Javascript must be enabled to continue!
Early evaluation of optic nerve head morphology and choroidal thickness after PreserFlo MicroShunt implantation
View through CrossRef
Abstract
Purpose
The aim of this study is to investigate changes in choroidal and optic nerve morphological parameters following MicroShunt PreserFlo implantation. The secondary aim is to investigate how the structural changes relate to the decrease in intraocular pressure (IOP).
Methods
Prospective observational study on 15 eyes with glaucoma requiring MicroShunt implantation. Optical coherence tomography was used to measure macular choroidal thickness (MCT), peripapillary choroidal thickness (PCT), lamina cribrosa depth (LCD), cup depth and prelaminar tissue thickness (PLT), before and one day after surgery. Results were expressed in median and interquartile range (IQR) and correlated with IOP results.
Results
The IOP decreased from a median of 25 (IQR = 11) mmHg to 8 (IQR = 2) mmHg the day after surgery. Median MCT increased after MicroShunt implantation from 252.1 (IQR = 156.4) µm to a postoperative value of 318.1 (IQR = 166.6) µm (p < 0.001), with a median increase of + 87.7 µm (+ 26.4%). PCT increased from 157.2 (IQR = 109.1) µm before surgery to 206.0 (IQR = 136.1) µm after surgery (p < 0.001). Moreover, we found a significant post-operative decrease in cup depth (median reduction of − 29.3 µm, p < 0.001) and an increase in PLT (median increase of 27.3 µm, p = 0.028). On the other side, LCD reduction 24 h after surgery didn’t reach any statistical significance.
Conclusion
PreserFlo implantation determines retinal structural changes which appear similar to those caused by traditional filtering surgery, confirming the effectiveness of this device, meantime carrying a much smaller complications rate when compared to trabeculectomy.
Springer Science and Business Media LLC
Title: Early evaluation of optic nerve head morphology and choroidal thickness after PreserFlo MicroShunt implantation
Description:
Abstract
Purpose
The aim of this study is to investigate changes in choroidal and optic nerve morphological parameters following MicroShunt PreserFlo implantation.
The secondary aim is to investigate how the structural changes relate to the decrease in intraocular pressure (IOP).
Methods
Prospective observational study on 15 eyes with glaucoma requiring MicroShunt implantation.
Optical coherence tomography was used to measure macular choroidal thickness (MCT), peripapillary choroidal thickness (PCT), lamina cribrosa depth (LCD), cup depth and prelaminar tissue thickness (PLT), before and one day after surgery.
Results were expressed in median and interquartile range (IQR) and correlated with IOP results.
Results
The IOP decreased from a median of 25 (IQR = 11) mmHg to 8 (IQR = 2) mmHg the day after surgery.
Median MCT increased after MicroShunt implantation from 252.
1 (IQR = 156.
4) µm to a postoperative value of 318.
1 (IQR = 166.
6) µm (p < 0.
001), with a median increase of + 87.
7 µm (+ 26.
4%).
PCT increased from 157.
2 (IQR = 109.
1) µm before surgery to 206.
0 (IQR = 136.
1) µm after surgery (p < 0.
001).
Moreover, we found a significant post-operative decrease in cup depth (median reduction of − 29.
3 µm, p < 0.
001) and an increase in PLT (median increase of 27.
3 µm, p = 0.
028).
On the other side, LCD reduction 24 h after surgery didn’t reach any statistical significance.
Conclusion
PreserFlo implantation determines retinal structural changes which appear similar to those caused by traditional filtering surgery, confirming the effectiveness of this device, meantime carrying a much smaller complications rate when compared to trabeculectomy.
Related Results
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Objective: To determine the frequency of common chromosomal aberrations in local population idiopathic determine the frequency of common chromosomal aberrations in local population...
Münsteraner Standard: modifizierte PAUL®-Implantation bei Zustand nach PreserFlo® MicroShunt-Implantation. Videobeitrag
Münsteraner Standard: modifizierte PAUL®-Implantation bei Zustand nach PreserFlo® MicroShunt-Implantation. Videobeitrag
Zusammenfassung
Operationsziel
Ziel dieser Operationstechnik ist die modifizierte Anlage eines PAUL®-Implantates (Fa. Advanced Ophthalmic Innovation...
Staphylococcus Capitus Blebitis Following Preserflo MicroShunt Implantation
Staphylococcus Capitus Blebitis Following Preserflo MicroShunt Implantation
Purpose:
To report a case of early-onset blebitis following Preserflo MicroShunt implantation.
Patients and Methods:
Case ...
Differential Diagnosis of Neurogenic Thoracic Outlet Syndrome: A Review
Differential Diagnosis of Neurogenic Thoracic Outlet Syndrome: A Review
Abstract
Thoracic outlet syndrome (TOS) is a complex and often overlooked condition caused by the compression of neurovascular structures as they pass through the thoracic outlet. ...
Anterior Segment-Optical Coherence Tomography Bleb Morphology Comparison in Minimally Invasive Glaucoma Surgery: XEN Gel Stent vs. PreserFlo MicroShunt
Anterior Segment-Optical Coherence Tomography Bleb Morphology Comparison in Minimally Invasive Glaucoma Surgery: XEN Gel Stent vs. PreserFlo MicroShunt
Background: The purpose of this study is to compare the morphology of six-month follow-up blebs created by a subconjunctival glaucoma surgical device (XEN45) to those created by a ...
Features of the Choroidal Structure in Children With Anisometropic Amblyopia
Features of the Choroidal Structure in Children With Anisometropic Amblyopia
Purpose:
To examine the choroidal structure in children with anisometropic amblyopia using the binarization method.
Methods:
...
Choroidal changes in patients with multiple sclerosis
Choroidal changes in patients with multiple sclerosis
PurposeTo evaluate the peripapillar choroidal thickness in patients with multiple sclerosis (MS) and to compare it with age‐ and sex‐ matched healthy controls, using a swept‐source...
PreserFlo® MicroShunt: An Overview of This Minimally Invasive Device for Open-Angle Glaucoma
PreserFlo® MicroShunt: An Overview of This Minimally Invasive Device for Open-Angle Glaucoma
For moderate-to-severe glaucoma, trabeculectomy remains the “gold standard” intraocular pressure (IOP)-lowering treatment; nonetheless, this method requires extensive post-operativ...