizes, namely ref.
0.
04/20, 0.
06/20, 0.
04/35 and 0.
06/35, were submitted to rotational bending tests at two different rotational speeds.
The planar radius of curvature imposed to the endodontic files was about 4.
5 mm, along an angle of 45°, in order to simulate an apical canal that would induce severe loading to the files during clinical treatment.
Additionally, 11 Hyflex CM™ files ref.
0.
04/20 were submitted to rotational bending tests at 500 rpm inside a mandible first molar manufactured through selective laser melting (SLM), aiming to simulate non-planar curvatures of a real tooth canal.
Findings
When considering planar curvature, the endodontic file ref.
0.
06/20, tested at 250 rpm, has shown the highest fatigue resistance (4,185 revolutions, 1,004 seconds), while the lowest fatigue resistance was registered for instrument ref.
0.
04/35 submitted to 500 rpm (747 revolutions, 89 seconds).
Hence, depending on the rotational speed, surface finish and the size of the endodontic files tested (taper and tip’s diameter), significant differences in fatigue resistance were noticed.
If non-planar curvatures were considered, the minimum fatigue resistance was equal to 107 seconds, and back and forth motion allowed extending the minimum fatigue lifetime to 140 seconds.
Originality/value
The fatigue resistance of endodontic files is frequently determined through in vitro fatigue tests carried out under single planar curvature or eventually under double planar curvatures.
However, non-planar loadings are frequently induced on endodontic instruments when treating root canals with severe multiplanar curvatures.
In the research herein presented, a mandible first molar was 3D printed by using the SLM technique in an AISI 316 L stainless steel and more realistic experimental fatigue tests were carried out.
Related Results
Influence of the Geometrical Cross-Section Design on the Dynamic Cyclic Fatigue Resistance of NiTi Endodontic Rotary Files—An In Vitro Study
Influence of the Geometrical Cross-Section Design on the Dynamic Cyclic Fatigue Resistance of NiTi Endodontic Rotary Files—An In Vitro Study
The aim of this study was to analyze and compare the influence of the geometrical cross-section design on the dynamic cyclic fatigue resistance of NiTi endodontic rotary files. Mat...
On the fatigue resistance of endodontic files subjected to electrochemical polishing and an autoclave’s sterilisation cycle
On the fatigue resistance of endodontic files subjected to electrochemical polishing and an autoclave’s sterilisation cycle
Purpose
The purpose of this paper is to improve the surface finish and, therefore, the fatigue resistance of some endodontic files by electrochemical polishing; the influence of ap...
EFEKTIVITAS YOGA UNTUK MENGURANGI FATIGUE PADA PASIEN KANKER YANG MENJALANI KEMOTERAPI
EFEKTIVITAS YOGA UNTUK MENGURANGI FATIGUE PADA PASIEN KANKER YANG MENJALANI KEMOTERAPI
ABSTRAKLatar Belakang : Cancer Related Fatigue (CRF) adalah manifestasi klinis yang serius dan gejala umum yang dialami oleh pasien kanker. Fatigue adalah salah satu yang paling se...
Pediatric Rotary Files: Evolution to Revolution
Pediatric Rotary Files: Evolution to Revolution
The main goal of pulp therapy in primary dentition is to preserve the primary tooth thus protecting future normal occlusion. Routinely, pulp debridement and canal shaping ar...
The Effect of Taper and Apical Diameter on the Cyclic Fatigue Resistance of Rotary Endodontic Files Using an Experimental Electronic Device
The Effect of Taper and Apical Diameter on the Cyclic Fatigue Resistance of Rotary Endodontic Files Using an Experimental Electronic Device
The aim of this study was to analyze the effect of the taper and apical diameter of nickel–titanium (NiTi) endodontic rotary files on the dynamic cyclic fatigue resistance. A total...
The Influence of NiTi Alloy on the Cyclic Fatigue Resistance of Endodontic Files
The Influence of NiTi Alloy on the Cyclic Fatigue Resistance of Endodontic Files
Background: The aim of this study was to analyze the influence of NiTi alloy in endodontic rotary instruments on cyclic fatigue resistance. Methods: One hundred and sixty-four (164...
Comparison of Cyclic Fatigue Resistance of Five Heat Treated NiTi Rotary Systems in Single and Double Curved Canals.
Comparison of Cyclic Fatigue Resistance of Five Heat Treated NiTi Rotary Systems in Single and Double Curved Canals.
Abstract
BackgroundThe aim of this study was to compare cyclic fatigue resistance (CFR) and separation patterns during rotating movement of five heat-treated nickel titaniu...
Differences in the confluence of mesial canals in mandibular molar teeth with three or four root canals
Differences in the confluence of mesial canals in mandibular molar teeth with three or four root canals
AbstractAim To investigate the correlation between number of root canals and their confluence in mandibular molar teeth.Methodology A total of 553 first and 383 second mandibular...