1322Impact of system impedance in ablation-index guided lesion formation in an in vitro model
View through CrossRef
Abstract
Funding Acknowledgements
Beca de la Sección de Electrofisiología y Arritmias para formación en investigación post-residencia en centros españoles
INTRODUCTION
Ablation Index and Lesion Size Index are novel markers of ablation lesion quality that incorporate power (W), contact force (CF), and time (T) in a weighted formula to accurately estimate lesion size. While LSI includes system impedance (I) in the formula, this parameter is not considered for AI calculation.
PURPOSE
The aim of our study was to analyse the impact of I in lesion formation.
METHODS
Radiofrequency (RF) lesions were created on porcine myocardial slabs by using an open-tip irrigated catheter with real-time monitoring of CF (a TactiCath™ Quartz Contact Force Ablation Catheter for LSI-guided lesions and a ThermoCool® SmartTouch® Catheter for AI-guided lesions). Catheters were mounted in an articulated arm and manually maneuvered over a platform placed within a tank filled with either a low impedance medium (saline solution) or a high impedance medium (running water) with a heating pump to produce non-pulsatile flow directed perpendicularly to the myocardium surface at a rate of 5 l/min and a temperature of 37ºC. For each catheter and medium, ten lesions were performed with a CF of 20 grams and a W of 25, 30, 35 and 40 watts aiming at a LSI of 5 or a AI of 400 (recommended targets for posterior wall pulmonary vein isolation for each index). After RF delivery, the myocardium was cross-sectioned at the level of each lesion and its blanched zone was measured with a dial calliper with a 0.1 mm resolution (image, panel A). Lesion volume was calculated using width and depth, according to a validated formula (image, panel B).
RESULTS
No difference in application time was noted in the AI-guided lesions depending on the medium (23.8 ± 1.6s, 17.2 ± 1.4s, 13.7 ± 0.8s and 12.2 ± 0.7s for 25W, 30W, 35W and 40W, respectively) while a significant reduction in application time was noted in the LSI-guided lesions performed in a low I medium (93.8 ± 6.6 ohms; 17.9 ± 1.9s, 11.3 ± 3.7s, 9.1 ± 0.9s and 7.3 ± 0.5s for 25W, 30W, 35W and 40W) compared with those performed in a high impedance medium (132.4 ± 6.6 ohms; 27.3 ± 1.4s, 17.9 ± 2.5s, 15.5 ± 1.4 and 13 ± 1.2 for the same programmed W (p < 0.0001 for all comparisons).
Both catheters showed differences in lesion size according to the I, with lower lesion volumes in the low I medium and higher lesion volumes in the high I medium. Although globally smaller in size, higher lesion volumes were noted in the AI-guided lesions compared to LSI-guided lesions in a low I medium (image, panel C). On the contrary, a better performance of the LSI-guided lesions was observed in the high I medium (image, panel D).
CONCLUSIONS
System impedance affected lesion size both in LSI-guided an AI-guided ablation lesions, with smaller lesions seen in a lower I medium. Further studies are needed to assess the clinical relevance of this findings in an in-vivo scenario.
Abstract Figure.
Oxford University Press (OUP)
Title: 1322Impact of system impedance in ablation-index guided lesion formation in an in vitro model
Description:
Abstract
Funding Acknowledgements
Beca de la Sección de Electrofisiología y Arritmias para formación en investigación post-residencia en centros españoles
INTRODUCTION
Ablation Index and Lesion Size Index are novel markers of ablation lesion quality that incorporate power (W), contact force (CF), and time (T) in a weighted formula to accurately estimate lesion size.
While LSI includes system impedance (I) in the formula, this parameter is not considered for AI calculation.
PURPOSE
The aim of our study was to analyse the impact of I in lesion formation.
METHODS
Radiofrequency (RF) lesions were created on porcine myocardial slabs by using an open-tip irrigated catheter with real-time monitoring of CF (a TactiCath™ Quartz Contact Force Ablation Catheter for LSI-guided lesions and a ThermoCool® SmartTouch® Catheter for AI-guided lesions).
Catheters were mounted in an articulated arm and manually maneuvered over a platform placed within a tank filled with either a low impedance medium (saline solution) or a high impedance medium (running water) with a heating pump to produce non-pulsatile flow directed perpendicularly to the myocardium surface at a rate of 5 l/min and a temperature of 37ºC.
For each catheter and medium, ten lesions were performed with a CF of 20 grams and a W of 25, 30, 35 and 40 watts aiming at a LSI of 5 or a AI of 400 (recommended targets for posterior wall pulmonary vein isolation for each index).
