Neisseria Species Identification Assay for the Confirmation of Neisseria gonorrhoeae -Positive Results of the COBAS Amplicor PCR
View through CrossRef
ABSTRACT
Screening assays for
Neisseria gonorrhoeae
exhibit low positive predictive values, particularly in low-prevalence populations. A new real-time PCR assay that detects and identifies individual
Neisseria
spp. using melt curve analysis was compared to two previously published supplementary assays. NsppID, a 16S rRNA real-time PCR/melt curve assay developed to distinguish
N. gonorrhoeae
from other
Neisseria
spp., was compared to real-time PCR assays targeting genes reportedly specific for
N. gonorrhoeae
, the
cppB
gene and the
porA
pseudogene. A total of 408 clinical specimens (324 female endocervical swabs and 84 male urine or urogenital swab specimens) were screened using the COBAS Amplicor assay for
Chlamydia trachomatis
and
N. gonorrhoeae
(CT/NG) (Roche Diagnostics, Indianapolis, IN) followed by confirmatory testing via real-time PCR. The NsppID assay detected
Neisseria
spp. in 150/181 COBAS-positive specimens (82%), including six dual infections, and identified
N. gonorrhoeae
in 102 (56%) specimens. Sixty-nine of 181 (38%) specimens were positive for
N. gonorrhoeae
by
porA
pseudogene, and 115/181 (64%) were positive for
cppB
. However,
cppB
was also positive in 15% of COBAS-negative specimens, more than either NsppID (4%) or
porA
pseudogene (2%) assays. The
porA
pseudogene assay had the highest specificity for both genders but the lowest sensitivity, especially in female specimens. NsppID had a slightly lower specificity but greater sensitivity and overall accuracy. The least desirable confirmatory assay was
cppB
, due to poor specificity. The NsppID assay is an accurate confirmatory assay for
N. gonorrhoeae
detection. In addition, the NsppID assay can identify the non-
N. gonorrhoeae
species responsible for the majority of false-positive results from the COBAS Amplicor CT/NG assay.
American Society for Microbiology
Title: Neisseria
Species Identification Assay for the Confirmation of
Neisseria gonorrhoeae
-Positive Results of the COBAS Amplicor PCR
Description:
ABSTRACT
Screening assays for
Neisseria gonorrhoeae
exhibit low positive predictive values, particularly in low-prevalence populations.
A new real-time PCR assay that detects and identifies individual
Neisseria
spp.
using melt curve analysis was compared to two previously published supplementary assays.
NsppID, a 16S rRNA real-time PCR/melt curve assay developed to distinguish
N.
gonorrhoeae
from other
Neisseria
spp.
, was compared to real-time PCR assays targeting genes reportedly specific for
N.
gonorrhoeae
, the
cppB
gene and the
porA
pseudogene.
A total of 408 clinical specimens (324 female endocervical swabs and 84 male urine or urogenital swab specimens) were screened using the COBAS Amplicor assay for
Chlamydia trachomatis
and
N.
gonorrhoeae
(CT/NG) (Roche Diagnostics, Indianapolis, IN) followed by confirmatory testing via real-time PCR.
The NsppID assay detected
Neisseria
spp.
in 150/181 COBAS-positive specimens (82%), including six dual infections, and identified
N.
gonorrhoeae
in 102 (56%) specimens.
Sixty-nine of 181 (38%) specimens were positive for
N.
gonorrhoeae
by
porA
pseudogene, and 115/181 (64%) were positive for
cppB
.
However,
cppB
was also positive in 15% of COBAS-negative specimens, more than either NsppID (4%) or
porA
pseudogene (2%) assays.
The
porA
pseudogene assay had the highest specificity for both genders but the lowest sensitivity, especially in female specimens.
NsppID had a slightly lower specificity but greater sensitivity and overall accuracy.
The least desirable confirmatory assay was
cppB
, due to poor specificity.
The NsppID assay is an accurate confirmatory assay for
N.
gonorrhoeae
detection.
