Evaluation of a Point-of-Care System for Quantitative Determination of Troponin T and Myoglobin
View through CrossRef
Abstract
We present the results of a multicenter evaluation of a new point-of-care system (Cardiac Reader) for the quantitative determination of cardiac troponin T (CARDIAC T Quantitative test) and myoglobin (CARDIAC M test) in whole blood samples.
The Cardiac Reader is a CCD camera that optically reads the immunochemical test strips. The measuring range is 0.1 to 3 μg/l for CARDIAC T Quantitative and 30 to 700 μg/l for CARDIAC M. Both tests are calibrated by the manufacturer. The reaction times of the tests are 12 or 8 minutes, respectively.
Method comparisons were performed with 281 heparinized blood samples from patients with suspected acute coronary syndromes. The results obtained with CARDIAC T Quantitative showed a good agreement compared with cardiac troponin T ELISA (r = 0.89; y = 0.93x + 0.02). The method comparison between CARDIAC M and Tina-quant Myoglobin also showed a good agreement between both assays (r = 0.98; y = 0.92x + 1.6). Test lot-to-lot comparisons yielded differences of 2% and 6% for CARDIACT Quantitative and of 0 to 11% for CARDIACM.
The within-run imprecision with blood samples and control materials was acceptable for CARDIAC T Quantitative (CV 10 to 15%) and good for CARDIAC M (CV 5 to 10%). The between-instrument CV was below 7% for CARDIACT Quantitative and below 5% for CARDIACM. The cross-reactivity of CARDIAC T Quantitative with skeletal troponin T was approximately 0.003%. No significant analytical interference was detected for any of the assays in investigations with biotin (up to 100 μg/l), hemoglobin (up to 0.125 mmol/l), hematocrit (26 to 52%), bilirubin (up to 340 μmol/l), triglycerides (up to 5.0 mmol/l), and 18 standard drugs.
With the Cardiac Reader reliable quantitative results can be easily obtained for both cardiac markers. The system is, therefore, particularly suitable for use in emergency rooms, coronary care units and small hospitals.
Title: Evaluation of a Point-of-Care System for Quantitative Determination of Troponin T and Myoglobin
Description:
Abstract
We present the results of a multicenter evaluation of a new point-of-care system (Cardiac Reader) for the quantitative determination of cardiac troponin T (CARDIAC T Quantitative test) and myoglobin (CARDIAC M test) in whole blood samples.
The Cardiac Reader is a CCD camera that optically reads the immunochemical test strips.
The measuring range is 0.
1 to 3 μg/l for CARDIAC T Quantitative and 30 to 700 μg/l for CARDIAC M.
Both tests are calibrated by the manufacturer.
The reaction times of the tests are 12 or 8 minutes, respectively.
Method comparisons were performed with 281 heparinized blood samples from patients with suspected acute coronary syndromes.
The results obtained with CARDIAC T Quantitative showed a good agreement compared with cardiac troponin T ELISA (r = 0.
89; y = 0.
93x + 0.
02).
The method comparison between CARDIAC M and Tina-quant Myoglobin also showed a good agreement between both assays (r = 0.
98; y = 0.
92x + 1.
6).
Test lot-to-lot comparisons yielded differences of 2% and 6% for CARDIACT Quantitative and of 0 to 11% for CARDIACM.
The within-run imprecision with blood samples and control materials was acceptable for CARDIAC T Quantitative (CV 10 to 15%) and good for CARDIAC M (CV 5 to 10%).
The between-instrument CV was below 7% for CARDIACT Quantitative and below 5% for CARDIACM.
The cross-reactivity of CARDIAC T Quantitative with skeletal troponin T was approximately 0.
003%.
No significant analytical interference was detected for any of the assays in investigations with biotin (up to 100 μg/l), hemoglobin (up to 0.
125 mmol/l), hematocrit (26 to 52%), bilirubin (up to 340 μmol/l), triglycerides (up to 5.
0 mmol/l), and 18 standard drugs.
With the Cardiac Reader reliable quantitative results can be easily obtained for both cardiac markers.
The system is, therefore, particularly suitable for use in emergency rooms, coronary care units and small hospitals.
Related Results
Detectable troponin below the 99th percentile predicts survival in patients with cardiovascular disease
Detectable troponin below the 99th percentile predicts survival in patients with cardiovascular disease
Abstract
Background
Patients with cardiovascular disease and elevated troponin above the 99th percentile of the upper reference ...
Additional PKA phosphorylation sites in human cardiac troponin I
Additional PKA phosphorylation sites in human cardiac troponin I
We used mass spectrometry to monitor cAMP‐dependent protein kinase catalysed phosphorylation of human cardiac troponin I in vitro. Phosphorylation of isolated troponin I by cAMP‐de...
Phosphorylation of troponin and the effects of interactions between the components of the complex
Phosphorylation of troponin and the effects of interactions between the components of the complex
1. The troponin complex from skeletal muscle contains approximately 1 mol of phosphate/80000g of complex, covalently bound to the troponin T component. 2. On prolonged incubation o...
Structural dynamics of the intrinsically disordered linker region of cardiac troponin T
Structural dynamics of the intrinsically disordered linker region of cardiac troponin T
ABSTRACT
The cardiac troponin complex, composed of troponins I, T, and C, plays a central role in regulating the calcium-dependent interactions between myosin and t...
Organisatie van geestelijke gezondheidszorg voor mensen met een ernstige en persisterende mentale aandoening
Organisatie van geestelijke gezondheidszorg voor mensen met een ernstige en persisterende mentale aandoening
1 INTRODUCTION AND RESEARCH QUESTIONS 5 -- 2 GENERAL BACKGROUND: DEFINITIONS AND SCOPE OF THE STUDY 7 -- 2.1 CHRONIC AND COMPLEX MENTAL DISORDERS: DEFINITIONS AND SCOPE OF THE -- S...
Frequency and outcome of elevated troponin level in patients presenting with acute ischemic stroke at a tertiary care hospital, Karachi
Frequency and outcome of elevated troponin level in patients presenting with acute ischemic stroke at a tertiary care hospital, Karachi
Background: Heart troponin has been observed to be raised in patients with acute ischemic stroke (AIS); its prognosis is controversial. The aim of this study was to review the inci...
Effect of <i>Newbouldia Leavis</i> Aqueous Extract on Cardiac Troponin, Creatinine Kinase, Myoglobin and Lactate Dehydrogenase in Albino Rats Treated with Diclofenac
Effect of <i>Newbouldia Leavis</i> Aqueous Extract on Cardiac Troponin, Creatinine Kinase, Myoglobin and Lactate Dehydrogenase in Albino Rats Treated with Diclofenac
Diclofenac is a commonly used non-steroidal anti-inflammatory drug (NSAID) associated with adverse cardiovascular effects, including myocardial injury marked by elevated levels of ...
Serum Myoglobin After Cardiac Surgery Predicts Postoperative Cardiogenic Shock Requiring Mechanical Circulatory Support Within 14 Days
Serum Myoglobin After Cardiac Surgery Predicts Postoperative Cardiogenic Shock Requiring Mechanical Circulatory Support Within 14 Days
Background:
Cardiogenic shock requiring mechanical circulatory support is a life-threatening complication of cardiac surgery with cardiopulmonary bypass (CPB). ...