A-117 Performance Evaluation of five Sentinel Diagnostics’ Assays on Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer

View through CrossRef
Abstract Objective The aim of this study is to evaluate the analytical performances of the following Sentinel Diagnostics’ assays: ACE Liquid, Lp(a) Ultra, Pancreatic Amylase, Lipase NG, Albumin BCP on the Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer. Methods Performance evaluation included Limit of Blank (LoB), Limit of Detection (LoD), Limit of Quantitation (LoQ), linearity, intra assay precision, inter assay precision, calibration stability, on board reagent stability and instrument correlation, following the current CLSI guidelines protocols. Analytical Measuring Range (AMR) was defined through linearity and sensitivity. Results Summary of the analytical performances for all the assays tested is reported in table below. Conclusions All assays showed good results for the parameters tested on the Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer making these kits very suitable for a routine measurement of these analytes.
Title: A-117 Performance Evaluation of five Sentinel Diagnostics’ Assays on Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer
Description:
Abstract Objective The aim of this study is to evaluate the analytical performances of the following Sentinel Diagnostics’ assays: ACE Liquid, Lp(a) Ultra, Pancreatic Amylase, Lipase NG, Albumin BCP on the Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer.
Methods Performance evaluation included Limit of Blank (LoB), Limit of Detection (LoD), Limit of Quantitation (LoQ), linearity, intra assay precision, inter assay precision, calibration stability, on board reagent stability and instrument correlation, following the current CLSI guidelines protocols.
Analytical Measuring Range (AMR) was defined through linearity and sensitivity.
Results Summary of the analytical performances for all the assays tested is reported in table below.
Conclusions All assays showed good results for the parameters tested on the Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer making these kits very suitable for a routine measurement of these analytes.

Related Results

Coulter Method for the detection of Physiological parameters v1
Coulter Method for the detection of Physiological parameters v1
Instrument A Coulter counter [1] [2] is an apparatus for counting and sizing particles suspended in electrolytes. It is used for cells, bacteria, prokaryotic cells and virus parti...
Multi-Resolution Ocean Color roducts to support the Copernicus Marine High-Resolution Coastal Service 
Multi-Resolution Ocean Color roducts to support the Copernicus Marine High-Resolution Coastal Service 
High-quality satellite-based ocean colour products can provide valuable support and insights in the management and monitoring of coastal ecosystems. Today’s availability ...
Blunt Chest Trauma and Chylothorax: A Systematic Review
Blunt Chest Trauma and Chylothorax: A Systematic Review
Abstract Introduction: Although traumatic chylothorax is predominantly associated with penetrating injuries, instances following blunt trauma, as a rare and challenging condition, ...
A-118 Sigma Metric Performance of Chemistry Assays on 4 Instrument Systems
A-118 Sigma Metric Performance of Chemistry Assays on 4 Instrument Systems
Abstract Background Chemistry panels play a crucial role in determining a person’s general health and well-being. These tests ev...
7 th International Symposium on Enabling Technologies for Life Sciences (ETP)
7 th International Symposium on Enabling Technologies for Life Sciences (ETP)
The seventh in the series of ETP Symposia (see Rapid Communications in Mass Spectrometry 2012, 26 , ...

Back to Top