In addition, the NsppID assay can identify the non-
N.
gonorrhoeae
species responsible for the majority of false-positive results from the COBAS Amplicor CT/NG assay.
Related Results
B-028 Result Precision on the New cobas pro integrated solutions cobas ISE neo and cobas c 703 Analytical Units Under Routine-Like Conditions Across Two Sites in Europe
B-028 Result Precision on the New cobas pro integrated solutions cobas ISE neo and cobas c 703 Analytical Units Under Routine-Like Conditions Across Two Sites in Europe
Abstract
Background
The precision of the new cobas® pro integrated solutions cobas® ISE neo and cobas c 703 analytical units (Ro...
Equivalent performance of the cobas® Cdiff test for use on the cobas® Liat® system and the cobas® 4800 system
Equivalent performance of the cobas® Cdiff test for use on the cobas® Liat® system and the cobas® 4800 system
Clostridium difficile infection is a significant health burden, and innovative solutions are needed to shorten time to diagnosis and improve infection control. We evaluated the per...
Development and Validation of a Rotor-Gene Real-Time PCR Assay for Detection, Identification, and Quantification of
Chlamydia trachomatis
in a Single Reaction
Development and Validation of a Rotor-Gene Real-Time PCR Assay for Detection, Identification, and Quantification of
Chlamydia trachomatis
in a Single Reaction
ABSTRACT
A multitarget real-time PCR (MRT-PCR) for detection of
Chlamydia trachomatis
DNA was developed and validated. There were thr...
Molecular Epidemiology and Antimicrobial Resistance Gene Distribution in Neisseria gonorrhoeae Isolates from Cervicitis Cases
Molecular Epidemiology and Antimicrobial Resistance Gene Distribution in Neisseria gonorrhoeae Isolates from Cervicitis Cases
Background: Cervicitis is frequently caused by sexually transmitted infections (STIs), including Neisseria gonorrhoeae, Chlamydia trachomatis and Trichomonas vaginalis. The emergen...
Comparison of the cobas Human Immunodeficiency Virus 1 (HIV-1) Test Using the cobas 4800 System With COBAS AmpliPrep/COBAS TaqMan HIV-1 Test and Abbott RealTime HIV-1 Assay and Performance Evaluation of cobas HIV-1
Comparison of the cobas Human Immunodeficiency Virus 1 (HIV-1) Test Using the cobas 4800 System With COBAS AmpliPrep/COBAS TaqMan HIV-1 Test and Abbott RealTime HIV-1 Assay and Performance Evaluation of cobas HIV-1
Abstract
Objectives
To compare quantified human immunodeficiency virus type 1 (HIV-1) viral load quantified using the cobas HIV-...
DNA Transformation and Type IV Pili in Neisseria gonorrhoeae
DNA Transformation and Type IV Pili in Neisseria gonorrhoeae
Transformation de l'ADN et pili de type IV chez Neisseria gonorrhoeae
La résistance aux antibiotiques, évidente chez des souches telles que Neisseria gonorrhoeae, e...
Cyclization increases bactericidal activity of arginine-rich cationic cell-penetrating peptide for
Neisseria gonorrhoeae
Cyclization increases bactericidal activity of arginine-rich cationic cell-penetrating peptide for
Neisseria gonorrhoeae
ABSTRACT
We previously reported that a linear cationic 12-amino acid cell-penetrating peptide (CPP) was bactericidal for
...
Performance evaluation of cobas HBV real-time PCR assay on Roche cobas 4800 System in comparison with COBAS AmpliPrep/COBAS TaqMan HBV Test
Performance evaluation of cobas HBV real-time PCR assay on Roche cobas 4800 System in comparison with COBAS AmpliPrep/COBAS TaqMan HBV Test
Abstract
Background:
Hepatitis B virus (HBV) nucleic acid amplification testing (NAAT) is important for the diagnosis and management of HBV infection. We evaluated the analytical ...