Feasibility study of shielded diode detector for small field dosimetry
View through CrossRef
Abstract
Improved imaging techniques and modern radiotherapy treatment delivery in the treatment field are reduced to the precise size of the tumor, which necessitates the need for small-field dosimetry. Dosimetry in small-field dosimetry is challenging because most of the available code of practice for dosimetry is based on the cavity theory concept. Some small-sized detectors show good spatial resolution and sensitivity. Of the available small detectors, the diamond detector's performance is remarkably good. Most of the centers for radiotherapy lack diamond detectors. In this situation, if a diode detector is available, we can use it for small-field dosimetry by applying the Daisy Chaining method correction methods. In this study, the diode detector's response is not over-responding because of the defective diode. So this diode cannot be used for further measurements, and we have to regularly check the performance of the diode before using it for measurements.
Oxford University Press (OUP)
Title: Feasibility study of shielded diode detector for small field dosimetry
Description:
Abstract
Improved imaging techniques and modern radiotherapy treatment delivery in the treatment field are reduced to the precise size of the tumor, which necessitates the need for small-field dosimetry.
Dosimetry in small-field dosimetry is challenging because most of the available code of practice for dosimetry is based on the cavity theory concept.
Some small-sized detectors show good spatial resolution and sensitivity.
Of the available small detectors, the diamond detector's performance is remarkably good.
Most of the centers for radiotherapy lack diamond detectors.
In this situation, if a diode detector is available, we can use it for small-field dosimetry by applying the Daisy Chaining method correction methods.
In this study, the diode detector's response is not over-responding because of the defective diode.
So this diode cannot be used for further measurements, and we have to regularly check the performance of the diode before using it for measurements.
Related Results
Characterization of a novel HgCdTe focal plane array for ground and space astronomy through innovative infrared setups
Characterization of a novel HgCdTe focal plane array for ground and space astronomy through innovative infrared setups
(English) Nowadays, mercury-cadmium-telluride (MCT) short-wave infrared (SWIR) detectors are widely used in cutting-edge space
missions and ground-based telescopes. They take adva...
Conceptual design report of the MPD Cosmic Ray Detector (MCORD)
Conceptual design report of the MPD Cosmic Ray Detector (MCORD)
Abstract
This report presents a concept of constructing a detector
dedicated for detection of muons observed during measurements
carried out at the MPD (Multi-Pu...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
An analytical formalism for the assessment of dose uncertainties due to positioning uncertainties
An analytical formalism for the assessment of dose uncertainties due to positioning uncertainties
PurposeTo present an analytical formalism for the in depth assessment of uncertainties of field output factors in small fields related to detector positioning based on dose profile...
SU‐GG‐T‐200: Comparison of a Novel Transmission Detector to a Standard Measurement Technique for Patient IMRT Quality Assurance
SU‐GG‐T‐200: Comparison of a Novel Transmission Detector to a Standard Measurement Technique for Patient IMRT Quality Assurance
Purpose: A new measurement‐based, pre‐treatment IMRT composite quality assurance (QA) procedure was compared to a standard QA method. The standard QA method consisted of delivering...
SU‐GG‐T‐173: Dosimetric Verification of IMRT Fields Using GafChromic® EBT‐2 Film
SU‐GG‐T‐173: Dosimetric Verification of IMRT Fields Using GafChromic® EBT‐2 Film
Purpose: Film based dosimetry often has advantages over detector based dosimetry when measurements with high spatial resolution are needed. We present comparisons of dosimetric ver...
On the implementation of a recently proposed dosimetric formalism to a robotic radiosurgery system
On the implementation of a recently proposed dosimetric formalism to a robotic radiosurgery system
Purpose:The aim of this work is to implement a recently proposed dosimetric formalism for nonstandard fields to the calibration and small field output factor measurement of a robot...