A pleural vacuum relief device for pleural drain unit use in the hyperbaric environment
View through CrossRef
Introduction: When a standard water-seal pleural drain unit (PDU) is used under hyperbaric conditions there are scenarios where excessive negative intrapleural pressure (IPP) and/or fluid reflux can be induced, risking significant morbidity. We developed and tested a pleural vacuum relief (PVR) device which automatically manages these risks, whilst allowing more rapid hyperbaric pressure change rates.
Methods: The custom-made PVR device consists of a one-way pressure relief valve connected in line with a sterile micro filter selected for its specific flow capacity. The PVR device is designed for connection to the patient side sampling port of a PDU system, allowing inflow of ambient air whenever negative pressure is present, creating a small, controlled air leak which prevents excessive negative pressure. The hyperbaric performance of a Pleur-Evac A-6000 intercostal drain was assessed with and without this added device by measuring simulated IPP with an electronic pressure monitor connected at the patient end of the PDU. IPP readings were taken at 10, 15, 20 and 30 cmH2O of suction (set on the drain unit) at compression rates of 10, 30, 60, 80, 90 and 180 kPa·min-1 to a pressure of 280 kPa.
Results: At any compression rate of > 10 kPa·min-1, the negative IPP generated by the Pleur-Evac A-6000 alone was excessive and resulted in back flow through the PDU water seal. By adding the PVR device, the generated negative IPP remains within a clinically acceptable range, allowing compression rates of at least 30 kPa·min-1 with suction settings up to -20 cmH2O during all phases of hyperbaric treatment.
Conclusions: The PDU PVR device we have developed works well, minimising attendant workload and automatically avoiding the excessive negative IPPs that can otherwise occur. This device should only be used with suction.
Diving and Hyperbaric Medicine Journal
Title: A pleural vacuum relief device for pleural drain unit use in the hyperbaric environment
Description:
Introduction: When a standard water-seal pleural drain unit (PDU) is used under hyperbaric conditions there are scenarios where excessive negative intrapleural pressure (IPP) and/or fluid reflux can be induced, risking significant morbidity.
We developed and tested a pleural vacuum relief (PVR) device which automatically manages these risks, whilst allowing more rapid hyperbaric pressure change rates.
Methods: The custom-made PVR device consists of a one-way pressure relief valve connected in line with a sterile micro filter selected for its specific flow capacity.
The PVR device is designed for connection to the patient side sampling port of a PDU system, allowing inflow of ambient air whenever negative pressure is present, creating a small, controlled air leak which prevents excessive negative pressure.
The hyperbaric performance of a Pleur-Evac A-6000 intercostal drain was assessed with and without this added device by measuring simulated IPP with an electronic pressure monitor connected at the patient end of the PDU.
IPP readings were taken at 10, 15, 20 and 30 cmH2O of suction (set on the drain unit) at compression rates of 10, 30, 60, 80, 90 and 180 kPa·min-1 to a pressure of 280 kPa.
Results: At any compression rate of > 10 kPa·min-1, the negative IPP generated by the Pleur-Evac A-6000 alone was excessive and resulted in back flow through the PDU water seal.
By adding the PVR device, the generated negative IPP remains within a clinically acceptable range, allowing compression rates of at least 30 kPa·min-1 with suction settings up to -20 cmH2O during all phases of hyperbaric treatment.
Conclusions: The PDU PVR device we have developed works well, minimising attendant workload and automatically avoiding the excessive negative IPPs that can otherwise occur.
This device should only be used with suction.
Related Results
Blunt Chest Trauma and Chylothorax: A Systematic Review
Blunt Chest Trauma and Chylothorax: A Systematic Review
Abstract
Introduction: Although traumatic chylothorax is predominantly associated with penetrating injuries, instances following blunt trauma, as a rare and challenging condition, ...
Emerging Evidence of IgG4-Related Disease in Pericarditis: A Systematic Review
Emerging Evidence of IgG4-Related Disease in Pericarditis: A Systematic Review
Abstract
Introduction
Immunoglobulin G4-related disease (IgG4-RD) is a recently identified immune-mediated condition that is debilitating and often overlooked. While IgG4-RD has be...
Challenging Management of Postoperative Empyema: A Case Report with Literature Review
Challenging Management of Postoperative Empyema: A Case Report with Literature Review
Abstract
Introduction: Pleural empyema is the collection of pus within the pleural cavity, typically arising as a complication of pneumonia, chest trauma, thoracic surgery, or bact...
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
EnvironmentalSurveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) This comprehensive protocol suite enables systematic environmental surveillance for avian influenza...
Perbedaan cost-effectiveness pengangkatan drain kurang dari 3 hari dengan lebih dari 3 hari pada modified radical mastectomy
Perbedaan cost-effectiveness pengangkatan drain kurang dari 3 hari dengan lebih dari 3 hari pada modified radical mastectomy
Introduction: Modified radical mastectomy (MRM) is a therapeutic choice for operable breast cancer. The most frequent post-surgery complication is seroma formation, and this can be...
Dynamic Simulation to Determine Governing Relief Load for De-Methanizer System
Dynamic Simulation to Determine Governing Relief Load for De-Methanizer System
Abstract
The existing design of de-methanizer column relief system is re-validated to ensure process safety and integrity. Relief load estimation for de-methanizer b...
Differences in vacuum suction drainage pressure following mastectomy and its impact on the clinical outcome of breast cancer patients
Differences in vacuum suction drainage pressure following mastectomy and its impact on the clinical outcome of breast cancer patients
Background: Vacuum suction drainage is an obligatory practice for long past following mastectomy for breast cancer. But recent studies are showing that the pressure of the vacuum s...
Hyperbaric medicine simulation education curriculum: CNS toxicity during Treatment Table 6 and intubated ICU patient with mucous plugging
Hyperbaric medicine simulation education curriculum: CNS toxicity during Treatment Table 6 and intubated ICU patient with mucous plugging
Introduction/Background: The incidence of complications and number of critically ill patients in hyperbaric medicine is relatively low [1]. This poses a challenge to those tasked w...