TURN-BY-TURN BEAM EXTRACTION DURING ACCELERATION IN A SYNCHROTRON
View through CrossRef
A synchrotron to accelerate protons or carbon ions for medical applications is being designed at Brookhaven National Laboratory (BNL). Single beam bunches with maximum beam energy of 1.18 GeV and 400 MeV/u for protons and carbon ions respectively will be extracted from the synchrotron at 15 Hz. For protons, the maximum required energy for irradiating a tumor is ∼206 MeV. A pencil-like proton beam containing ∼5.4×10
7
p/bunch delivers a therapeutic dose of 2.5 Gy in ∼1.5 minutes to treat a tumor of 1 liter volume. It will take ∼80 minutes with bunches containing 4.5×10
4
ions/bunch to deliver the same dose of 2.5 Gy with a 400 MeV/u pencil-like carbon beam. This extended treatment time when using carbon ions is not acceptable. In addition, the synchrotron cannot be controlled with a beam bunch containing such a low number of carbon ions. To overcome these two problems of the extended treatment time and the low bunch intensity required for the treatment when carbon ions are used, we have devised a method to “peel” the required 4.5×10
4
carbon-ions/bunch from the accelerating carbon beam bunch containing ∼10
8
ions/bunch and deliver them to the tumor on a “turn-by-turn” basis. Unlike other methods of beam extraction from a synchrotron, such as resonance extraction, this method does not allow for any beam losses during the extraction and the carbon beam can be peeled off in less than 15 ms during the acceleration or deceleration cycle of the synchrotron. Thus, this turn-by-turn beam extraction method provides beam with variable energy and precisely controlled beam current during the 30 ms acceleration or deceleration time.
World Scientific Pub Co Pte Ltd
Title: TURN-BY-TURN BEAM EXTRACTION DURING ACCELERATION IN A SYNCHROTRON
Description:
A synchrotron to accelerate protons or carbon ions for medical applications is being designed at Brookhaven National Laboratory (BNL).
Single beam bunches with maximum beam energy of 1.
18 GeV and 400 MeV/u for protons and carbon ions respectively will be extracted from the synchrotron at 15 Hz.
For protons, the maximum required energy for irradiating a tumor is ∼206 MeV.
A pencil-like proton beam containing ∼5.
4×10
7
p/bunch delivers a therapeutic dose of 2.
5 Gy in ∼1.
5 minutes to treat a tumor of 1 liter volume.
It will take ∼80 minutes with bunches containing 4.
5×10
4
ions/bunch to deliver the same dose of 2.
5 Gy with a 400 MeV/u pencil-like carbon beam.
This extended treatment time when using carbon ions is not acceptable.
In addition, the synchrotron cannot be controlled with a beam bunch containing such a low number of carbon ions.
To overcome these two problems of the extended treatment time and the low bunch intensity required for the treatment when carbon ions are used, we have devised a method to “peel” the required 4.
5×10
4
carbon-ions/bunch from the accelerating carbon beam bunch containing ∼10
8
ions/bunch and deliver them to the tumor on a “turn-by-turn” basis.
Unlike other methods of beam extraction from a synchrotron, such as resonance extraction, this method does not allow for any beam losses during the extraction and the carbon beam can be peeled off in less than 15 ms during the acceleration or deceleration cycle of the synchrotron.
Thus, this turn-by-turn beam extraction method provides beam with variable energy and precisely controlled beam current during the 30 ms acceleration or deceleration time.
Related Results
Analisis Numerik Kuat Lentur Balok Support Beam Curve Tile Beton Semi Pracetak Dengan Variasi Panjang Bentangan
Analisis Numerik Kuat Lentur Balok Support Beam Curve Tile Beton Semi Pracetak Dengan Variasi Panjang Bentangan
Semi precast slab is a combination of precast concrete which consist of the support beam and curve tile with the cast in place concrete. During the working process, support beam wi...
Advancements in optics control at the CERN Proton Synchrotron
Advancements in optics control at the CERN Proton Synchrotron
The CERN Proton Synchrotron (PS), operational since 1959, remains a critical component in CERN's accelerator complex, supporting downstream experiments and serving as an injector f...
Vævning over gruber
Vævning over gruber
Pits, Looms, and Loom Pits? In the archaeological quarterly Skalk (1959, no 2), Mogens Ørsnes described "The Problem of the Hundred Holes," pits "large and small, shallow and deep"...
Effects of width and density of supersonic molecule beam on penetration depth of tokamak
Effects of width and density of supersonic molecule beam on penetration depth of tokamak
The penetration depth and the fueling efficiency of the supersonic molecular beam injection (SMBI) are affected by both the intrinsic parameters of the SMBI and the parameters of b...
Development of a high-voltage power supply with ultra-low ripple of beam current on electron beam gun
Development of a high-voltage power supply with ultra-low ripple of beam current on electron beam gun
Electron beam guns with grid-controlled electrodes and hot-cathodes are widely used in accelerator physics facilities, high-voltage electron beam welding machines, electron beam 3D...
Design and early operation of a new-generation internal beam dump for CERN’s Super Proton Synchrotron
Design and early operation of a new-generation internal beam dump for CERN’s Super Proton Synchrotron
The Super Proton Synchrotron (SPS) is the last stage in the injector chain for CERN’s Large Hadron Collider, and it also provides proton and ion beams for several fixed-target expe...
SU‐E‐T‐47: A Monte Carlo Model of a Spot Scanning Proton Beam Based On a Synchrotron Proton Therapy Accelerator
SU‐E‐T‐47: A Monte Carlo Model of a Spot Scanning Proton Beam Based On a Synchrotron Proton Therapy Accelerator
Purpose:To build the model of a spot scanning proton beam for the dose calculation of a synchrotron proton therapy accelerator, which is capable of accelerating protons from 50 up ...
The Use of Synchrotron Radiation in Virology
The Use of Synchrotron Radiation in Virology
Introduction: The demand for new rapid methods for development of medicines for health care is becoming more relevant. Detailed knowledge of the spatial structure of viral proteins...