Prospective use of Ablation Index targets improves clinical outcomes following ablation for atrial fibrillation
View through CrossRef
AbstractAimsLate recovery of ablated tissue leading to reconnection of pulmonary veins remains common following radiofrequency catheter ablation for AF. Ablation Index (AI), a novel ablation quality marker, incorporates contact force (CF), time, and power in a weighted formula. We hypothesized that prospective use of our previously published derived AI targets would result in better outcomes when compared to CF‐guided ablation.MethodsEighty‐nine consecutive drug‐refractory AF patients (49% paroxysmal) underwent AI‐guided ablation (AI‐group). AI targets were 550 for anterior/roof and 400 for posterior/inferior left atrial segments. Procedural and clinical outcomes of these patients were compared to 89 propensity‐matched controls who underwent CF‐guided ablation (CF‐group). All 178 procedures were otherwise similar, and both groups were followed‐up for 12 months. The last 25 patients from each group underwent analysis of all VisiTags™ for ablation duration, CF, Force‐Time Integral, and impedance drop.ResultsFirst‐pass pulmonary vein isolation (PVI) was more frequent in AI‐group than in CF‐group (173 [97%] vs. 149 [84%] circles, P < 0.001), and acute PV reconnection was lower (11 [6%] vs. 24 [13%] circles, P = 0.02). Mean PVI ablation time was similar (AI‐group: 42 ± 9 vs. CF‐group: 45 ± 14 minutes, P = 0.14). Median impedance drop for AI‐group was significantly higher than in CF‐group (13.7 [9‐19] Ω vs. 8.8 [5.2‐13] Ω, P < 0.001). Two major complications occurred in CF‐group and none in AI‐group. Atrial tachyarrhythmia recurrence was significantly lower in AI‐group (15 of 89 [17%]) than in CF‐group (33 of 89 [37%], P = 0.002).ConclusionAI‐guided ablation is associated with significant improvements in the incidence of acute PV reconnection and atrial tachyarrhythmia recurrence rate compared to CF‐guided ablation, potentially due to creation of better quality lesions as suggested by greater impedance drop.
Title: Prospective use of Ablation Index targets improves clinical outcomes following ablation for atrial fibrillation
Description:
AbstractAimsLate recovery of ablated tissue leading to reconnection of pulmonary veins remains common following radiofrequency catheter ablation for AF.
Ablation Index (AI), a novel ablation quality marker, incorporates contact force (CF), time, and power in a weighted formula.
We hypothesized that prospective use of our previously published derived AI targets would result in better outcomes when compared to CF‐guided ablation.
MethodsEighty‐nine consecutive drug‐refractory AF patients (49% paroxysmal) underwent AI‐guided ablation (AI‐group).
AI targets were 550 for anterior/roof and 400 for posterior/inferior left atrial segments.
Procedural and clinical outcomes of these patients were compared to 89 propensity‐matched controls who underwent CF‐guided ablation (CF‐group).
All 178 procedures were otherwise similar, and both groups were followed‐up for 12 months.
The last 25 patients from each group underwent analysis of all VisiTags™ for ablation duration, CF, Force‐Time Integral, and impedance drop.
ResultsFirst‐pass pulmonary vein isolation (PVI) was more frequent in AI‐group than in CF‐group (173 [97%] vs.
149 [84%] circles, P < 0.
001), and acute PV reconnection was lower (11 [6%] vs.
24 [13%] circles, P = 0.
02).
Mean PVI ablation time was similar (AI‐group: 42 ± 9 vs.
CF‐group: 45 ± 14 minutes, P = 0.
14).
Median impedance drop for AI‐group was significantly higher than in CF‐group (13.
7 [9‐19] Ω vs.
8.
8 [5.
2‐13] Ω, P < 0.
001).
Two major complications occurred in CF‐group and none in AI‐group.
Atrial tachyarrhythmia recurrence was significantly lower in AI‐group (15 of 89 [17%]) than in CF‐group (33 of 89 [37%], P = 0.
002).
ConclusionAI‐guided ablation is associated with significant improvements in the incidence of acute PV reconnection and atrial tachyarrhythmia recurrence rate compared to CF‐guided ablation, potentially due to creation of better quality lesions as suggested by greater impedance drop.
Related Results
Rethinking Anticoagulation in Kidney Disease and Kidney Transplantation
Rethinking Anticoagulation in Kidney Disease and Kidney Transplantation
BackgroundA clinical dilemma exists in patients with CKD, ESKD, and after kidney transplantation, as they face concurrent risks of both thromboembolic and hemorrhagic complications...
Inter‐Relationships Between Atrial Flutter and Atrial Fibrillation
Inter‐Relationships Between Atrial Flutter and Atrial Fibrillation
It has been appreciated for a long time that atrial flutter and atrial fibrillation have a clinical relationship. Now, with the technological advances that permit more sophisticate...
Adiponectin and Lone atrial fibrillation
Adiponectin and Lone atrial fibrillation
Objective: Lone atrial fibrillation is an idiopathic arrhythmia seen in younger individuals without any secondary disease. Adiponectin is an endogenous adipocytokine that increases...
Network Pharmacology and Molecular Docking Reveal Wenxin Granule’s Anti-Atrial Fibrillation Mechanisms
Network Pharmacology and Molecular Docking Reveal Wenxin Granule’s Anti-Atrial Fibrillation Mechanisms
Objective: Wenxin Granules (WXG) is a Chinese medicine preparation made from five traditional Chinese medicines, including Polygonal circumscribe, Synopsis peninsula, Panax nothing...
Beyond Coronary Risk: Clinical Scores as Predictors of Atrial Fibrillation in Chronic Coronary Syndrome
Beyond Coronary Risk: Clinical Scores as Predictors of Atrial Fibrillation in Chronic Coronary Syndrome
Atrial fibrillation frequently coexists with chronic coronary syndrome, sharing common cardiovascular risk factors and pathophysiological mechanisms. Identifying patients with chro...
La enfermera experta en fibrilación auricular
La enfermera experta en fibrilación auricular
La fibrilación auricular es la arritmia cardiaca más prevalente de nuestro entorno. Las clínicas de fibrilación auricular lideradas por enfermería nacen para garantizar el manejo i...
Case report: Para Hisian Atrial Tachycardia
Case report: Para Hisian Atrial Tachycardia
Atrial tachycardia accounts for around 10% of supraventricular tachycardias (1). Atrial tachycardia can either be focal or macrorenentrant. Focal AT usually start at a focal p...
RELATIONSHIP BETWEEN ATRIAL FIBRILLATION CARDIOVERSION AND F
RELATIONSHIP BETWEEN ATRIAL FIBRILLATION CARDIOVERSION AND F
Objectives
To investigate the relationship between atrial fibrillation cardioversion and f wave in electrocardiogram, providing an ordinary and noninvasive method...