Abstract 6108: Cardiac 64-slice Multidetector Computerized Tomography in the Management of Prosthetic Heart Valve Obstruction

View through CrossRef
Differentiation of pannus (P) and thrombus (THR) in prosthetic heart valve obstruction (PHVO) is still a confusive state at the present cardiological practise. Transesophageal echocardiography (TEE) is not fully capable method in this condition. We considered that cardiac 64-slice multidetector computerized tomography (MDCT) may have a value in the management of PHVO. Thirtytwo of 37 pts (22 male, mean age 47±11 years, 17 with atrial fibrillation) with PHVO (35 mitral, 2 tricuspid) were detected with TEE and also with MDCT before and after thrombolytic therapy (TT). Five pts having monoleaflet valves were excluded due to excessive artifacts. Pre-TT MDCT revealed at least 1 culprit mass (CM) which was considered to be the cause of PHVO and the Hounsfield unit (HU) values of CM’s were measured. All pts underwent TT under the guidance of serial TEE. The study consisted of 2 groups up to response of CM’s to TT demonstrated by post-TT MDCT. Group I: Successfully lysed CM’s, Group II: Nonresponder CM’s. TEE disclosed THR in 26 pts, however the cause of PHVO in 6 pts remained undefined. Pre-TT MDCT disclosed 45 distinct CM’s on 32 valves. Post-TT MDCT disclosed the number of CM’s for groups I and II: 26 and 19, respectively. Significant difference between mean HU values of groups was found [87 (range 11–180), 331 (range 168–510), respectively, p<0.001]. A cut-off value of HU>200 predicted nonresponder CM’s (sensitivity 94%, specificity 96%, p<0.001). CM’s of 6 pts with undefined TEE findings were all in group II and pure P was demonstrated surgically. In group I, 17 of 26 (%65) CM’s were completely and the remaining partially lysed which was confirmed by TEE. A cut-off value of HU<100 predicted complete lysis (sensitivity 100%, specificity %95, p<0.001). Subgroup analysis of CM’s with HU=100–200 revealed 10% completely and 80% partially lysis and 10% remained unresponded as confirmed by TEE. MDCT can be a useful technique for evaluation of PHVO caused by THR or P formation. Higher HU values denote P whereas the lower values THR. MDCT can also predict the response to TT. The higher the HU value the less the successful outcome. The CM’s with HU>200 are almost always resistant to TT.
Title: Abstract 6108: Cardiac 64-slice Multidetector Computerized Tomography in the Management of Prosthetic Heart Valve Obstruction
Description:
Differentiation of pannus (P) and thrombus (THR) in prosthetic heart valve obstruction (PHVO) is still a confusive state at the present cardiological practise.
Transesophageal echocardiography (TEE) is not fully capable method in this condition.
We considered that cardiac 64-slice multidetector computerized tomography (MDCT) may have a value in the management of PHVO.
Thirtytwo of 37 pts (22 male, mean age 47±11 years, 17 with atrial fibrillation) with PHVO (35 mitral, 2 tricuspid) were detected with TEE and also with MDCT before and after thrombolytic therapy (TT).
Five pts having monoleaflet valves were excluded due to excessive artifacts.
Pre-TT MDCT revealed at least 1 culprit mass (CM) which was considered to be the cause of PHVO and the Hounsfield unit (HU) values of CM’s were measured.
All pts underwent TT under the guidance of serial TEE.
The study consisted of 2 groups up to response of CM’s to TT demonstrated by post-TT MDCT.
Group I: Successfully lysed CM’s, Group II: Nonresponder CM’s.
TEE disclosed THR in 26 pts, however the cause of PHVO in 6 pts remained undefined.
Pre-TT MDCT disclosed 45 distinct CM’s on 32 valves.
Post-TT MDCT disclosed the number of CM’s for groups I and II: 26 and 19, respectively.
Significant difference between mean HU values of groups was found [87 (range 11–180), 331 (range 168–510), respectively, p<0.
001].
A cut-off value of HU>200 predicted nonresponder CM’s (sensitivity 94%, specificity 96%, p<0.
001).
CM’s of 6 pts with undefined TEE findings were all in group II and pure P was demonstrated surgically.
In group I, 17 of 26 (%65) CM’s were completely and the remaining partially lysed which was confirmed by TEE.
A cut-off value of HU<100 predicted complete lysis (sensitivity 100%, specificity %95, p<0.
001).
Subgroup analysis of CM’s with HU=100–200 revealed 10% completely and 80% partially lysis and 10% remained unresponded as confirmed by TEE.
MDCT can be a useful technique for evaluation of PHVO caused by THR or P formation.
Higher HU values denote P whereas the lower values THR.
MDCT can also predict the response to TT.
The higher the HU value the less the successful outcome.
The CM’s with HU>200 are almost always resistant to TT.

Related Results

Comparison of Heart Valve Circumference Examined Before and After 10% Formalin Fixation
Comparison of Heart Valve Circumference Examined Before and After 10% Formalin Fixation
Objective: To compare the heart valve circumference before and after 10% formalin fixation. Materials and Methods: The study analyzed 63 Thai human cadaveric hearts. Each heart val...
An Acoustically Controlled Down-Hole Safety Valve (SCSSSV)
An Acoustically Controlled Down-Hole Safety Valve (SCSSSV)
American Institute of Mining, Metallurgical, and Petroleum Engineers, Inc. Abstract A surface-controlled subsurface safety valve...
Cardiac Myxoma Post-Transseptal Ablation: Coincidence or Causation?
Cardiac Myxoma Post-Transseptal Ablation: Coincidence or Causation?
Background: Cardiac myxomas are benign cardiac neoplasms usually found solitarily located within a single cardiac chamber, most commonly in the left atrium. With no established cau...
AN UNCOMMON CAUSE OF MECHANICAL PROSTHETIC VALVE OBSTRUCTION
AN UNCOMMON CAUSE OF MECHANICAL PROSTHETIC VALVE OBSTRUCTION
Abstract An 84–years–old man presented to the ED for worsening exertional dyspnoea and fever. Medical history included CKD, hypertension, permanent AF, type 2 diabet...
Advanced engineering materials and materials processing technologies in dental implant and prosthetic treatment with clinical cases
Advanced engineering materials and materials processing technologies in dental implant and prosthetic treatment with clinical cases
The article deals with materials science issues concerning the application areas in dental engineering. The monograph aims to present the results of the Author’s work against the b...
Role of multi-slice computed tomography in evaluation and management of intestinal obstruction
Role of multi-slice computed tomography in evaluation and management of intestinal obstruction
Background: The purpose of this study is to evaluate the role of multi-slice computed tomography (CT) with reference to presence or absence of intestinal obstruction, level of obst...

Back to Top