Microdose EOS Imaging: Just as Good, and Safer
View through CrossRef
Abstract
Purpose:
Children with scoliosis are exposed to significant lifetime radiation doses because of multiple imaging studies as part of their investigation and treatment. EOS scanning has been shown to significantly reduce overall radiation and cancer risk. To reduce radiation further, a microdose setting can be used on the EOS scanner. This study aimed to quantify the reduction in radiation dose and assess image quality compared to the standard EOS settings.
Methods:
A consecutive group of paediatric patients who received both a standard EOS dose and a microdose EOS scan for investigating their scoliosis within a 6-month period. The radiation doses were compared, and cancer risk was estimated using a validated risk calculator. Assessment of image quality was blindly evaluated by six surgeons, using a five-point visual analogue scale (VAS) between standard and microdose studies.
Results:
Twenty-five children who received both studies within a six month period were included, median age 13 (IQR 4). The total doses in the standard EOS group were 0.0895 mSv (additional lifetime cancer risk per study: 0.00197%); compared to the microdose group of 0.0167 mSv (additional lifetime cancer risk per study: 0.00037%) (p <.0001). Image quality VAS was not statistically different between settings (standard: 4.2, microdose 4.0) (p = 0.074).
Conclusion:
Microdose EOS resulted in a five-fold reduction in radiation dose compared to Standard dose EOS and a subsequent 532% decrease in lifetime cancer risk per study. This builds on the fact that Standard dose EOS already carries five times less radiation and cancer risk than plain films. Therefore, compared to traditional radiographs, using a microdose EOS would result in a 40-fold reduction in radiation dose, and a large lifetime reduction in iatrogenic cancer risk. Microdose EOS was found to be both safe and accurate for purpose. A larger study is needed to support these findings.
Springer Science and Business Media LLC
Title: Microdose EOS Imaging: Just as Good, and Safer
Description:
Abstract
Purpose:
Children with scoliosis are exposed to significant lifetime radiation doses because of multiple imaging studies as part of their investigation and treatment.
EOS scanning has been shown to significantly reduce overall radiation and cancer risk.
To reduce radiation further, a microdose setting can be used on the EOS scanner.
This study aimed to quantify the reduction in radiation dose and assess image quality compared to the standard EOS settings.
Methods:
A consecutive group of paediatric patients who received both a standard EOS dose and a microdose EOS scan for investigating their scoliosis within a 6-month period.
The radiation doses were compared, and cancer risk was estimated using a validated risk calculator.
Assessment of image quality was blindly evaluated by six surgeons, using a five-point visual analogue scale (VAS) between standard and microdose studies.
Results:
Twenty-five children who received both studies within a six month period were included, median age 13 (IQR 4).
The total doses in the standard EOS group were 0.
0895 mSv (additional lifetime cancer risk per study: 0.
00197%); compared to the microdose group of 0.
0167 mSv (additional lifetime cancer risk per study: 0.
00037%) (p <.
0001).
Image quality VAS was not statistically different between settings (standard: 4.
2, microdose 4.
0) (p = 0.
074).
Conclusion:
Microdose EOS resulted in a five-fold reduction in radiation dose compared to Standard dose EOS and a subsequent 532% decrease in lifetime cancer risk per study.
This builds on the fact that Standard dose EOS already carries five times less radiation and cancer risk than plain films.
Therefore, compared to traditional radiographs, using a microdose EOS would result in a 40-fold reduction in radiation dose, and a large lifetime reduction in iatrogenic cancer risk.
Microdose EOS was found to be both safe and accurate for purpose.
A larger study is needed to support these findings.
Related Results
Apoptotic response of eosinophils in chronic eosinophilic pneumonia
Apoptotic response of eosinophils in chronic eosinophilic pneumonia
To clarify the pathogenesis of chronic eosinophilic pneumonia (CEP), the apoptosis of eosinophils from bronchoalveolar lavage (BAL-Eos) was compared with that of eosinophils from p...
Impact of Common Anticoagulants on Complete Blood Count Parameters Among Humans
Impact of Common Anticoagulants on Complete Blood Count Parameters Among Humans
Abstract
Introduction
Among the most frequently used anticoagulants in hematological testing are tetra-acetic acid (EDTA), sodium citrate, and sodium heparin. However, there is a n...
Effect of microdose chemical fertilization method on soil chemical properties and productivity of maize (Zea mays L.) in West-Central Senegal
Effect of microdose chemical fertilization method on soil chemical properties and productivity of maize (Zea mays L.) in West-Central Senegal
This study was conducted in West-Central Senegal (Thies region) during two cropping seasons (2018 and 2019) to evaluate the effect of different doses of chemical fertilizer on soil...
Fluid Characterization and EoS Modelling of PVT Experiments
Fluid Characterization and EoS Modelling of PVT Experiments
Abstract
Equations of state (EoS) are used to simulate laboratory pressure, volume and temperature (PVT) experiments. Each PVT experiment mimics a form of phase beha...
Critical Properties Estimation From an Equation of State
Critical Properties Estimation From an Equation of State
Whitson, C.H., Consultant
Member SPE-AIME
Abstract
This paper describes a new method for calculating critical pro...
Effectiveness of Microdosing in Comparison to Broadcasting and Banding Vermicompost Application for Maize (Zea mays L.) Production in Smallholder Farming of Semiarid Tigray Region of Ethiopia
Effectiveness of Microdosing in Comparison to Broadcasting and Banding Vermicompost Application for Maize (Zea mays L.) Production in Smallholder Farming of Semiarid Tigray Region of Ethiopia
Optimizing organic fertilizer recommendations can be cost‐effective and affordable for resource‐poor farmers by increasing crop yield and income. This field experiment was conducte...
The predictive value of eosinophils in children with febrile seizures
The predictive value of eosinophils in children with febrile seizures
Abstract
Objective
To evaluate the predictive value of eosinophil count (EOS) in children with febrile seizures (FS).
Methods
a total of 92 children with FS who visited t...
The Adoption and Applications of Earth Observations and Allied Geoinformation Systems for Environmental Management in Ghana
The Adoption and Applications of Earth Observations and Allied Geoinformation Systems for Environmental Management in Ghana
Earth Observations (EOs) and allied technologies were initially adopted and applied to inform sustainable environmental resource management in Ghana. This paper examines the initia...