PENENTUAN TIPIKAL DOSIS PADA PEMERIKSAAN RADIOGRAFI THORAKS PA BERDASARKAN KURVA REFERENSI-DRL MENGGUNAKAN DATABASE SI-INTAN BAPETEN
View through CrossRef
Consideration of patient weight as an essential indicator in optimization efforts. This is because the radiation dose varies not only for age groups but is accompanied by changes in body weight. This research method uses a retrospective study of PA chest radiography examination results in the BAPETEN DRL database between 2018-2022. PA chest radiography examination data used: examination projection, body weight, age, source image distance, exposure factor consisting of kV and mAs, and entrance surface air kerma (ESAK) dose. Data were analyzed to create a DRL-reference curve by grouping body weight from 30-101 kg with a range of 5 kg. The first stage of data fitting is to determine the relationship between body weight (kg) and radiation dose (mGy). The second stage creates a DRL reference curve using the exponential function. Next, determine the DRL value for each weight group with the 25th percentile, 50th percentile, and 75th percentile values. Continue by determining descriptive statistical values: average, standard deviation, and range. As well as comparing the 75th percentile typical dose value between the Jakarta DRL Local value and the Indonesian National DRL value. As a result of the research, a DRL reference curve for body weight groups against radiation dose was created. DRL reference curves can be compared according to further research needs or as an optimization effort. The 75th percentile DRL reference curve for the Jakarta DRL Local data is higher than the Indonesian DRL National data. Therefore, collaboration between all parties is needed to optimize radiography procedures to control the Jakarta DRL-reference curve. Data limitations are also an important factor that can cause the shape of the exponential curve to change. Keywords: DRL-reference curve, diagnostic reference level, Thorax PA.
Badan Pengawas Tenaga Nuklir (BAPETEN)
Title: PENENTUAN TIPIKAL DOSIS PADA PEMERIKSAAN RADIOGRAFI THORAKS PA BERDASARKAN KURVA REFERENSI-DRL MENGGUNAKAN DATABASE SI-INTAN BAPETEN
Description:
Consideration of patient weight as an essential indicator in optimization efforts.
This is because the radiation dose varies not only for age groups but is accompanied by changes in body weight.
This research method uses a retrospective study of PA chest radiography examination results in the BAPETEN DRL database between 2018-2022.
PA chest radiography examination data used: examination projection, body weight, age, source image distance, exposure factor consisting of kV and mAs, and entrance surface air kerma (ESAK) dose.
Data were analyzed to create a DRL-reference curve by grouping body weight from 30-101 kg with a range of 5 kg.
The first stage of data fitting is to determine the relationship between body weight (kg) and radiation dose (mGy).
The second stage creates a DRL reference curve using the exponential function.
Next, determine the DRL value for each weight group with the 25th percentile, 50th percentile, and 75th percentile values.
Continue by determining descriptive statistical values: average, standard deviation, and range.
As well as comparing the 75th percentile typical dose value between the Jakarta DRL Local value and the Indonesian National DRL value.
As a result of the research, a DRL reference curve for body weight groups against radiation dose was created.
DRL reference curves can be compared according to further research needs or as an optimization effort.
The 75th percentile DRL reference curve for the Jakarta DRL Local data is higher than the Indonesian DRL National data.
Therefore, collaboration between all parties is needed to optimize radiography procedures to control the Jakarta DRL-reference curve.
Data limitations are also an important factor that can cause the shape of the exponential curve to change.
Keywords: DRL-reference curve, diagnostic reference level, Thorax PA.
Related Results
RADIOGRAFI SINAR-X INTERMITEN DAN KONSTAN PADA LAS
RADIOGRAFI SINAR-X INTERMITEN DAN KONSTAN PADA LAS
RADIOGRAFI SINAR-X INTERMITEN DAN KONSTAN PADA LAS. Pengaplikasian radio-grafi sinar-X sudah berkembang dan sudah banyak dimanfaatkan pada bahan metal las. Radiografi ini menggun...
