Intraocular Pressure Measurements in Standing, Sitting, and Supine Position: Comparison between Tono-Pen Avia and Icare Pro Tonometers
View through CrossRef
Background: Intraocular pressure (IOP) is influenced by body position. The purpose of this study is to compare the IOP measurements obtained with two different devices, to investigate IOP changes in standing, sitting, and supine positions. Methods: In this comparative prospective case series, IOP was measured in sitting, supine, prone, and standing (standing 1) positions and again five minutes after standing (standing 2), utilizing an Icare Pro (ICP) and a Tono-Pen Avia (TPA) in the 64 eyes of 32 healthy subjects. Results: Compared to the sitting position, both devices showed an increase in the IOP both in supine and standing 2 positions (p < 0.05). The mean IOP difference between the two devices was: in the sitting position, 0.57 ± 2.10 mmHg (range: −3.80 to 6.60 mmHg) (p < 0.05), in the supine position, 0.93 ± 2.49 mmHg (range: −4.50 to 7.10 mmHg) (p < 0.05), in the standing 1 position, 0.37 ± 1.96 mmHg (range: −5.20 to 5.00 mmHg) (p = 0.102), and in the standing 2 position 0.73 ± 2.03 mmHg (range: −4.5 to 6.4 mmHg) (p < 0.001). Conclusions: The results highlight an agreement between the TPA and ICP, both confirming not only the increase in IOP in the supine position, but also showing an increase in the standing 2 position. Therefore, it is suggested to perform such measurements in patients with glaucoma, to explain its progression in an apparently normal tension or in compensated patients.
Title: Intraocular Pressure Measurements in Standing, Sitting, and Supine Position: Comparison between Tono-Pen Avia and Icare Pro Tonometers
Description:
Background: Intraocular pressure (IOP) is influenced by body position.
The purpose of this study is to compare the IOP measurements obtained with two different devices, to investigate IOP changes in standing, sitting, and supine positions.
Methods: In this comparative prospective case series, IOP was measured in sitting, supine, prone, and standing (standing 1) positions and again five minutes after standing (standing 2), utilizing an Icare Pro (ICP) and a Tono-Pen Avia (TPA) in the 64 eyes of 32 healthy subjects.
Results: Compared to the sitting position, both devices showed an increase in the IOP both in supine and standing 2 positions (p < 0.
05).
The mean IOP difference between the two devices was: in the sitting position, 0.
57 ± 2.
10 mmHg (range: −3.
80 to 6.
60 mmHg) (p < 0.
05), in the supine position, 0.
93 ± 2.
49 mmHg (range: −4.
50 to 7.
10 mmHg) (p < 0.
05), in the standing 1 position, 0.
37 ± 1.
96 mmHg (range: −5.
20 to 5.
00 mmHg) (p = 0.
102), and in the standing 2 position 0.
73 ± 2.
03 mmHg (range: −4.
5 to 6.
4 mmHg) (p < 0.
001).
Conclusions: The results highlight an agreement between the TPA and ICP, both confirming not only the increase in IOP in the supine position, but also showing an increase in the standing 2 position.
Therefore, it is suggested to perform such measurements in patients with glaucoma, to explain its progression in an apparently normal tension or in compensated patients.
Related Results
COMPARISON BETWEEN SUPINE POSITION VERSUS PRONE POSITION IN PERCUTANEOUS NEPHROLITHOTOMY: A SINGLE CENTERED ANALYSIS OF 623 CASES
COMPARISON BETWEEN SUPINE POSITION VERSUS PRONE POSITION IN PERCUTANEOUS NEPHROLITHOTOMY: A SINGLE CENTERED ANALYSIS OF 623 CASES
Background: The ideal urological method for treating complex, large renal calculi is Percutaneous nephrolithotomy (PCNL). Its instruments, surgical techniques, and positions have a...
Changes in Intraocular Pressure While Using Electronic Devices in Sitting and Supine Positions
Changes in Intraocular Pressure While Using Electronic Devices in Sitting and Supine Positions
Objective: This study aimed to compare intraocular pressure (IOP) measurements in sitting and supine positions while exposed to electronic devices. Material and Methods: A total of...
Intraocular Pressure Variations in Postural Changes: Comparison between Obese and Non-Obese Controls
Intraocular Pressure Variations in Postural Changes: Comparison between Obese and Non-Obese Controls
Background: Comparing intraocular pressure (IOP) changes (ΔIOP) between obese subjects and non-obese controls in relation to different positions: standing, sitting, supine. Methods...
PAC System for Radiographic Comparison of Standing and Supine Lateral Views in Patients with Low-Grade Spondylolisthesis
PAC System for Radiographic Comparison of Standing and Supine Lateral Views in Patients with Low-Grade Spondylolisthesis
Objective: Supine views may not adequately depict low-degree spondylolisthesis, thereby confusing surgeons in deciding the best therapy plan. The study aimed to compare the standi...
Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer
Comparison of Intraocular Pressure Measurement by iCare Tonometer and Goldman Applanation Tonometer
Objective: To compare the Intraocular Pressure measured by the Goldmann Applanation tonometer with by iCare ic100 tonometer in normal human eye to determine the accuracy and reli...
Numerical analysis of the pulse wave in a hydro-cuff blood pressure measurement system
Numerical analysis of the pulse wave in a hydro-cuff blood pressure measurement system
For people suffering from hypertension, the procedure of constant monitoring of blood pressure values becomes mandatory. Accurate and timely assessment of their values allows to ev...
Assessing the True Intraocular Pressure in the Non-human Primate
Assessing the True Intraocular Pressure in the Non-human Primate
SIGNIFICANCEFor glaucoma patients, high intraocular pressure (IOP) is a risk factor for progressive neuropathy. Similarly, animal models used to study the disease are based on an e...
O9-5 Workplace sitting associated with self-rated perceived global health
O9-5 Workplace sitting associated with self-rated perceived global health
Abstract
Introduction
Total self-reported sitting time is associated with higher risk for cardiometabolic disease and mortality,...