Biomechanical Study: Determining the Optimum Insertion Angle for Screw‐In Suture Anchors—Is Deadman's Angle Correct?
View through CrossRef
Purpose
To assess the effect of the insertion angle and the angle of applied load on the pullout strength of screw‐in suture anchors.
Methods
Screw‐in metallic suture anchors were inserted into a 10–lb/cu ft synthetic cancellous bone block at 30°, 45°, 60°, and 90° to the surface. The suture pull angle was then varied in 30° increments between 0° and 180°. Five constructs were tested to failure (anchor pullout) for each combination of angles using a Zwick tensile testing machine (Zwick Roell, Ulm, Germany).
Results
There were a total of 25 combinations. The greatest pullout strength was seen with a suture anchor inserted at 90° to the bone block with a pull angle of 90° to the bone (mean, 306 N; standard deviation [SD], 9 N). The weakest pullout strength was seen with a suture anchor inserted at 30° with the angle of pull at 120° (i.e., opposite to the direction of insertion of the anchor) (mean, 97 N; SD, 11 N). A simulated deadman's angle of 45° with an angle of pull of 150° produced a pullout strength of 127 N (SD, 4 N). The pullout strengths for each insertion angle were greatest when the angle of pull was similar to the angle of insertion (
P
< .0001, repeated‐measures analysis of variance).
Conclusions
The angle of applied load to a suture anchor and the insertion angle significantly influence the biomechanical pullout strength of screw‐in suture anchors. The insertion angle of the suture anchor should replicate the angle of applied load to ensure the optimum pullout strength.
Clinical Relevance
The screw‐in anchor insertion angle and angle of applied load may have an influence on pullout strength.
Title: Biomechanical Study: Determining the Optimum Insertion Angle for Screw‐In Suture Anchors—Is Deadman's Angle Correct?
Description:
Purpose
To assess the effect of the insertion angle and the angle of applied load on the pullout strength of screw‐in suture anchors.
Methods
Screw‐in metallic suture anchors were inserted into a 10–lb/cu ft synthetic cancellous bone block at 30°, 45°, 60°, and 90° to the surface.
The suture pull angle was then varied in 30° increments between 0° and 180°.
Five constructs were tested to failure (anchor pullout) for each combination of angles using a Zwick tensile testing machine (Zwick Roell, Ulm, Germany).
Results
There were a total of 25 combinations.
The greatest pullout strength was seen with a suture anchor inserted at 90° to the bone block with a pull angle of 90° to the bone (mean, 306 N; standard deviation [SD], 9 N).
The weakest pullout strength was seen with a suture anchor inserted at 30° with the angle of pull at 120° (i.
e.
, opposite to the direction of insertion of the anchor) (mean, 97 N; SD, 11 N).
A simulated deadman's angle of 45° with an angle of pull of 150° produced a pullout strength of 127 N (SD, 4 N).
The pullout strengths for each insertion angle were greatest when the angle of pull was similar to the angle of insertion (
P
< .
0001, repeated‐measures analysis of variance).
Conclusions
The angle of applied load to a suture anchor and the insertion angle significantly influence the biomechanical pullout strength of screw‐in suture anchors.
The insertion angle of the suture anchor should replicate the angle of applied load to ensure the optimum pullout strength.
Clinical Relevance
The screw‐in anchor insertion angle and angle of applied load may have an influence on pullout strength.
Related Results
Biomechanical Analysis of Pullout Strengths of Rotator Cuff and Glenoid Anchors: 2011 Update
Biomechanical Analysis of Pullout Strengths of Rotator Cuff and Glenoid Anchors: 2011 Update
Purpose
To evaluate the biomechanical and design characteristics of newer suture anchors.
Methods
...
Concomitant Medial Meniscal Root Repair with Extrusion Repair (Centralization Technique)
Concomitant Medial Meniscal Root Repair with Extrusion Repair (Centralization Technique)
Background:
Meniscal extrusion is a phenomenon in which a degenerative posterior horn tear, radial tear, or root tear results in displacement of the body of the meniscu...
Arthroscopic Suture Anchor Capsulorrhaphy Versus Suture Capsulorrhaphy in a Cadaveric Model (SS‐01)
Arthroscopic Suture Anchor Capsulorrhaphy Versus Suture Capsulorrhaphy in a Cadaveric Model (SS‐01)
Introduction
Recurrent anterior glenohumeral instability is challenging to treat. Arthroscopic capsulorrhaphy utilizing either a solitary suture or suture ancho...
Suture-Induced Tubo-Ovarian Abscess: A Case Report with Literature Review
Suture-Induced Tubo-Ovarian Abscess: A Case Report with Literature Review
Abstract
Introduction
Suture is an underreported cause for tubo-ovarian abscess (TOA) that can cause significant morbidity. This report describes a case of TOA arising from a silk ...
#1089 Outcomes of medical and surgical peritoneal dialysis catheter insertion: a single centre experience
#1089 Outcomes of medical and surgical peritoneal dialysis catheter insertion: a single centre experience
Abstract
Background and Aims
Medical insertion of peritoneal dialysis (PD) catheters has been common practice within the ...
SS: MODU Anchor: Improved Predictions of MODU Fluke Anchors Increases the Confidence in Mooring System Design
SS: MODU Anchor: Improved Predictions of MODU Fluke Anchors Increases the Confidence in Mooring System Design
Abstract
An overview of the current Norwegian and international requirements for design and installation of fluke anchors is given. Important design and installat...
Field Experimental Study on the Uplift and Lateral Capacity of Deep Helical Anchors and Grouped Helical Anchors in Clays
Field Experimental Study on the Uplift and Lateral Capacity of Deep Helical Anchors and Grouped Helical Anchors in Clays
This research aims to investigate the bearing capability of deep helical anchors and grouped helical anchors under uplift or lateral loads using field experiments. Grouped helical ...
Development of porous swelling copolymeric bone anchor for orthopedic applications
Development of porous swelling copolymeric bone anchor for orthopedic applications
Bone anchors are crucial in orthopedic surgeries, aiding in fracture fixation, soft tissue reattachment, deformity correction, osteochondral defect repair, and fusion. Existing bon...