SU‐FF‐T‐353: Influence of a Novel Transmission Detector On 6 MV X‐Ray Beam Characteristics

View through CrossRef
Purpose: A new 2‐D ionisation chamber array from IBA Dosimetry has been investigated at our center for its use as a dosimetric verification tool. This study evaluates the influence of this detector on 6MV X‐ray beam properties. Materials & Methods: COMPASS is a 2‐D transmission array of 1600 plane parallel ionisation chambers with an active area of 40 × 40 cm2. The transmission characteristics of the chamber and the influence of the detector on the 6MV X‐ray beam parameters in the build‐up and beyond were studied. PTW Markus Model 329 chamber was used for measurements in the build‐up region and diode was used for relative dosimetry beyond dmax. The following relative dose parameters were compared with and without the transmission detector in the path of the beam: central axis PDD; profiles at 1.4 and 10 cm depths; transmission factor at 10 cm. Results: For moderately narrow beam geometric conditions (smaller fields, larger SSDs) the increase in surface dose with COMPASS was small. The surface dose profiles also showed expected behavior within the field i.e., increased doses for COMPASS compared to open fields. Beyond dmax, the differences between open and COMPASS profiles were negligible. The transmission factors were very consistent throughout the range of field sizes and SSDs investigated. The mean value was 0.967 with a standard deviation of 0.002. Comparison to measurements with an acrylic block tray has shown that the insertion of the transmission detector has a similar effect on the radiation beam as a standard block tray. Conclusions: With the exception of the transmission factor, the only other influence of the transmission detector on a 6MV beam is an increase in dose in the buildup region. Our measurements show that the differences between open and COMPASS fields were found to be insignificant beyond dmax. Research partially sponsored by IBA Dosimetry.
Title: SU‐FF‐T‐353: Influence of a Novel Transmission Detector On 6 MV X‐Ray Beam Characteristics
Description:
Purpose: A new 2‐D ionisation chamber array from IBA Dosimetry has been investigated at our center for its use as a dosimetric verification tool.
This study evaluates the influence of this detector on 6MV X‐ray beam properties.
Materials & Methods: COMPASS is a 2‐D transmission array of 1600 plane parallel ionisation chambers with an active area of 40 × 40 cm2.
The transmission characteristics of the chamber and the influence of the detector on the 6MV X‐ray beam parameters in the build‐up and beyond were studied.
PTW Markus Model 329 chamber was used for measurements in the build‐up region and diode was used for relative dosimetry beyond dmax.
The following relative dose parameters were compared with and without the transmission detector in the path of the beam: central axis PDD; profiles at 1.
4 and 10 cm depths; transmission factor at 10 cm.
Results: For moderately narrow beam geometric conditions (smaller fields, larger SSDs) the increase in surface dose with COMPASS was small.
The surface dose profiles also showed expected behavior within the field i.
e.
, increased doses for COMPASS compared to open fields.
Beyond dmax, the differences between open and COMPASS profiles were negligible.
The transmission factors were very consistent throughout the range of field sizes and SSDs investigated.
The mean value was 0.
967 with a standard deviation of 0.
002.
Comparison to measurements with an acrylic block tray has shown that the insertion of the transmission detector has a similar effect on the radiation beam as a standard block tray.
Conclusions: With the exception of the transmission factor, the only other influence of the transmission detector on a 6MV beam is an increase in dose in the buildup region.
Our measurements show that the differences between open and COMPASS fields were found to be insignificant beyond dmax.
Research partially sponsored by IBA Dosimetry.

Related Results

Conceptual design report of the MPD Cosmic Ray Detector (MCORD)
Conceptual design report of the MPD Cosmic Ray Detector (MCORD)
Abstract This report presents a concept of constructing a detector dedicated for detection of muons observed during measurements carried out at the MPD (Multi-Pu...
A method of calibrating effective area of focusing X-ray detector by using normal spectrum of Crab pulsar
A method of calibrating effective area of focusing X-ray detector by using normal spectrum of Crab pulsar
X-ray detector is a core component for X-ray astronomical observation and pulsar navigation.The on-orbit observation performance of X-ray detector will change gradually,owing to th...
Characterization of a novel HgCdTe focal plane array for ground and space astronomy through innovative infrared setups
Characterization of a novel HgCdTe focal plane array for ground and space astronomy through innovative infrared setups
(English) Nowadays, mercury-cadmium-telluride (MCT) short-wave infrared (SWIR) detectors are widely used in cutting-edge space missions and ground-based telescopes. They take adva...
Neutron holography simulation based on different sample rotations
Neutron holography simulation based on different sample rotations
Neutron holography is a new imaging technique based on the recording of the interference pattern of two coherent waves emitted by the same source, which allows observing the spatia...
Analisis Numerik Kuat Lentur Balok Support Beam Curve Tile Beton Semi Pracetak Dengan Variasi Panjang Bentangan
Analisis Numerik Kuat Lentur Balok Support Beam Curve Tile Beton Semi Pracetak Dengan Variasi Panjang Bentangan
Semi precast slab is a combination of precast concrete which consist of the support beam and curve tile with the cast in place concrete. During the working process, support beam wi...
Feasibility of energy-resolved angiography
Feasibility of energy-resolved angiography
<p dir="ltr">Cardiovascular disease (CVD) is the leading cause of death globally, with coronary heart disease being one of the most common types of CVD. Two-dimensional x-ray...
Feasibility of energy-resolved angiography
Feasibility of energy-resolved angiography
<p dir="ltr">Cardiovascular disease (CVD) is the leading cause of death globally, with coronary heart disease being one of the most common types of CVD. Two-dimensional x-ray...
Vævning over gruber
Vævning over gruber
Pits, Looms, and Loom Pits? In the archaeological quarterly Skalk (1959, no 2), Mogens Ørsnes described "The Problem of the Hundred Holes," pits "large and small, shallow and deep"...

Back to Top