Shortened 'Thrombin Time' Monitoring on QCM-D: A Better Substitute of 'Gold Standard'
View through CrossRef
Thrombin is the central enzyme in the coagulation cascade. Recently, modern technologies for thrombin generation measurement have shown a multidisciplinary interest in methods and assays. This is first report to recognize and optimize for human plasma application of 'thrombin time' (TT) assay for quartz crystal microbalance with dissipation (QCM-D) technique. QCM-D technique has been studied comparatively with mechanical coagulometer (which is regarded as 'gold standard' for coagulation assays). The optimized protocol for TT has been applied for plasma samples (n=54) on both platforms. Thrombin times (TTs) on QCM-D platform versus those of 'gold standard' yielded a promising correlation line with R2 value of 0.86. For TT assay, QCM-D platform proved superior as compared to 'gold standard' platform due to following four edges. Firstly, QCM-D platform provides whole kinetic information, including monitoring of total coagulation on its measurement curve. 'Gold standard' cannot yield this information because it picks one point during coagulation. Secondly, TTs on QCM-D platform are 30% shorter as compared to those of 'gold standard'. Thirdly, TTs on QCM-D technique produced 16% lower %RSD demonstrating lower variability. Lastly, a historical lowest sample volume consumption of 2.50 µl has been applied on QCM-D platform. ‘Gold standard’ employs 40 times greater sample volume consumption for laboratory experiments of TT. Additionally, 40 times lower reagent volume consumption has been employed on QCM-D platform in comparison to its counterpart’s. These advantageous features are substantial support for point of care (POC) settings for TT assay via QCM-D technique.
UK Journal of Pharmaceutical and Biosciences
Title: Shortened 'Thrombin Time' Monitoring on QCM-D: A Better Substitute of 'Gold Standard'
Description:
Thrombin is the central enzyme in the coagulation cascade.
Recently, modern technologies for thrombin generation measurement have shown a multidisciplinary interest in methods and assays.
This is first report to recognize and optimize for human plasma application of 'thrombin time' (TT) assay for quartz crystal microbalance with dissipation (QCM-D) technique.
QCM-D technique has been studied comparatively with mechanical coagulometer (which is regarded as 'gold standard' for coagulation assays).
The optimized protocol for TT has been applied for plasma samples (n=54) on both platforms.
Thrombin times (TTs) on QCM-D platform versus those of 'gold standard' yielded a promising correlation line with R2 value of 0.
86.
For TT assay, QCM-D platform proved superior as compared to 'gold standard' platform due to following four edges.
Firstly, QCM-D platform provides whole kinetic information, including monitoring of total coagulation on its measurement curve.
'Gold standard' cannot yield this information because it picks one point during coagulation.
Secondly, TTs on QCM-D platform are 30% shorter as compared to those of 'gold standard'.
Thirdly, TTs on QCM-D technique produced 16% lower %RSD demonstrating lower variability.
Lastly, a historical lowest sample volume consumption of 2.
50 µl has been applied on QCM-D platform.
‘Gold standard’ employs 40 times greater sample volume consumption for laboratory experiments of TT.
Additionally, 40 times lower reagent volume consumption has been employed on QCM-D platform in comparison to its counterpart’s.
These advantageous features are substantial support for point of care (POC) settings for TT assay via QCM-D technique.
Related Results
Thrombin Induces Protease Activated Receptor (PAR)-3/-4 Interaction in Human Podocytes
Thrombin Induces Protease Activated Receptor (PAR)-3/-4 Interaction in Human Podocytes
Abstract
INTRODUCTION
Nephrotic Syndrome, one of the most common forms of glomerular disease, is characterized by massive proteinuria with structural ...
Thrombin interaction with platelet glycoprotein Ib: effect of glycocalicin on thrombin specificity
Thrombin interaction with platelet glycoprotein Ib: effect of glycocalicin on thrombin specificity
We describe here the alteration of thrombin specificity induced by its interaction with glycocalicin. Glycocalicin is the external part of platelet glycoprotein Ib alpha (GPIb alph...
A Simultaneous Monitoring of Coagulation Time and Fibrinogen via PiCT on QCM-D
A Simultaneous Monitoring of Coagulation Time and Fibrinogen via PiCT on QCM-D
‘Prothrombinase induced Clotting Time’ (PiCT) has potential to detect all anticoagulants in clinics. In the present study, PiCT has been used as a tool for anticoagulant detection ...
Decreased Immunogenicity of Purified Topical Bovine Thrombin Preparations.
Decreased Immunogenicity of Purified Topical Bovine Thrombin Preparations.
Abstract
Abstract 4209
It has been reported that severe coagulopathy following exposure to topical bovine thrombin may be attributed to the impurities...
An Antibody to Tissue Factor Pathway Inhibitor (PF-06741086) in Combination with Recombinant Factor VIIa Increases Hemostasis in Hemophilia Plasma without Excessive Thrombin Generation
An Antibody to Tissue Factor Pathway Inhibitor (PF-06741086) in Combination with Recombinant Factor VIIa Increases Hemostasis in Hemophilia Plasma without Excessive Thrombin Generation
Abstract
Hemophilia is a hereditary bleeding disorder caused by intrinsic coagulation pathway deficiencies of Factor VIII (hemophilia A) or Factor IX (hemophilia B)....
Thrombin induces endothelial type II activation in vitro: IL-1 and TNF-alpha-independent IL-8 secretion and E-selectin expression
Thrombin induces endothelial type II activation in vitro: IL-1 and TNF-alpha-independent IL-8 secretion and E-selectin expression
Abstract
In addition to its role in coagulation, thrombin is involved in the inflammatory process by inducing vessel neutrophilic infiltration. Thrombin induces e...
Prothrombin Time (PT) for Human Plasma on QCM-D Platform: A Better Alternative to 'Gold Standard'
Prothrombin Time (PT) for Human Plasma on QCM-D Platform: A Better Alternative to 'Gold Standard'
Measurements of hemostasis parameters such as Prothrombin time (PT) are substantial in various clinical cases of extensive surgery, dialysis or innate hemostasis disorders. This is...
Thrombin Interaction with Platelet GPIB: Role of the Heparin Binding Domain
Thrombin Interaction with Platelet GPIB: Role of the Heparin Binding Domain
SummaryThe platelet membrane glycoprotein lb (Gplb) has a high affinity binding site for α-thrombin whose occupancy is thought to positively modulate the thrombin-induced platelet ...