Objective: To assess prospectively the presence and degree of tricuspid regurgitation (TR) after right ventricular (RV) lead placement utilizing 2-dimensional and doppler echocardiography in a group of patients implanted with Permanent pace-maker or implantable cardioverter defibrillator (ICD). Study Design: Descriptive cross sectional study. Place and Duration of Study: Study was conducted at OPD of AFIC/NIHD Rawalpindi from Jan to Jun 2019. Methodology: One hundred patients after cardiac stimulation system implantation were included in this study. Patients with severe valve disease, heart failure, congenital heart disease, pre-existing pulmonary hypertension and presence of moderate or severe tricuspid regurgitation were excluded. M-mode, 2 –dimensional and doppler echocardiographic studies were performed. Continuous-wave doppler measurements were made from apical four-chamber view in order to obtain maximum tricuspid flow velocities. A complete echocardiographic study was performed following device implantation. Echocardiographic measurements were repeated at 6-months of follow-up period. Those patients who developed tricuspid regurgitation were further assessed by color doppler, taking into account density and contour of the jet on continuous-wave doppler. Tricuspid regurgitation severity was classified into three groups: mild, moderate and severe according to the recommendations for non-invasive evaluation of native valvular regurgitation by the American Society of Echocardiography. The collected data was analysed by using SPSS-23. Results: Out of 100 patients, 67 (67%) were males and 33(33%) were females with the mean age of 65.29 ± 12.02 years. All the patients had normal chambers dimensions before the procedure. 65 (65%) had leads screwed to RV apex while 35 (35%) had through RV septum. 83 (83%) patients received a dual chamber device, while 17 (17%) patients got single chamber (ventricular) device implanted. Following device implantation after 6 months mild TR was noted in 4 (4%), moderate TR in 4 (4%) and severe TR 2 (2%) in cases. Conclusion: Tricuspid incompetence following endocardial leads implantation is not rare and occurred in approximately 10% of our patients. This complication may be preventable, because it is likely due to the interference of the endocardial lead with the TV.
Title: TRICUSPID VALVE REGURGITATION IN PATIENTS UNDERGOING ENDOCARDIAL LEADS PLACEMENT
Description:
Objective: To assess prospectively the presence and degree of tricuspid regurgitation (TR) after right ventricular (RV) lead placement utilizing 2-dimensional and doppler echocardiography in a group of patients implanted with Permanent pace-maker or implantable cardioverter defibrillator (ICD).
Study Design: Descriptive cross sectional study.
Place and Duration of Study: Study was conducted at OPD of AFIC/NIHD Rawalpindi from Jan to Jun 2019.
Methodology: One hundred patients after cardiac stimulation system implantation were included in this study.
Patients with severe valve disease, heart failure, congenital heart disease, pre-existing pulmonary hypertension and presence of moderate or severe tricuspid regurgitation were excluded.
M-mode, 2 –dimensional and doppler echocardiographic studies were performed.
Continuous-wave doppler measurements were made from apical four-chamber view in order to obtain maximum tricuspid flow velocities.
A complete echocardiographic study was performed following device implantation.
Echocardiographic measurements were repeated at 6-months of follow-up period.
Those patients who developed tricuspid regurgitation were further assessed by color doppler, taking into account density and contour of the jet on continuous-wave doppler.
Tricuspid regurgitation severity was classified into three groups: mild, moderate and severe according to the recommendations for non-invasive evaluation of native valvular regurgitation by the American Society of Echocardiography.
The collected data was analysed by using SPSS-23.
Results: Out of 100 patients, 67 (67%) were males and 33(33%) were females with the mean age of 65.
29 ± 12.
02 years.
All the patients had normal chambers dimensions before the procedure.
65 (65%) had leads screwed to RV apex while 35 (35%) had through RV septum.
83 (83%) patients received a dual chamber device, while 17 (17%) patients got single chamber (ventricular) device implanted.
Following device implantation after 6 months mild TR was noted in 4 (4%), moderate TR in 4 (4%) and severe TR 2 (2%) in cases.
Conclusion: Tricuspid incompetence following endocardial leads implantation is not rare and occurred in approximately 10% of our patients.
This complication may be preventable, because it is likely due to the interference of the endocardial lead with the TV.

Related Results

P221 Carcinoid heart disease
P221 Carcinoid heart disease
Abstract Carcinoid heart disease is a rare disease, which develops in 20-50% of patients with carcinoid syndrome and is a main predictor of clinical outcome in those...
Comparison of Valveplasty and Replacement for Surgical Treatment of Infective Tricuspid Valve Endocarditis
Comparison of Valveplasty and Replacement for Surgical Treatment of Infective Tricuspid Valve Endocarditis
Abstract Background In recent years, due to the increase in intravenous drug injection and intracardiac and vascular interventional treatments among drug users, infective ...
Comparison of Heart Valve Circumference Examined Before and After 10% Formalin Fixation
Comparison of Heart Valve Circumference Examined Before and After 10% Formalin Fixation
Objective: To compare the heart valve circumference before and after 10% formalin fixation. Materials and Methods: The study analyzed 63 Thai human cadaveric hearts. Each heart val...
P1720 Hammock mitral valve, a challenging echocardiographic diagnosis
P1720 Hammock mitral valve, a challenging echocardiographic diagnosis
Abstract We report a 43 year-old female with a past TTE echocardiography of rheumatic valve disease performed in her district h...
The medtronic melody® transcatheter pulmonary valve implanted at 24‐mm diameter—it works
The medtronic melody® transcatheter pulmonary valve implanted at 24‐mm diameter—it works
ObjectivesWe report the Melody valve implanted and/or expanded to 24‐mm diameter.BackgroundThe Medtronic Melody valve has been implanted up to 22 mm in the pulmonary position for o...

Back to Top