Analysis of the causes of discrepancies in troponin I concentrations as measured by ARCHITECT High-Sensitive Troponin I ST and STACIA CLEIA cTnI

View through CrossRef
Background Recently developed reagents for the highly sensitive measurement of cardiac troponin I are useful for early diagnosis of acute coronary syndrome. However, differences in measured values between these new reagents and previously used reagents have not been well studied. In this study, we aimed to compare the values between ARCHITECT High-Sensitive Troponin I ST (newly developed reagents), ARCHITECT Troponin I ST and STACIA CLEIA cardiac troponin I (two previously developed reagent kits). Methods Gel filtration high-performance liquid chromatography was used to analyse the causes of differences in measured values. Results The measured values differed between ARCHITECT High-Sensitive Troponin I ST and STACIA CLEIA cardiac troponin I reagents (r = 0.82). Cross-reactivity tests using plasma with added skeletal-muscle troponin I resulted in higher reactivity (2.17–3.03%) for the STACIA CLEIA cardiac troponin I reagents compared with that for the ARCHITECT High-Sensitive Troponin I ST reagents (less than 0.014%). In addition, analysis of three representative samples using gel filtration high-performance liquid chromatography revealed reagent-specific differences in the reactivity against each cardiac troponin I complex; this could explain the differences in values observed for some of the samples. Conclusion The newly developed ARCHITECT High-Sensitive Troponin I ST reagents were not affected by the presence of skeletal-muscle troponin I in the blood and may be useful for routine examinations.
Title: Analysis of the causes of discrepancies in troponin I concentrations as measured by ARCHITECT High-Sensitive Troponin I ST and STACIA CLEIA cTnI
Description:
Background Recently developed reagents for the highly sensitive measurement of cardiac troponin I are useful for early diagnosis of acute coronary syndrome.
However, differences in measured values between these new reagents and previously used reagents have not been well studied.
In this study, we aimed to compare the values between ARCHITECT High-Sensitive Troponin I ST (newly developed reagents), ARCHITECT Troponin I ST and STACIA CLEIA cardiac troponin I (two previously developed reagent kits).
Methods Gel filtration high-performance liquid chromatography was used to analyse the causes of differences in measured values.
Results The measured values differed between ARCHITECT High-Sensitive Troponin I ST and STACIA CLEIA cardiac troponin I reagents (r = 0.
82).
Cross-reactivity tests using plasma with added skeletal-muscle troponin I resulted in higher reactivity (2.
17–3.
03%) for the STACIA CLEIA cardiac troponin I reagents compared with that for the ARCHITECT High-Sensitive Troponin I ST reagents (less than 0.
014%).
In addition, analysis of three representative samples using gel filtration high-performance liquid chromatography revealed reagent-specific differences in the reactivity against each cardiac troponin I complex; this could explain the differences in values observed for some of the samples.
Conclusion The newly developed ARCHITECT High-Sensitive Troponin I ST reagents were not affected by the presence of skeletal-muscle troponin I in the blood and may be useful for routine examinations.

Related Results

Possible effects and changes of myocardial injury markers in percutaneous coronary intervention
Possible effects and changes of myocardial injury markers in percutaneous coronary intervention
Objective Percutaneous Coronary Intervention (PCI) is a technique of cardiac intervention treatment in revascularisation of coronary vessel. The use of PCI in pat...
THE REASEARCH INTO THE CLINICAL CHARACTERISTICS OF PERIPARTUM CARDIOMYOPATHY
THE REASEARCH INTO THE CLINICAL CHARACTERISTICS OF PERIPARTUM CARDIOMYOPATHY
Objectives The research is to find the related risk factors that may lead to PPCM through analysing and summarising the general clinical characteristics of PPCM p...
CORRELATION OF CARDIAC TROPONIN I WITH VENTRICULAR ARRHYTHMIA AND LATE PROGNOSIS IN THE PATIENTS WITH CHRONIC HEART FAILURE
CORRELATION OF CARDIAC TROPONIN I WITH VENTRICULAR ARRHYTHMIA AND LATE PROGNOSIS IN THE PATIENTS WITH CHRONIC HEART FAILURE
Objectives TO investigate correlation of cardiac troponin I (cTnI) with ventricular arrhythmia and late prognosis in the patients with chronic heart failure (CHF)...
Determinants of Interindividual Variation in Exercise‐Induced Cardiac Troponin I Levels
Determinants of Interindividual Variation in Exercise‐Induced Cardiac Troponin I Levels
Background Postexercise cardiac troponin levels show considerable interindividual variations. This study aimed to identify the major determinants of this postexercise v...
Abstract 14211: Positive Predictive Value of Troponin for NSTEMI Based on Chief Complaint
Abstract 14211: Positive Predictive Value of Troponin for NSTEMI Based on Chief Complaint
Introduction: Serum troponin is a ubiquitous biomarker used to diagnose acute myocardial infarction (AMI). The positive predictive value (PPV) of high-sensitivity tropo...
Additional PKA phosphorylation sites in human cardiac troponin I
Additional PKA phosphorylation sites in human cardiac troponin I
We used mass spectrometry to monitor cAMP‐dependent protein kinase catalysed phosphorylation of human cardiac troponin I in vitro. Phosphorylation of isolated troponin I by cAMP‐de...
e0447 Combination of conventional biomarkers for risk stratification in non ST elevation acute coronary syndrome
e0447 Combination of conventional biomarkers for risk stratification in non ST elevation acute coronary syndrome
Objective To investigate the predicting value of electrocardiography (ECG) deviation, plasma levels of troponin I (cTnI) and Combination of conventional biomarker...

Back to Top