Influence of Build Platform Position on the Trueness and Precision of LCD 3D-Printed Orthognathic Surgical Splints
View through CrossRef
Background/Objectives: Among vat polymerization technologies, liquid-crystal display (LCD) 3D printing has gained popularity in dentistry because of its affordability and acceptable resolution. However, the factors influencing the dimensional accuracy of LCD-printed surgical splints, particularly build platform position, remain insufficiently investigated. This study aimed to evaluate the influence of build platform position on the trueness and precision of orthognathic surgical splints fabricated using LCD 3D printing technology. Methods: Thirty-six surgical splints were printed from a master digital file using an LCD 3D printer. All surgical splints were printed with a 90-degree layer orientation to the building platform. The layer thickness was set at 100 μm. The surgical splints were divided into three groups according to their printing position on the building platform: middle (M), left (Lt), and right (Rt). Each 3D-printed surgical splint was sprayed with an opaque scanning spray and then rescanned to create digital files for testing. A surface-based superimposition and deviation analysis was performed using specialized 3D software to evaluate accuracy of surgical splints. Root mean square error (RMSE) values were statistically analyzed. Results: There were no statistically significant differences in trueness among the middle, left, and right printing positions on the build platform (p > 0.05). In contrast, printing position significantly affected precision, with surgical splints printed at the center of the build platform demonstrating significantly lower RMSE values compared with those printed at the left and right positions (p < 0.001). In addition, no significant difference in precision was detected between the left and right positions. Conclusions: The printing position on the build platform significantly influences the precision of orthognathic surgical splints fabricated using LCD 3D printing technology. Splints printed at the central region of the build platform exhibited the highest precision, whereas trueness was not significantly affected by printing position. These findings suggest that preferential placement of surgical splints at the center of the build platform may improve fabrication consistency and predictability in digital orthognathic surgery workflows.
Title: Influence of Build Platform Position on the Trueness and Precision of LCD 3D-Printed Orthognathic Surgical Splints
Description:
Background/Objectives: Among vat polymerization technologies, liquid-crystal display (LCD) 3D printing has gained popularity in dentistry because of its affordability and acceptable resolution.
However, the factors influencing the dimensional accuracy of LCD-printed surgical splints, particularly build platform position, remain insufficiently investigated.
This study aimed to evaluate the influence of build platform position on the trueness and precision of orthognathic surgical splints fabricated using LCD 3D printing technology.
Methods: Thirty-six surgical splints were printed from a master digital file using an LCD 3D printer.
All surgical splints were printed with a 90-degree layer orientation to the building platform.
The layer thickness was set at 100 μm.
The surgical splints were divided into three groups according to their printing position on the building platform: middle (M), left (Lt), and right (Rt).
Each 3D-printed surgical splint was sprayed with an opaque scanning spray and then rescanned to create digital files for testing.
A surface-based superimposition and deviation analysis was performed using specialized 3D software to evaluate accuracy of surgical splints.
Root mean square error (RMSE) values were statistically analyzed.
Results: There were no statistically significant differences in trueness among the middle, left, and right printing positions on the build platform (p > 0.
05).
In contrast, printing position significantly affected precision, with surgical splints printed at the center of the build platform demonstrating significantly lower RMSE values compared with those printed at the left and right positions (p < 0.
001).
In addition, no significant difference in precision was detected between the left and right positions.
Conclusions: The printing position on the build platform significantly influences the precision of orthognathic surgical splints fabricated using LCD 3D printing technology.
Splints printed at the central region of the build platform exhibited the highest precision, whereas trueness was not significantly affected by printing position.
These findings suggest that preferential placement of surgical splints at the center of the build platform may improve fabrication consistency and predictability in digital orthognathic surgery workflows.
Related Results
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Abstarct
Introduction
Isolated brain hydatid disease (BHD) is an extremely rare form of echinococcosis. A prompt and timely diagnosis is a crucial step in disease management. This ...
Rigidity of commonly used dental trauma splints
Rigidity of commonly used dental trauma splints
Abstract – Background/Aims: The stability of immobilization devices varies from flexible to rigid, depending on the trauma. We evaluated the rigidity of various commonly used spl...
Effect of Milling Protocols on Trueness and Precision of Ceramic Crowns
Effect of Milling Protocols on Trueness and Precision of Ceramic Crowns
AbstractPurposeTo investigate the effect of different milling protocols for different ceramic materials on the trueness and precision of milled ceramic crowns.Materials and Methods...
Current Concepts in Orthognathic Surgery
Current Concepts in Orthognathic Surgery
Learning Objectives:
After studying this article, the participant should be able to: 1. Identify skeletal differences that are treated with orthognathic surgery...
Fit, Precision, and Trueness of 3D-Printed Zirconia Crowns Compared to Milled Counterparts
Fit, Precision, and Trueness of 3D-Printed Zirconia Crowns Compared to Milled Counterparts
Precise fit of a crown and accurate reproduction of the digital design are paramount for successful treatment outcomes and preservation of clinician and technician time. The study ...
ecision Farming and Predictive Analytics in Precision Farming and Predictive Analytics in Precision Farming and Predictive Analytics in Precision Farming and Predictive Analytics in Precision Farming and Predictive Analytics in Precision Farming and Predi
ecision Farming and Predictive Analytics in Precision Farming and Predictive Analytics in Precision Farming and Predictive Analytics in Precision Farming and Predictive Analytics in Precision Farming and Predictive Analytics in Precision Farming and Predi
The scope of sensor networks and the Internet of Things spanning rapidly to diversified domains but not limited to sports, health, and business trading. In recent past, the sensors...
Compliance with Hand Splints Among Patients of Rheumatoid Arthritis at Benazir Bhutto Hospital Rawalpindi Pakistan
Compliance with Hand Splints Among Patients of Rheumatoid Arthritis at Benazir Bhutto Hospital Rawalpindi Pakistan
Objectives: To determine the compliance with hand splints among patients of Rheumatoid arthritis at Benazir Bhutto Hospital Rawalpindi and their satisfaction with usage of splints....
Comparative Study of Outcome in Septoplasty with and Without Postoperative Intranasal Slastic Splints
Comparative Study of Outcome in Septoplasty with and Without Postoperative Intranasal Slastic Splints
Background: Septal deviation and chronic nasal obstruction are commonly seen and affect most of the population, affecting respiratory function and life quality. The use of intranas...