Aortic valve morphology and the double envelope phenomenon in continuous Doppler echocardiography of aortic stenosis: a computerized model

View through CrossRef
Abstract Introduction Visualization of a double envelope (DE) is a relatively common issue in aortic transvalvular continuous-wave Doppler (CW) traces. In 1997, Song et al proposed that velocity time integral (VTI) measured at the edge of the inner envelope (IE) might equal that of the left ventricular outflow tract (LVOT). This surrogate measurement would be especially interesting in patients with irregular heart rhythms, as both LVOT and valvular VTIs could be obtained from the same beat, thus avoiding inter-beat variability when assessing aortic valve area (AVA) with continuity equation. In contrast, based on clinical data and 3D-printed models, they raised concern that DE method in dome-shaped valves could lead to significant overestimation of LVOT velocities (about 40%). Later studies, including a Partner trial subanalysis, question the validity of this measurement. Purpose We created a computerized aortic valve model to reproduce the CW tracing, and analysed how valve morphology may lead to a valid, well defined inner envelope or, by contrast, produce overestimation of LVOT velocities. Methods The simulation code was created using Python (v3.9.6) along with NumPy and Matplotlib libraries. For each simulation, an aortic valve scheme was created (LVOT diameter: 2 cm; TSVI length: 1 cm; LVOT peak velocity 1.1 m/s; varying morphologies and lengths of the aortic cusps). The longitudinal axis between LVOT base and valve tip was divided in sections or levels (at least 300 per emulation). To recreate the CW trace, for each level, luminal diameter and peak velocity (according to continuity equation) were calculated, and an array of points was generated representing the aortic waveform during the entire systole. Slight random dispersion (standard deviation 0.05 times original values) was added to produce a more realistic image. Data points were represented as scatter plot over black background with transparency, simulating CW tracings (Figure 1) Results Figure 2 represents the plots of various aortic valve configurations (1st row AVA 1.6 cm2; 2nd: 1.2 cm2; 3rd: 0.8 cm2). Note that the innermost portion of the trace results in hollow space, as these velocities are produced in portions of ventricle and aorta which are not included in the simulation. Pink dashed line represents actual LVOT velocities. In a short-length flat valve (A), the IE is well defined and shows good correlation with LVOT velocity. In a cone-shaped 6mm length valve (B), there is a progressive brightness reduction, with no recognizable contour. With long (6 mm) dome-shaped valves, an ill-defined IE is produced at higher velocities than LVOT, which could result in overestimation. Conclusion DE method could accurately assess LVOT velocities when a sharp-defined IE is present and a dome-shaped valve can be excluded (e.g. rheumatic or some bicuspid valves). Measurements in a blurred DE or in a dome-shaped valve would produce overestimation of LVOT velocities and AVA. Funding Acknowledgement Type of funding sources: None.
Title: Aortic valve morphology and the double envelope phenomenon in continuous Doppler echocardiography of aortic stenosis: a computerized model
Description:
Abstract Introduction Visualization of a double envelope (DE) is a relatively common issue in aortic transvalvular continuous-wave Doppler (CW) traces.
In 1997, Song et al proposed that velocity time integral (VTI) measured at the edge of the inner envelope (IE) might equal that of the left ventricular outflow tract (LVOT).
This surrogate measurement would be especially interesting in patients with irregular heart rhythms, as both LVOT and valvular VTIs could be obtained from the same beat, thus avoiding inter-beat variability when assessing aortic valve area (AVA) with continuity equation.
In contrast, based on clinical data and 3D-printed models, they raised concern that DE method in dome-shaped valves could lead to significant overestimation of LVOT velocities (about 40%).
Later studies, including a Partner trial subanalysis, question the validity of this measurement.
Purpose We created a computerized aortic valve model to reproduce the CW tracing, and analysed how valve morphology may lead to a valid, well defined inner envelope or, by contrast, produce overestimation of LVOT velocities.
Methods The simulation code was created using Python (v3.
9.
6) along with NumPy and Matplotlib libraries.
For each simulation, an aortic valve scheme was created (LVOT diameter: 2 cm; TSVI length: 1 cm; LVOT peak velocity 1.
1 m/s; varying morphologies and lengths of the aortic cusps).
The longitudinal axis between LVOT base and valve tip was divided in sections or levels (at least 300 per emulation).
To recreate the CW trace, for each level, luminal diameter and peak velocity (according to continuity equation) were calculated, and an array of points was generated representing the aortic waveform during the entire systole.
Slight random dispersion (standard deviation 0.
05 times original values) was added to produce a more realistic image.
Data points were represented as scatter plot over black background with transparency, simulating CW tracings (Figure 1) Results Figure 2 represents the plots of various aortic valve configurations (1st row AVA 1.
6 cm2; 2nd: 1.
2 cm2; 3rd: 0.
8 cm2).
Note that the innermost portion of the trace results in hollow space, as these velocities are produced in portions of ventricle and aorta which are not included in the simulation.
Pink dashed line represents actual LVOT velocities.
In a short-length flat valve (A), the IE is well defined and shows good correlation with LVOT velocity.
In a cone-shaped 6mm length valve (B), there is a progressive brightness reduction, with no recognizable contour.
With long (6 mm) dome-shaped valves, an ill-defined IE is produced at higher velocities than LVOT, which could result in overestimation.
Conclusion DE method could accurately assess LVOT velocities when a sharp-defined IE is present and a dome-shaped valve can be excluded (e.
g.
rheumatic or some bicuspid valves).
Measurements in a blurred DE or in a dome-shaped valve would produce overestimation of LVOT velocities and AVA.
Funding Acknowledgement Type of funding sources: None.

Related Results

Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Objective: To determine the frequency of common chromosomal aberrations in local population idiopathic determine the frequency of common chromosomal aberrations in local population...
A case of Unicuspid Aortic Valve with Hidden Subaortic Ridge
A case of Unicuspid Aortic Valve with Hidden Subaortic Ridge
57-years old man presented with exertional dyspnea. An early systolic murmur was heard over the aortic areas 2D and 3D Echocardiography revealed unicuspid , unicommissural aortic v...
CFD Based on The Visualisation of Aortic Valve Mechanism in Aortic Valve Stenosis for Risk Prediction at The Peak Velocity
CFD Based on The Visualisation of Aortic Valve Mechanism in Aortic Valve Stenosis for Risk Prediction at The Peak Velocity
Aortic valve disease plays a crucial role in the development of cardiovascular disease (CVD), leading to increased rates of mortality and morbidity. Two diseases, aortic valve regu...
Critical Arterial Stenosis Revisited
Critical Arterial Stenosis Revisited
Abstract Introduction Stenosis of an organ/tissue primary artery can produce ischemia or only reduce blood flow reserve. Despit...
Abstract 4369914: Successful Transfemoral TAVR in a Patient with Chronic Aortic Dissection and Severe Aortic Insufficiency
Abstract 4369914: Successful Transfemoral TAVR in a Patient with Chronic Aortic Dissection and Severe Aortic Insufficiency
Transcatheter aortic valve replacement is a well-established treatment for high-surgical-risk patients with severe aortic disease, providing a less invasive alternative to traditio...

Back to Top