Corneal densitometry changes after femtosecond laser-assisted intracorneal ring segments implantation in keratoconus
View through CrossRef
Background: Intrastromal corneal ring segments are commonly implanted in the corneas of eyes with mild-to-moderate keratoconus; however, changes in corneal densitometry (CD) after implantation are a matter of debate in the current literature. We evaluated the changes in CD 1 and 3 months after femtosecond laser-assisted Keraring implantation.
Methods: This retrospective, non-comparative, multicenter, case series study included patients with keratoconus who underwent femtosecond laser-assisted implantation of double segments with 90° and 160° arc lengths or two 160° arc length Keraring segments. Demographic and baseline clinical ophthalmic data were recorded. Corneal topography and tomography data acquired using a Pentacam HR Scheimpflug tomography system (Pentacam High Resolution; Oculus, Wetzlar, Germany) with a best-fit sphere were used as a reference surface. Using the Pentacam HR, CD measurements were acquired over a corneal area of 12 mm in total and at four concentric zones (0–2, 2–6, 6–10, and 10–12 mm) of three corneal stromal depths: 120 micrometers of the anterior corneal stromal layer, 60 micrometers of the posterior corneal stromal layer, and the central layer of stroma lying between these two layers.
Results: We included 40 eyes of 40 patients, including 8 (20%) male and 32 (80%) female individuals, with a mean (standard deviation) age of 21.0 (6.4) years. We observed a significant improvement in the topographic values of steep keratometry (K), flat K, maximum K, and corneal astigmatism (all P < 0.05), but not in the mean K, thinnest corneal pachymetry, corneal thickness at the apex, back elevation, or front elevation (all P > 0.05). The mean total anterior, central, and posterior CD differed significantly among the time points, with a significant increase from the preoperative to the 1-month and 3-month postoperative visits (all P < 0.05) and no difference between those of the 1-month and 3-month postoperative visits (all P > 0.05). The mean CD for the anterior layer in the central, paracentral, and mid-peripheral zones, and the central layer in all four zones, differed significantly among time points, with a significant increase from the preoperative to the 1-month and 3-month postoperative visits (all P < 0.05), which remained unchanged from the 1-month to the 3-month postoperative visit (all P < 0.05), except for the central 2–6-mm zone, which decreased significantly from the 1-month to the 3-month postoperative visit (P < 0.001). The CD of the central 10–12-mm zone did not differ significantly in each pairwise comparison (all P > 0.05). In contrast, CD for the posterior layer in the paracentral zone decreased significantly from the preoperative to the 1-month and 3-month postoperative visits but increased, to a lesser extent, from the 1-month to the 3-month postoperative visit (all P < 0.05).
Conclusions: Femtosecond laser-assisted Keraring implantation significantly changes CD, with improvement in most topography parameters. Further longitudinal studies with larger sample sizes are required to verify these preliminary findings.
International Virtual Ophthalmic Research Center
Title: Corneal densitometry changes after femtosecond laser-assisted intracorneal ring segments implantation in keratoconus
Description:
Background: Intrastromal corneal ring segments are commonly implanted in the corneas of eyes with mild-to-moderate keratoconus; however, changes in corneal densitometry (CD) after implantation are a matter of debate in the current literature.
We evaluated the changes in CD 1 and 3 months after femtosecond laser-assisted Keraring implantation.
Methods: This retrospective, non-comparative, multicenter, case series study included patients with keratoconus who underwent femtosecond laser-assisted implantation of double segments with 90° and 160° arc lengths or two 160° arc length Keraring segments.
Demographic and baseline clinical ophthalmic data were recorded.
Corneal topography and tomography data acquired using a Pentacam HR Scheimpflug tomography system (Pentacam High Resolution; Oculus, Wetzlar, Germany) with a best-fit sphere were used as a reference surface.
Using the Pentacam HR, CD measurements were acquired over a corneal area of 12 mm in total and at four concentric zones (0–2, 2–6, 6–10, and 10–12 mm) of three corneal stromal depths: 120 micrometers of the anterior corneal stromal layer, 60 micrometers of the posterior corneal stromal layer, and the central layer of stroma lying between these two layers.
Results: We included 40 eyes of 40 patients, including 8 (20%) male and 32 (80%) female individuals, with a mean (standard deviation) age of 21.