FAKTOR-FAKTOR YANG MEMPENGARUHI MORTALITAS PADA PASIEN DENGAN FRAKTUR COSTA: Literature Review
FAKTOR-FAKTOR YANG MEMPENGARUHI MORTALITAS PADA PASIEN DENGAN FRAKTUR COSTA: Literature Review
FAKTOR-FAKTOR YANG MEMPENGARUHI MORTALITAS PADA PASIEN DENGAN FRAKTUR COSTA: Literature Review Anna Tri Wahyuni1), Masfuri2), Liya Arista3)1,2,3 Fakultas Ilmu Keperawatan Univers...
Verifikasi Dosis Radiasi Teknik Penyinaran 3D-CRT pada Pasien Kanker Payudara Menggunakan Film EBT3 di Rumah Sakit UNAND
Verifikasi Dosis Radiasi Teknik Penyinaran 3D-CRT pada Pasien Kanker Payudara Menggunakan Film EBT3 di Rumah Sakit UNAND
Verifikasi dosis radiasi teknik penyinaran Three-Dimensional Conformal Radiation Therapy  (3D-CRT) pada pasien kanker payudara telah dilakukan menggunakan film EBT3. Verifikasi do...
Karakteristik dan Gambaran Radiologi pada Pasien Low Back Pain
Karakteristik dan Gambaran Radiologi pada Pasien Low Back Pain
Mobilitas yang salah, penyakit psikologis, dan berbagai masalah muskuloskeletal semuanya dapat menyebabkan low back pain (LBP). Disabilitas adalah penyebab utama LBP dan berdampak ...
STUDI ANALISIS AUDIT PAPARAN MEDIK SEBAGAI UPAYA OPTIMISASI PADA CT HEAD DI RUMAH SAKIT UMUM ELISABET MEDAN
STUDI ANALISIS AUDIT PAPARAN MEDIK SEBAGAI UPAYA OPTIMISASI PADA CT HEAD DI RUMAH SAKIT UMUM ELISABET MEDAN
Studi analisis audit paparan medik sebagai upaya optimisasi pada CT HEAD di Rumah Sakit Umum Elisabet Medan. Audit dosis dilakukan pada pemeriksaan CT scan melalui metrik dosis, un...
PENERAPAN EDUKASI DAN KOMUNIKASI PASIEN MENGENAI DOSIS RADIASI SEBAGAI PENGEMBANGAN TINGKAT PANDUAN DIAGNOSTIK LOKAL
PENERAPAN EDUKASI DAN KOMUNIKASI PASIEN MENGENAI DOSIS RADIASI SEBAGAI PENGEMBANGAN TINGKAT PANDUAN DIAGNOSTIK LOKAL
Pemantauan dosis radiasi pada pasien merupakan aspek penting dalam upaya optimisasi proteksi radiasi. Penelitian ini bertujuan untuk menerapkan edukasi dan komunikasi pasien mengen...
Penerapan Optimisasi Paparan Medik CT-Scan Melalui Tingkat Panduan Diagnostik di RSUP dr. Soeradji Tirtonegoro
Penerapan Optimisasi Paparan Medik CT-Scan Melalui Tingkat Panduan Diagnostik di RSUP dr. Soeradji Tirtonegoro
Latar belakang: Pemanfaatan CT-Scan di RSUP dr. Soeradji Tirtonegoro menunjukkan adanya peningkatan yang signifikan, pemanfaatan tersebut harus dilakukan pengawasan untuk menjamin ...
Evaluasi dosis CT Simulator pada Radioterapi Head and Neck dan Pelvis
Evaluasi dosis CT Simulator pada Radioterapi Head and Neck dan Pelvis
CT Simulator merupakan modalitas pencitraan dalam Radioterapi yang digunakan untuk delineasi dan kalkulasi dosis pada perencanaan terapi. AAPM TG.204 menyediakan faktor konversi ya...