Fluid and Solid Mechanics in Stented Arteries
View through CrossRef
Abstract
Vascular stents are being used in increasing numbers to correct arterial flow limiting disorders. Although data exist on the in vivo performance of these devices, comparatively little is known about how these devices affect the arterial mechanical environment. The long-term success of stenting is very likely affected by the blood flow patterns and artery wall stresses that follow stent placement.
This study was undertaken to identify the mechanical environment in stented arteries and to propose a stent design that minimizes the “mechanical trauma” of stent implantation. A series of pulsatile flow visualization experiments were performed with a Johnson & Johnson Palmaz/Schatz stent inserted in a straight compliant tube. A flow loop was constructed that was capable of applying a physiologic pulsatile flow indicative of conditions in the femoral or coronary arteries. The results showed that this stent design creates complex flow patterns including large-scale vortices, and that these patterns are caused by the compliance mismatch between the stented vessel and adjacent unstented vessel.
The compliance mismatch also would be expected to create abnormal stress concentrations in the artery wall near the ends of the stent. A simplified model of an artery was constructed to estimate the stress in the artery wall. The stent/artery structure was assumed to be an axisymmetric thin shell made of a linear elastic, orthotropic material that undergoes small deformations. The stent diameter was assumed to be equal to the artery systolic diameter. The stresses were estimated under diastolic pressure, when the deflection of the artery due to the presence of the stent is greatest. It was found that the circumferential and axial stresses at the ends of the stent were 2 to 3 times higher than the stress far from the stent.
Based on these mechanical studies, a new stent design was proposed that provides a smooth transition in compliance at the proximal and distal ends of the tube. Flow visualization studies with this new stent indicate that the flow disturbances are greatly reduced with this new design. The transition zone was incorporated into the shell model to estimate the artery wall stresses induced by this new model. It was found that the stress concentration was reduced by as much as 36% from the rigid stent value. These studies demonstrate the potential complexity of flow in stented arteries and provide some recommendations for optimizing stent design from a biomechanical point of view.
This work was supported in part by a Linkage Grant from NATO.
American Society of Mechanical Engineers
Title: Fluid and Solid Mechanics in Stented Arteries
Description:
Abstract
Vascular stents are being used in increasing numbers to correct arterial flow limiting disorders.
Although data exist on the in vivo performance of these devices, comparatively little is known about how these devices affect the arterial mechanical environment.
The long-term success of stenting is very likely affected by the blood flow patterns and artery wall stresses that follow stent placement.
This study was undertaken to identify the mechanical environment in stented arteries and to propose a stent design that minimizes the “mechanical trauma” of stent implantation.
A series of pulsatile flow visualization experiments were performed with a Johnson & Johnson Palmaz/Schatz stent inserted in a straight compliant tube.
A flow loop was constructed that was capable of applying a physiologic pulsatile flow indicative of conditions in the femoral or coronary arteries.
The results showed that this stent design creates complex flow patterns including large-scale vortices, and that these patterns are caused by the compliance mismatch between the stented vessel and adjacent unstented vessel.
The compliance mismatch also would be expected to create abnormal stress concentrations in the artery wall near the ends of the stent.
A simplified model of an artery was constructed to estimate the stress in the artery wall.
The stent/artery structure was assumed to be an axisymmetric thin shell made of a linear elastic, orthotropic material that undergoes small deformations.
The stent diameter was assumed to be equal to the artery systolic diameter.
The stresses were estimated under diastolic pressure, when the deflection of the artery due to the presence of the stent is greatest.
It was found that the circumferential and axial stresses at the ends of the stent were 2 to 3 times higher than the stress far from the stent.
Based on these mechanical studies, a new stent design was proposed that provides a smooth transition in compliance at the proximal and distal ends of the tube.
Flow visualization studies with this new stent indicate that the flow disturbances are greatly reduced with this new design.
The transition zone was incorporated into the shell model to estimate the artery wall stresses induced by this new model.
It was found that the stress concentration was reduced by as much as 36% from the rigid stent value.
These studies demonstrate the potential complexity of flow in stented arteries and provide some recommendations for optimizing stent design from a biomechanical point of view.
This work was supported in part by a Linkage Grant from NATO.
Related Results
Comparison of Hospital Stay in Stented Versus Non-Stented Pyeloplasty in Children with Pelviureteric Junction Obstruction
Comparison of Hospital Stay in Stented Versus Non-Stented Pyeloplasty in Children with Pelviureteric Junction Obstruction
Objective: To compare the average hospital, stay between stented vs non-stented pyeloplasty in children with hydronephrosis due to pelviureteric junction obstruction at our setup. ...
COMPARISON OF STENTED VERSUS NON-STENTED PATIENTS OF URETERIC CALCULI AFTER INTRACORPOREAL LITHOTRIPSY
COMPARISON OF STENTED VERSUS NON-STENTED PATIENTS OF URETERIC CALCULI AFTER INTRACORPOREAL LITHOTRIPSY
Background: Urolithiasis is a prevalent urological condition resulting from crystalline deposits in the urinary tract, often manifesting as acute flank pain and lower urinary tract...
e0517 Forearm arteries with ultrasound for percutaneous coronary procedures
e0517 Forearm arteries with ultrasound for percutaneous coronary procedures
Background
The radial artery has become a widely used approach for coronary angiography and intervention in patients, and the ulnar artery is another approach for...
A RANDOMISED CLINICAL TRIAL COMPARING OUTCOME OF STENTED VERSUS NON-STENTED URSL
A RANDOMISED CLINICAL TRIAL COMPARING OUTCOME OF STENTED VERSUS NON-STENTED URSL
Purpose: Ureteroscopic lithotripsy (URSL) is procedure of choice for managing ureteric calculi. However placing ureteric stent after every URSL
is controversial. Routine stenting e...
Response to Endothelin-1 in Arteries from Human Colorectal Tumours: Role of Endothelin Receptors
Response to Endothelin-1 in Arteries from Human Colorectal Tumours: Role of Endothelin Receptors
To examine the reaction of tumour arteries to endothelin-1, we obtained arteries supplying blood flow to colorectal tumours from patients, as well as mesenteric arteries supplying ...
External Stenting of Vein Grafts in Coronary Artery Bypass Grating: Interim Results from a Two Centers Prospective Study
External Stenting of Vein Grafts in Coronary Artery Bypass Grating: Interim Results from a Two Centers Prospective Study
Abstract
Background – previous studies evaluating external stents for saphenous vein grafts (SVG) in CABG were limited to on-pump isolated CABG and single grafting techniqu...
CHANGES IN THE INTENSITY OF UTERINE CIRCULATION IN PREGNANT WOMEN WITH A HISTORY OF CHRONIC ENDOMETRITIS
CHANGES IN THE INTENSITY OF UTERINE CIRCULATION IN PREGNANT WOMEN WITH A HISTORY OF CHRONIC ENDOMETRITIS
Introduction. An essential prerequisite for the normal course of pregnancy is a complete gestational remodeling of the blood circulation, when the spiral arteries are transformed i...
Hybrid continuum-molecular modeling of fluid slip flow
Hybrid continuum-molecular modeling of fluid slip flow
Experiments on fluid systems in micro-/nano-scale solid conveyors have shown a violation of the no-slip assumption that has been adopted by the classical fluid mechanics. To correc...