After RF delivery, the myocardium was cross-sectioned at the level of each lesion and its blanched zone was measured with a dial calliper with a 0.
1 mm resolution (image, panel A).
Lesion volume was calculated using width and depth, according to a validated formula (image, panel B).
RESULTS
No difference in application time was noted in the AI-guided lesions depending on the medium (23.
8 ± 1.
6s, 17.
2 ± 1.
4s, 13.
7 ± 0.
8s and 12.
2 ± 0.
7s for 25W, 30W, 35W and 40W, respectively) while a significant reduction in application time was noted in the LSI-guided lesions performed in a low I medium (93.
8 ± 6.
6 ohms; 17.
9 ± 1.
9s, 11.
3 ± 3.
7s, 9.
1 ± 0.
9s and 7.
3 ± 0.
5s for 25W, 30W, 35W and 40W) compared with those performed in a high impedance medium (132.
4 ± 6.
6 ohms; 27.
3 ± 1.
4s, 17.
9 ± 2.
5s, 15.
5 ± 1.
4 and 13 ± 1.
2 for the same programmed W (p < 0.
0001 for all comparisons).
Both catheters showed differences in lesion size according to the I, with lower lesion volumes in the low I medium and higher lesion volumes in the high I medium.
Although globally smaller in size, higher lesion volumes were noted in the AI-guided lesions compared to LSI-guided lesions in a low I medium (image, panel C).
On the contrary, a better performance of the LSI-guided lesions was observed in the high I medium (image, panel D).
CONCLUSIONS
System impedance affected lesion size both in LSI-guided an AI-guided ablation lesions, with smaller lesions seen in a lower I medium.
Further studies are needed to assess the clinical relevance of this findings in an in-vivo scenario.
Abstract Figure.
Related Results
Microwave Ablation with or Without Chemotherapy in Management of Non-Small Cell Lung Cancer: A Systematic Review
Microwave Ablation with or Without Chemotherapy in Management of Non-Small Cell Lung Cancer: A Systematic Review
Abstract
Introduction
Microwave ablation (MWA) has emerged as a minimally invasive treatment for patients with inoperable non-small cell lung cancer (NSCLC). However, whether it i...
EPD Electronic Pathogen Detection v1
EPD Electronic Pathogen Detection v1
Electronic pathogen detection (EPD) is a non - invasive, rapid, affordable, point- of- care test, for Covid 19 resulting from infection with SARS-CoV-2 virus. EPD scanning techno...
Ablation Performance of C/SiC-ZrC Composites
Ablation Performance of C/SiC-ZrC Composites
Abstract
To reveal the ablation performance of C/SiC-ZrC composites under different ablation methods, C/SiC-ZrC composites were prepared by chemical vapor deposition and pr...
Ablation index-guided ablation with milder targets for atrial fibrillation: Comparison between high power and low power ablation
Ablation index-guided ablation with milder targets for atrial fibrillation: Comparison between high power and low power ablation
BackgroundHigh power-ablation index (HP-AI)-guided ablation for atrial fibrillation (AF) targeting high AIs has been implemented in European countries. However, milder AI targets a...
Abstract 12300: Impact of Catheter Contact Angle on Local Impedance Drop and Lesion Size Using the INTELLANAV STABLEPOINT Ablation Catheter: Insight From a Porcine Experimental Study
Abstract 12300: Impact of Catheter Contact Angle on Local Impedance Drop and Lesion Size Using the INTELLANAV STABLEPOINT Ablation Catheter: Insight From a Porcine Experimental Study
Background:
Local impedance (LI) at a distal tip of the ablation catheter can indirectly measure tissue temperature during radiofrequency catheter ablation (RFCA). LI d...
Prospective use of Ablation Index targets improves clinical outcomes following ablation for atrial fibrillation
Prospective use of Ablation Index targets improves clinical outcomes following ablation for atrial fibrillation
AbstractAimsLate recovery of ablated tissue leading to reconnection of pulmonary veins remains common following radiofrequency catheter ablation for AF. Ablation Index (AI), a nove...
Entirely anatomical high-power ablation for refractory left ventricular summit arrhythmias
Entirely anatomical high-power ablation for refractory left ventricular summit arrhythmias
Abstract
Introduction
Ventricular arrhythmias (VA) originating from the left ventricular summit (LVS) can be difficult to...
Local multifrequency impedance to characterize RF ablations
Local multifrequency impedance to characterize RF ablations
Abstract
Introduction
Radiofrequency (RF) ablation is one of the principal treatments of cardiac arrhythmias. The electroanatomi...