Evaluation of modified window-based scatter compensation in quantitative 177Lu-SPECT for a ring-configured CZT SPECT-CT

View through CrossRef
Abstract Background : The aim was to evaluate bias and precision for 177 Lu-activity-concentration estimation for window-based scatter compensation in a ring-configured CZT gamma camera. This paper extends a previous study by applying modified scatter compensation (MSC) methods. Whilst the original compensation addressed tailing in the 208 keV peak only, separating primary and scatter photons within the scatter window, the current adaption extents the concept to the 113 keV peak. In the modified version, the estimated primary signal in the scatter window can optionally also be added to measured projections to boost the signal-to-noise ratio (MSC + P). Methods : No additional measurements were acquired, and only previously collected listmode-data for a uniform cylindrical phantom, an image quality NEMA phantom and an anthropomorphic phantom were reframed to accommodate the updated window settings recommended for MSC and MSC + P. Reconstructions were performed using OS-EM with two to 30 iterations (10 subsets) with compensation for attenuation, scatter (DEW/TEW, MSC and MSC + P), distance-dependent spatial resolution, and penetration at 208 keV. Volume-of-interest following the physical sphere size were defined and activity-concentration estimates for each sphere, a liver region and for total activity in the phantoms were assessed in terms of bias and precision for short (10 min) and long timeframes (60 min). Results : Imaging at 208 keV generally result in similar bias and precision and total activity estimates for all scatter-compensation methods. For 113 keV, a slightly netter precision is obtained for MSC and MSC + P but at the cost of larger bias compared with TEW. The major difference is seen for total activity where the MSC and MSC + P manages to decrease bias considerably. The MSC + P yields a decrease in bias of about 10 percentage points when comparing short versus long time frames for activity concentration estimates at 113 keV. Conclusions : Despite the lack of broader improvement in bias and precision when using MSC and MSC + P the findings suggest potential value in more targeted applications. For instance, MSC and MSC + P improves preservation of total activity. Although the findings indicate that using MSC + P is not advisable for estimating high activity concentrations, it may positively influence the estimation of low activity concentrations.
Title: Evaluation of modified window-based scatter compensation in quantitative 177Lu-SPECT for a ring-configured CZT SPECT-CT
Description:
Abstract Background : The aim was to evaluate bias and precision for 177 Lu-activity-concentration estimation for window-based scatter compensation in a ring-configured CZT gamma camera.
This paper extends a previous study by applying modified scatter compensation (MSC) methods.
Whilst the original compensation addressed tailing in the 208 keV peak only, separating primary and scatter photons within the scatter window, the current adaption extents the concept to the 113 keV peak.
In the modified version, the estimated primary signal in the scatter window can optionally also be added to measured projections to boost the signal-to-noise ratio (MSC + P).
Methods : No additional measurements were acquired, and only previously collected listmode-data for a uniform cylindrical phantom, an image quality NEMA phantom and an anthropomorphic phantom were reframed to accommodate the updated window settings recommended for MSC and MSC + P.
Reconstructions were performed using OS-EM with two to 30 iterations (10 subsets) with compensation for attenuation, scatter (DEW/TEW, MSC and MSC + P), distance-dependent spatial resolution, and penetration at 208 keV.
Volume-of-interest following the physical sphere size were defined and activity-concentration estimates for each sphere, a liver region and for total activity in the phantoms were assessed in terms of bias and precision for short (10 min) and long timeframes (60 min).
Results : Imaging at 208 keV generally result in similar bias and precision and total activity estimates for all scatter-compensation methods.
For 113 keV, a slightly netter precision is obtained for MSC and MSC + P but at the cost of larger bias compared with TEW.
The major difference is seen for total activity where the MSC and MSC + P manages to decrease bias considerably.
The MSC + P yields a decrease in bias of about 10 percentage points when comparing short versus long time frames for activity concentration estimates at 113 keV.
Conclusions : Despite the lack of broader improvement in bias and precision when using MSC and MSC + P the findings suggest potential value in more targeted applications.
For instance, MSC and MSC + P improves preservation of total activity.
Although the findings indicate that using MSC + P is not advisable for estimating high activity concentrations, it may positively influence the estimation of low activity concentrations.

Related Results

Study on the Microstructure of Au/CdZnTe Interface by HRTEM
Study on the Microstructure of Au/CdZnTe Interface by HRTEM
As a new type of detector material, the CdZnTe (CZT) crystal has the advantages such as a wide energy gap, a large average atomic number, the strong ability to stop X- ray and high...
Characterization of Extended Defects Observed in Cadmium Zinc Telluride (CZT) Crystal
Characterization of Extended Defects Observed in Cadmium Zinc Telluride (CZT) Crystal
ABSTRACTCadmium Zinc Telluride (CZT) semiconductor crystal properties have been studied extensively with a focus on correlations to their radiation detector performance. The need f...
Absolute quantification in brain SPECT imaging
Absolute quantification in brain SPECT imaging
Certes malalties neurològiques estan associades amb problemes en els sistemes de neurotransmissió. Una aproximació a l'estudi d'aquests sistemes és la tomografia d'emissió SPECT (S...
Emerging Role of 177Lu in Nuclear Oncology: A Brief Review
Emerging Role of 177Lu in Nuclear Oncology: A Brief Review
 With the innovations in nuclear medicine techniques, Lutetium 177 (177Lu) has epitomized as a revolutionary theranostic agent- with both scintigraphic and therapeutic properties. ...
Evaluation of the sensitivity of SPECT/CT systems for 225Ac
Evaluation of the sensitivity of SPECT/CT systems for 225Ac
Radiopharmaceuticals labelled with alpha-emitting radionuclides, particularly 225Ac, are the most promising and interesting radiopharmaceuticals for the cancer treatment. Evaluatio...
SU‐E‐I‐154: Cardiac SPECT/CT: Comparing Slow‐Rotation and Fast‐Rotation CT for Attenuation Correction in Cardiac Imaging
SU‐E‐I‐154: Cardiac SPECT/CT: Comparing Slow‐Rotation and Fast‐Rotation CT for Attenuation Correction in Cardiac Imaging
Purpose: The hybridisation of SPECT with X‐ray CT is expanding the utility of SPECT. In addition to image fusion, CT enables improved attenuation maps (AM) for SPECT. Two CT design...

Back to Top