Infrared Skin Thermometry: Validating and Comparing Techniques to Detect Periwound Skin Infection
View through CrossRef
ABSTRACTOBJECTIVE:Diagnosis of wound infection can be challenging because it relies on a combination of clinical signs and symptoms that are often nonspecific. Increased periwound cutaneous temperature is a classic sign of deep and surrounding wound infection, and its quantitative measurement is one of the most reliable and valid clinical signs of deep and surrounding skin infection at the bedside. Skin surface temperature differences may be detected using commercially available noncontact infrared thermometers. However, techniques to detect temperature using noncontact infrared thermometers vary, and no studies have evaluated these methods. Two such measurement techniques include the “4-point” and “whole-wound” scanning methods. This study assessed the ability of different infrared thermometers using the aforementioned techniques to detect clinically meaningful differences in periwound cutaneous temperatures used in the diagnosis of wound infection.METHODS:A prospective cohort study was conducted from 2015 to 2016 of consenting adult patients 18 years or older with an open wound attending a regional wound care clinic. One hundred patients with wounds underwent surface temperature measurement. Infection was not a specific inclusion criterion as the primary objective was to conduct a comparative assessment of infrared thermometry devices. Demographic data (age, height, weight, gender, and ethnicity) were also collected. Each wound was measured using 4 different noncontact infrared thermometers: Exergen DermaTemp 1001 (Exergen Corporation, Watertown, Massachusetts), Mastercraft Digital Temperature Reader (Mastercraft, Toronto, Ontario, Canada), Mastercool MSC52224-A (Mastercool Inc, Randolph, New Jersey), and Etekcity ETC-8250 Temperature Heat Pen (Etekcity, Anaheim, California). Data analysis was based on a comparison of 4 periwound skin surface temperature measurement differences (ΔTin degrees Fahrenheit) between the wound site and an equivalent contralateral control site.OUTCOME MEASURES:The primary outcome of the ability of each thermometer to detect a clinically significant difference in temperature was assessed withχ2analysis. Pairedttests were conducted to compare the ΔTmeasurements of each specific thermometry device between the 2 measurement techniques. Pearson product moment correlation coefficients were calculated for the temperature ΔTfor both measuring techniques for all devices to determine level of agreement. A 1-way analysis of variance was conducted to compare temperature measurements among the infrared thermometry devices.MAIN RESULTS:There was no significant difference in the ability of each thermometer to detect a clinically meaningful difference of 3° F by either the 4-point (P= .10) or whole-wound techniques (P= .67). When a definition of 4° F was used, results were similar (4-point,P= .15; whole wound,P= .20). Comparisons among devices and techniques showed strong correlations (>0.80). Etekcity ETC-8250 and the 4-point measurement compared with the Exergen device using the whole-wound technique had a correlation coefficient of 0.72. Spearmanρdemonstrated a similarly high degree of correlation between techniques and devices, and only the Etekcity ETC-8250 device had a coefficient of 0.71 to 0.90 when compared with others. Pairedttesting for each thermometry device comparing measurement techniques for raw temperatures ΔTdemonstrated no significant difference (P> .05). No statistical differences among the ΔTvalues for the 3 different thermometers were noted when using the whole-wound method (P= .095). Similarly, no significant differences among the ΔTvalues were noted for the 4 different thermometers when using the 4-point method (P= .10).CONCLUSIONS:The results of this study demonstrate that both the 4-point and whole-wound methods of temperature acquisition using cost-efficient infrared thermometers provide accurate and similar results in clinical wound care settings.
Ovid Technologies (Wolters Kluwer Health)
Title: Infrared Skin Thermometry: Validating and Comparing Techniques to Detect Periwound Skin Infection
Description:
ABSTRACTOBJECTIVE:Diagnosis of wound infection can be challenging because it relies on a combination of clinical signs and symptoms that are often nonspecific.
Increased periwound cutaneous temperature is a classic sign of deep and surrounding wound infection, and its quantitative measurement is one of the most reliable and valid clinical signs of deep and surrounding skin infection at the bedside.
Skin surface temperature differences may be detected using commercially available noncontact infrared thermometers.
However, techniques to detect temperature using noncontact infrared thermometers vary, and no studies have evaluated these methods.
Two such measurement techniques include the “4-point” and “whole-wound” scanning methods.
This study assessed the ability of different infrared thermometers using the aforementioned techniques to detect clinically meaningful differences in periwound cutaneous temperatures used in the diagnosis of wound infection.
METHODS:A prospective cohort study was conducted from 2015 to 2016 of consenting adult patients 18 years or older with an open wound attending a regional wound care clinic.
One hundred patients with wounds underwent surface temperature measurement.
Infection was not a specific inclusion criterion as the primary objective was to conduct a comparative assessment of infrared thermometry devices.
Demographic data (age, height, weight, gender, and ethnicity) were also collected.
Each wound was measured using 4 different noncontact infrared thermometers: Exergen DermaTemp 1001 (Exergen Corporation, Watertown, Massachusetts), Mastercraft Digital Temperature Reader (Mastercraft, Toronto, Ontario, Canada), Mastercool MSC52224-A (Mastercool Inc, Randolph, New Jersey), and Etekcity ETC-8250 Temperature Heat Pen (Etekcity, Anaheim, California).
