Effect of Pedicle Fill on Axial Pullout Strength in Spinal Fixation After Rod Reduction

View through CrossRef
Rod reduction to pedicle screws is used for a variety of spinal fixation procedures; however, it can alter the integrity of the screw-bone interface. The authors investigated the effect of pedicle fill (ratio of pedicle screw diameter to pedicle diameter) on the strength of the screw-bone interface after simulated rod reduction on 17 vertebrae (3 thoracolumbar spine specimens). Pedicle diameter was measured with standard clinical computed tomography scan protocols. The authors determined the minimum pedicle diameter for each level. Polyaxial pedicle screws were surgically placed bilaterally with a freehand technique with standard clinical anatomic landmarks. The pedicle pairs were instrumented with pedicle screws of predetermined diameter, 1 with greater than 80% fill and 1 with less than 80% fill. A simulated reduction maneuver was performed with a 5-mm gap followed by an axial pullout test to assess screw interface strength. Comparison of insertion torque between less than 80% fill and greater than 80% fill did not show significant increases. A significant difference in pullout load ( P =.043) occurred with greater than 80% fill (791±637 N) compared with less than 80% fill (636±492 N). No significant difference in stiffness was noted ( P =.154) with pedicle fill of greater than 80% (427±134 N/mm) compared with less than 80% (376±178 N/mm). The current findings support the use of greater than 80% pedicle fill for optimal screw anchoring in pedicle screw-based constructs involving rod reduction. Surgeons should consider placing screws that can safely fill vertebral pedicles, especially at the apex of the curve and the proximal and distal levels of constructs, where excessive forces are imparted to the screws. [ Orthopedics. 2017; 40(6):e990–e995.]
Title: Effect of Pedicle Fill on Axial Pullout Strength in Spinal Fixation After Rod Reduction
Description:
Rod reduction to pedicle screws is used for a variety of spinal fixation procedures; however, it can alter the integrity of the screw-bone interface.
The authors investigated the effect of pedicle fill (ratio of pedicle screw diameter to pedicle diameter) on the strength of the screw-bone interface after simulated rod reduction on 17 vertebrae (3 thoracolumbar spine specimens).
Pedicle diameter was measured with standard clinical computed tomography scan protocols.
The authors determined the minimum pedicle diameter for each level.
Polyaxial pedicle screws were surgically placed bilaterally with a freehand technique with standard clinical anatomic landmarks.
The pedicle pairs were instrumented with pedicle screws of predetermined diameter, 1 with greater than 80% fill and 1 with less than 80% fill.
A simulated reduction maneuver was performed with a 5-mm gap followed by an axial pullout test to assess screw interface strength.
Comparison of insertion torque between less than 80% fill and greater than 80% fill did not show significant increases.
A significant difference in pullout load ( P =.
043) occurred with greater than 80% fill (791±637 N) compared with less than 80% fill (636±492 N).
No significant difference in stiffness was noted ( P =.
154) with pedicle fill of greater than 80% (427±134 N/mm) compared with less than 80% (376±178 N/mm).
The current findings support the use of greater than 80% pedicle fill for optimal screw anchoring in pedicle screw-based constructs involving rod reduction.
Surgeons should consider placing screws that can safely fill vertebral pedicles, especially at the apex of the curve and the proximal and distal levels of constructs, where excessive forces are imparted to the screws.
[ Orthopedics.
2017; 40(6):e990–e995.
].

Related Results

Biomechanics of Artificial Pedicle Fixation in a 3d-printed Prosthesis After Total en Bloc Spondylectomy: A Finite Element Analysis
Biomechanics of Artificial Pedicle Fixation in a 3d-printed Prosthesis After Total en Bloc Spondylectomy: A Finite Element Analysis
Abstract Background: This study compared the biomechanics of artificial pedicle fixation in spine reconstruction with a 3-dimensional (3D)-printed prosthesis after total en...
Minimizing Pedicle Screw Pullout Risks
Minimizing Pedicle Screw Pullout Risks
Study Design: Detailed biomechanical analysis of the anchorage performance provided by different pedicle screw designs and placement strategies under pullout loading. ...
Biomechanical characterization of bilateral pedicle screw internal fixation combinations on lumbar vertebrae
Biomechanical characterization of bilateral pedicle screw internal fixation combinations on lumbar vertebrae
Purpose Pedicle screw fixation has been considered a suitable surgical intervention for addressing a diverse range of indications involving the lumbar spinal segments, but the...
Influence of Different Fixed System in Osteoporotic Lumbar Vertebrae in Fusion:A 3D Finite Element Analysis
Influence of Different Fixed System in Osteoporotic Lumbar Vertebrae in Fusion:A 3D Finite Element Analysis
Abstract Background:Posterior lumbar interbody fusion(PLIF) is the most common surgical method for lumbar fusion surgery, and pedicle screws(PS)can provide effective fixati...
Experimental study on the pullout characteristics of toothed warp knitted geogrid in different sand fillers
Experimental study on the pullout characteristics of toothed warp knitted geogrid in different sand fillers
In order to investigate the pullout characteristics of a toothed geogrid, a commercialized warp-knitted geogrid was combined with rigid angle aluminum to form a qualified toothed g...
Rod bipolar cells in the macaque monkey retina: immunoreactivity and connectivity
Rod bipolar cells in the macaque monkey retina: immunoreactivity and connectivity
Rod bipolar cells in the macaque monkey retina were labeled by three antibodies: an antibody against the alpha- and beta-subspecies of protein kinase C (PKC), a polyclonal antiseru...

Back to Top