0 (6.
4) years.
We observed a significant improvement in the topographic values of steep keratometry (K), flat K, maximum K, and corneal astigmatism (all P < 0.
05), but not in the mean K, thinnest corneal pachymetry, corneal thickness at the apex, back elevation, or front elevation (all P > 0.
05).
The mean total anterior, central, and posterior CD differed significantly among the time points, with a significant increase from the preoperative to the 1-month and 3-month postoperative visits (all P < 0.
05) and no difference between those of the 1-month and 3-month postoperative visits (all P > 0.
05).
The mean CD for the anterior layer in the central, paracentral, and mid-peripheral zones, and the central layer in all four zones, differed significantly among time points, with a significant increase from the preoperative to the 1-month and 3-month postoperative visits (all P < 0.
05), which remained unchanged from the 1-month to the 3-month postoperative visit (all P < 0.
05), except for the central 2–6-mm zone, which decreased significantly from the 1-month to the 3-month postoperative visit (P < 0.
001).
The CD of the central 10–12-mm zone did not differ significantly in each pairwise comparison (all P > 0.
05).
In contrast, CD for the posterior layer in the paracentral zone decreased significantly from the preoperative to the 1-month and 3-month postoperative visits but increased, to a lesser extent, from the 1-month to the 3-month postoperative visit (all P < 0.
05).
Conclusions: Femtosecond laser-assisted Keraring implantation significantly changes CD, with improvement in most topography parameters.
Further longitudinal studies with larger sample sizes are required to verify these preliminary findings.
Related Results
Corneal densitometry changes in keratoconus patients after intrastromal corneal ring segments implantation
Corneal densitometry changes in keratoconus patients after intrastromal corneal ring segments implantation
BackgroundKeratoconus (KC) is an ectatic corneal disorder characterized by progressive corneal thinning with myopic astigmatism. Intrastromal corneal ring segments (ICRS) have been...
Corneal densitometry analysis in patients diagnosed with keratoconus
Corneal densitometry analysis in patients diagnosed with keratoconus
AbstractPurposeThe purpose of this study is to carry out a comparative study of corneal densitometry (COD) using Pentacam HR in two groups of patients. A control group of normal ey...
Diagnostic value of corneal optical densitometry in keratoconus
Diagnostic value of corneal optical densitometry in keratoconus
Abstract
Purpose
To investigate the clinical significance of the correlation between optical densitometry and both biomechanical and morphological parameters in keratoconu...
Corneal Physiology and Responses to Cross-Linking and Intracorneal Rings in Keratoconus
Corneal Physiology and Responses to Cross-Linking and Intracorneal Rings in Keratoconus
ABSTRACT
Keratoconus is a noninflammatory, usually, bilateral progressive disease. It is a pathology characterized by a progressive thinning and ectasia of the stroma that results ...
<b>Corneal Biomechanical Integrity After Femtosecond Laser-Assisted Versus Manual Incisions in Cataract Surgery: A Systematic Review</b>
<b>Corneal Biomechanical Integrity After Femtosecond Laser-Assisted Versus Manual Incisions in Cataract Surgery: A Systematic Review</b>
Background: The integrity of the clear corneal incision is fundamental to safety, wound sealing, and refractive stability in cataract surgery. Although femtosecond laser-assisted c...
Corneal Scheimpflug Densitometry in Photorefractive Keratectomy Candidates
Corneal Scheimpflug Densitometry in Photorefractive Keratectomy Candidates
Purpose:
To describe the normative data of corneal densitometry in photorefractive Keratectomy (PRK) candidates by using a Scheimpflug camera.
...
Corneal Deformation Response in Patients With keratoconus and in Healthy Subjects
Corneal Deformation Response in Patients With keratoconus and in Healthy Subjects
Abstract
Background
To compare the corneal deformation response between patients with keratoconus and healthy subjects and to identify potential characteristics associated...
Early changes in corneal densitometry after FS-LASIK combined with accelerated corneal cross-linking for correction of high myopia
Early changes in corneal densitometry after FS-LASIK combined with accelerated corneal cross-linking for correction of high myopia
AIM: To observe the effects of femtosecond laser-assisted excimer laser in situ keratomileusis combined with accelerated corneal cross-linking (FS-LASIK Xtra) on corneal densitomet...