Data analysis was based on a comparison of 4 periwound skin surface temperature measurement differences (ΔTin degrees Fahrenheit) between the wound site and an equivalent contralateral control site.
OUTCOME MEASURES:The primary outcome of the ability of each thermometer to detect a clinically significant difference in temperature was assessed withχ2analysis.
Pairedttests were conducted to compare the ΔTmeasurements of each specific thermometry device between the 2 measurement techniques.
Pearson product moment correlation coefficients were calculated for the temperature ΔTfor both measuring techniques for all devices to determine level of agreement.
A 1-way analysis of variance was conducted to compare temperature measurements among the infrared thermometry devices.
MAIN RESULTS:There was no significant difference in the ability of each thermometer to detect a clinically meaningful difference of 3° F by either the 4-point (P= .
10) or whole-wound techniques (P= .
67).
When a definition of 4° F was used, results were similar (4-point,P= .
15; whole wound,P= .
20).
Comparisons among devices and techniques showed strong correlations (>0.
80).
Etekcity ETC-8250 and the 4-point measurement compared with the Exergen device using the whole-wound technique had a correlation coefficient of 0.
72.
Spearmanρdemonstrated a similarly high degree of correlation between techniques and devices, and only the Etekcity ETC-8250 device had a coefficient of 0.
71 to 0.
90 when compared with others.
Pairedttesting for each thermometry device comparing measurement techniques for raw temperatures ΔTdemonstrated no significant difference (P> .
05).
No statistical differences among the ΔTvalues for the 3 different thermometers were noted when using the whole-wound method (P= .
095).
Similarly, no significant differences among the ΔTvalues were noted for the 4 different thermometers when using the 4-point method (P= .
10).
CONCLUSIONS:The results of this study demonstrate that both the 4-point and whole-wound methods of temperature acquisition using cost-efficient infrared thermometers provide accurate and similar results in clinical wound care settings.
Related Results
Comparison of skin and rectal thermometry in Sokoto Gudali cattle
Comparison of skin and rectal thermometry in Sokoto Gudali cattle
Measuring temperature with a rectal thermometer in conscious animals can be exacting and time-consuming, especially in a herd without adequate restraint. This research compared the...
Newborn friendly thermometry – Comparative study of body temperature with an infrared versus digital thermometer
Newborn friendly thermometry – Comparative study of body temperature with an infrared versus digital thermometer
Background: Accurate measurement of body temperature is of great importance in day to day neonatology practice. One touch infrared thermometry is safe, accurate, and easy to use. O...
Accuracy of Daily Foot Temperature Monitoring for Patients with Recently Healed Diabetic Foot Ulcers or History of Amputation
Accuracy of Daily Foot Temperature Monitoring for Patients with Recently Healed Diabetic Foot Ulcers or History of Amputation
Diabetic foot ulcers (DFU) result in significant morbidity, mortality, and healthcare costs. Fortunately, recent data suggest as many as 75% of DFU and 60% of lower extremity amput...
EPD Electronic Pathogen Detection v1
EPD Electronic Pathogen Detection v1
Electronic pathogen detection (EPD) is a non - invasive, rapid, affordable, point- of- care test, for Covid 19 resulting from infection with SARS-CoV-2 virus. EPD scanning techno...
Rancang Bangun Sistem Terapi Infrared Otomatis Untuk Terapi Far-Infrared Pada Spinal Cord
Rancang Bangun Sistem Terapi Infrared Otomatis Untuk Terapi Far-Infrared Pada Spinal Cord
Terapi infrared adalah terapi yang banyak dilakukan dengan memanfaatkan efek panas yang ditimbulkan pada tubuh manusia terutama pada permukaan kulit yang terpapar secara langsung. ...
Design of high transparent infrared stealth thin films based on FTO/Ag/FTO structure
Design of high transparent infrared stealth thin films based on FTO/Ag/FTO structure
Multi-spectral compatible stealth materials have become an imperative development trend, especially visible and infrared compatible stealth materials have become the most important...
Association between skin-to-skin contact post-birth and breastfeeding behaviour: a cross-sectional study of Nigerian women using the 2018 Demographic Health Survey
Association between skin-to-skin contact post-birth and breastfeeding behaviour: a cross-sectional study of Nigerian women using the 2018 Demographic Health Survey
Background The World Health Organisation (WHO) recommends skin-to-skin contact at birth as part of the essential newborn care processes that can help improve breastfeeding behaviou...
[RETRACTED] Kerassentials Toenail Fungus Treatment, Kerassentials for Toenail Fungus, Kerasentials Nail Treatment v1
[RETRACTED] Kerassentials Toenail Fungus Treatment, Kerassentials for Toenail Fungus, Kerasentials Nail Treatment v1
[RETRACTED]kerassentials reviewsis another nail and skin support dietary enhancement would you say you are tired of your foot parasite? Do you need to confront a steady humiliation...