Mini-circuit cardiopulmonary bypass with vacuum assisted venous drainage: feasibility of an asanguineous prime in the neonate

View through CrossRef
Conventional cardiopulmonary bypass (CPB) in neonates results in increased transfusion requirements and hemodilution. There has been little advancement in CPB for the neonatal population. There is evidence that increased priming volumes and blood product transfusion enhances inflammatory response to CPB and increases myocardial and pulmonary dysfunction. We have devised a miniaturized CPB circuit that utilizes vacuum-assisted venous drainage (VAVD) in an effort to decrease priming volume and avoid transfusion requirements. The purpose of this study was to evaluate the safety and efficacy of this miniaturized CPB system and determine the feasibility of an asanguineous prime. Ten 1-week-old piglets were randomized to five mini- and five conventional CPB pump circuits. Subjects were supported with CPB at 100 ml/kg/min, cooled to 28°C, exposed to 10 min aortic crossclamp with cardioplegic arrest, rewarmed to 37°C, weaned from bypass, and subjected to modified-ultrafiltration (MUF) for approximately 10 min. This method was chosen to simulate a situation with all the elements of clinical CPB. Blood transfusion trigger was a hematocrit <15 on CPB. Serum samples were obtained pre-CPB, at 15 min of CPB onset, immediately post-CPB completion, and immediately post-MUF. Indices of hemolysis (SGOT, LDH), production of inflammatory mediators (interleukin (IL)-8, tumor necrosis factor-alpha (TNFα)), and physiologic parameters of inflammation were measured. The overall blood requirement was significantly less in the mini-circuit compared to conventional CPB (47.0 ± 5.8 ml vs 314.2 ± 31.6 ml; p < 0.0001). The only significant blood requirement in the mini-circuit was to replace the volume removed for samples. During the study, mean arterial pressure (MAP) ( p = 0.004), static pulmonary compliance ( p = 0.04), platelets ( p = 0.0003), and white blood cells ( p = 0.003) significantly decreased across the groups. Lung water content ( p = 0.02), TNFα levels ( p = 0.05), and SGOT ( p = 0.009) increased significantly during the study, across the groups. Among all parameters tested, except for blood requirement and hematocrit post-CPB, there were no significant differences between the two circuits. VAVD makes asanguineous prime in neonates feasible. When used in this study to miniaturize a conventional-CPB circuit, VAVD with a reconfigured neonatal CPB console and circuit resulted in no detrimental effects, and allowed for markedly decreased priming volumes and blood transfusion requirements.
Title: Mini-circuit cardiopulmonary bypass with vacuum assisted venous drainage: feasibility of an asanguineous prime in the neonate
Description:
Conventional cardiopulmonary bypass (CPB) in neonates results in increased transfusion requirements and hemodilution.
There has been little advancement in CPB for the neonatal population.
There is evidence that increased priming volumes and blood product transfusion enhances inflammatory response to CPB and increases myocardial and pulmonary dysfunction.
We have devised a miniaturized CPB circuit that utilizes vacuum-assisted venous drainage (VAVD) in an effort to decrease priming volume and avoid transfusion requirements.
The purpose of this study was to evaluate the safety and efficacy of this miniaturized CPB system and determine the feasibility of an asanguineous prime.
Ten 1-week-old piglets were randomized to five mini- and five conventional CPB pump circuits.
Subjects were supported with CPB at 100 ml/kg/min, cooled to 28°C, exposed to 10 min aortic crossclamp with cardioplegic arrest, rewarmed to 37°C, weaned from bypass, and subjected to modified-ultrafiltration (MUF) for approximately 10 min.
This method was chosen to simulate a situation with all the elements of clinical CPB.
Blood transfusion trigger was a hematocrit <15 on CPB.
Serum samples were obtained pre-CPB, at 15 min of CPB onset, immediately post-CPB completion, and immediately post-MUF.
Indices of hemolysis (SGOT, LDH), production of inflammatory mediators (interleukin (IL)-8, tumor necrosis factor-alpha (TNFα)), and physiologic parameters of inflammation were measured.
The overall blood requirement was significantly less in the mini-circuit compared to conventional CPB (47.
0 ± 5.
8 ml vs 314.
2 ± 31.
6 ml; p < 0.
0001).
The only significant blood requirement in the mini-circuit was to replace the volume removed for samples.
During the study, mean arterial pressure (MAP) ( p = 0.
004), static pulmonary compliance ( p = 0.
04), platelets ( p = 0.
0003), and white blood cells ( p = 0.
003) significantly decreased across the groups.
Lung water content ( p = 0.
02), TNFα levels ( p = 0.
05), and SGOT ( p = 0.
009) increased significantly during the study, across the groups.
Among all parameters tested, except for blood requirement and hematocrit post-CPB, there were no significant differences between the two circuits.
VAVD makes asanguineous prime in neonates feasible.
When used in this study to miniaturize a conventional-CPB circuit, VAVD with a reconfigured neonatal CPB console and circuit resulted in no detrimental effects, and allowed for markedly decreased priming volumes and blood transfusion requirements.

Related Results

Blunt Chest Trauma and Chylothorax: A Systematic Review
Blunt Chest Trauma and Chylothorax: A Systematic Review
Abstract Introduction: Although traumatic chylothorax is predominantly associated with penetrating injuries, instances following blunt trauma, as a rare and challenging condition, ...
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
EnvironmentalSurveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) This comprehensive protocol suite enables systematic environmental surveillance for avian influenza...
Endovascular Transvenous versus Open Femoropopliteal Bypass
Endovascular Transvenous versus Open Femoropopliteal Bypass
Background and Objectives: Lower extremity arterial disease is one of the most prevalent manifestations of atherosclerosis. The results from numerous studies regarding the best rev...
STRUKTUR DRAMATIK PERTUNJUKAN DRAMA KLASIK SANGGAR TEATER MINI LAKON DEWA RUCI KAJIAN BENTUK DAN FUNGSI
STRUKTUR DRAMATIK PERTUNJUKAN DRAMA KLASIK SANGGAR TEATER MINI LAKON DEWA RUCI KAJIAN BENTUK DAN FUNGSI
ABSTRAK   Pada dasarnya nilai pendidikan karakter mempunyai tiga bagian yang saling bekaitan, yaitu pengetahuan moral, penghayatan moral dan perilaku moral. Oleh karena...
Anticoagulation in Cardiopulmonary Bypass
Anticoagulation in Cardiopulmonary Bypass
Prolonging the activated clotting time (ACT) 2 to 3 times normal is said to provide a “safe” level of anticoagulation during cardiopulmonary bypass. To test this level of anticoagu...
Comparative study between aortic valve replacement through full sternotomy versus mini-sternotomy
Comparative study between aortic valve replacement through full sternotomy versus mini-sternotomy
Background: The superiority of minimally invasive aortic valve replacement (AVR) over the standard approach is the subject of ongoing research. The aim of this study was to compare...
Differences in vacuum suction drainage pressure following mastectomy and its impact on the clinical outcome of breast cancer patients
Differences in vacuum suction drainage pressure following mastectomy and its impact on the clinical outcome of breast cancer patients
Background: Vacuum suction drainage is an obligatory practice for long past following mastectomy for breast cancer. But recent studies are showing that the pressure of the vacuum s...

Back to Top